Gas chromatography with capillary flow technology An effective analytical method for detecting pesticide residues in olive oil

被引:0
作者
Smith, Doris [1 ]
Lynam, Ken [1 ]
机构
[1] Agilent Technol, Wilmington, DE 19809 USA
来源
AGRO FOOD INDUSTRY HI-TECH | 2012年 / 23卷 / 01期
关键词
D O I
暂无
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The detection of residual organophosphorous pesticides in processed olive oil is complicated by the chromatographically active nature of these compounds, which compromises chromatographic resolution. This study demonstrates a quick and effective analytical method for the determination of low ppm and trace level OP pesticide residues in an olive oil extract. A J&W DB-35ms Ultra Inert (UI) 30 m x 0.25 mm, 0.25 mu m column resolved the pesticides of interest in less than 16 minutes, yielding excellent peak shape for even the more problematic OP pesticides. The detection limits for most of the pesticides were 10-15 ng/mL. A simplified QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) method provided sufficient sample matrix clean-up while preserving low-level analyte detection. A capillary flow technology (CFT) device was installed post column to split the effluent between the MSD and FPD and implement an automated back flush to diminish residual sample carryover and reduce instrument cycle times.
引用
收藏
页码:21 / 23
页数:3