Rapid SARS-CoV-2 Inactivation in a Simulated Hospital Room Using a Mobile and Autonomous Robot Emitting Ultraviolet-C Light

被引:16
作者
Lorca-Oro, Cristina [1 ]
Vila, Jordi [2 ,3 ]
Pleguezuelos, Patricia [1 ]
Vergara-Alert, Julia [1 ]
Rodon, Jordi [1 ]
Majo, Natalia [1 ,4 ]
Lopez, Sergio [1 ]
Segales, Joaquim [1 ,4 ]
Saldana-Buesa, Francesc [5 ]
Visa-Boladeras, Maria [6 ]
Vea-Baro, Andreu [2 ,5 ]
Maria Campistol, Josep [2 ,3 ]
Abad, Xavier [1 ]
机构
[1] Ctr Recerca Sanitat Anim, Inst Recerca & Tecnol Agroalimentaries, Campus Univ Autonoma Barcelona, Barcelona, Spain
[2] Univ Barcelona, Hosp Clin, August Pi & Sunyer Biomed Res Inst, Barcelona, Spain
[3] Univ Barcelona, Inst Global Hlth ISGlobal Hosp Clin, Barcelona, Spain
[4] Univ Autonoma Barcelona, Fac Vet, Dept Sanitat & Anat Anim, Barcelona, Spain
[5] CovidWarriors, Barcelona, Spain
[6] MTS Tech, Barcelona, Spain
关键词
SARS-CoV-2; inactivation; UV-C radiation; autonomous robots; exponential viral-load reduction; virus inactivation; CORONAVIRUS;
D O I
10.1093/infdis/jiab551
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
The spread of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) since 2019 has made mask-wearing, physical distancing, hygiene, and disinfection complementary measures to control virus transmission. Especially for health facilities, we evaluated the efficacy of an UV-C autonomous robot to inactivate SARS-CoV-2 desiccated on potentially contaminated surfaces. ASSUM (autonomous sanitary sterilization ultraviolet machine) robot was used in an experimental box simulating a hospital intensive care unit room. Desiccated SARS-CoV-2 samples were exposed to UV-C in 2 independent runs of 5, 12, and 20 minutes. Residual virus was eluted from surfaces and viral titration was carried out in Vero E6 cells. ASSUM inactivated SARS-CoV-2 by >= 99.91% to >= 99.99% titer reduction with 12 minutes or longer of UV-C exposure and onwards and a minimum distance of 100 cm between the device and the SARS-CoV-2 desiccated samples. This study demonstrates that ASSUM UV-C device is able to inactivate SARS-CoV-2 within a few minutes.
引用
收藏
页码:587 / 592
页数:6
相关论文
共 23 条
[1]   Stability of SARS-CoV-2 and other coronaviruses in the environment and on common touch surfaces and the influence of climatic conditions: A review [J].
Aboubakr, Hamada A. ;
Sharafeldin, Tamer A. ;
Goyal, Sagar M. .
TRANSBOUNDARY AND EMERGING DISEASES, 2021, 68 (02) :296-312
[2]   UV-C irradiation is highly effective in inactivating SARS-CoV-2 replication [J].
Biasin, Mara ;
Bianco, Andrea ;
Pareschi, Giovanni ;
Cavalleri, Adalberto ;
Cavatorta, Claudia ;
Fenizia, Claudio ;
Galli, Paola ;
Lessio, Luigi ;
Lualdi, Manuela ;
Tombetti, Enrico ;
Ambrosi, Alessandro ;
Redaelli, Edoardo Maria Alberto ;
Saulle, Irma ;
Trabattoni, Daria ;
Zanutta, Alessio ;
Clerici, Mario .
SCIENTIFIC REPORTS, 2021, 11 (01)
[3]   Evaluation of an Ultraviolet C (UVC) Light-Emitting Device for Disinfection of High Touch Surfaces in Hospital Critical Areas [J].
Casini, Beatrice ;
Tuvo, Benedetta ;
Cristina, Maria Luisa ;
Spagnolo, Anna Maria ;
Totaro, Michele ;
Baggiani, Angelo ;
Privitera, Gaetano Pierpaolo .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2019, 16 (19)
[4]   Specific risk factors for SARS-CoV-2 transmission among health care workers in a university hospital [J].
Celebi, Guven ;
Piskin, Nihal ;
Beklevic, Arzum Celik ;
Altunay, Yurdagul ;
Keles, Aysegul Salci ;
Tuz, Mehmet Ali ;
Altinsoy, Bulent ;
Haciseyitoglu, Demet .
AMERICAN JOURNAL OF INFECTION CONTROL, 2020, 48 (10) :1225-1230
[5]   Inactivation of the coronavirus that induces severe acute respiratory syndrome, SARS-CoV [J].
Darnell, MER ;
Subbarao, K ;
Feinstone, SM ;
Taylor, DR .
JOURNAL OF VIROLOGICAL METHODS, 2004, 121 (01) :85-91
[6]   Rapid inactivation of SARS-CoV-2 with deep-UV LED irradiation [J].
Inagaki, Hiroko ;
Saito, Akatsuki ;
Sugiyama, Hironobu ;
Okabayashi, Tamaki ;
Fujimoto, Shouichi .
EMERGING MICROBES & INFECTIONS, 2020, 9 (01) :1744-1747
[7]   Article on the collective handling of pharmacological array effort. [J].
Karber, G .
NAUNYN-SCHMIEDEBERGS ARCHIV FUR EXPERIMENTELLE PATHOLOGIE UND PHARMAKOLOGIE, 1931, 162 :480-483
[8]  
Kowalski W, 2009, ULTRAVIOLET GERMICIDAL IRRADIATION HANDBOOK, P17, DOI 10.1007/978-3-642-01999-9_2
[9]   Temperature-dependent surface stability of SARS-CoV-2 [J].
Kratzel, Annika ;
Steiner, Silvio ;
Todt, Daniel ;
V'kovski, Philip ;
Brueggemann, Yannick ;
Steinmann, Joerg ;
Steinmann, Eike ;
Thiel, Volker ;
Pfaender, Stephanie .
JOURNAL OF INFECTION, 2020, 81 (03) :474-476
[10]   COVID-19 RARELY SPREADS THROUGH SURFACES. SO WHY ARE WE STILL DEEP CLEANING? [J].
Lewis, Dyani .
NATURE, 2021, 590 (7844) :26-28