Combined Prospective and Retrospective Motion Correction to Relax Navigator Requirements

被引:26
作者
Maclaren, Julian [1 ]
Lee, Kuan J. [1 ]
Luengviriya, Chaiya [2 ]
Speck, Oliver [2 ,3 ]
Zaitsev, Maxim [1 ]
机构
[1] Univ Med Ctr, Dept Radiol, D-79106 Freiburg, Germany
[2] Otto Von Guericke Univ, Dept Biomed Magnet Resonance, Magdeburg, Germany
[3] Kasetsart Univ, Dept Phys, Bangkok, Thailand
关键词
prospective motion correction; retrospective motion correction; Kalman filtering; tracking accuracy; real-time; conjugate gradient method; SENSE RECONSTRUCTION; TIME; MRI;
D O I
10.1002/mrm.22754
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Prospective motion correction can prevent motion artifacts in magnetic resonance imaging of the brain. However, for high-resolution imaging, the technique relies on precise tracking of head motion. This precision is often limited by tracking noise, which leads to residual errors in the prospectively-corrected k-space data and artifacts in the image. This work shows that it is possible to estimate these tracking errors, and hence the true k-space sample locations, by applying a two-sided filter to the tracking data after imaging. A conjugate gradient reconstruction is compared to gridding as a means of using this information to retrospectively correct for the effects of the residual errors. Magn Reson Med 65:1724-1732, 2011. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:1724 / 1732
页数:9
相关论文
共 27 条
[1]  
Aksoy M, 2010, P 18 SCI M INT SOC M, P499
[2]  
Armstrong B, 2002, 2002 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS I-IV, PROCEEDINGS, P2938, DOI 10.1109/ROBOT.2002.1013678
[3]  
Armstrong BS, 2010, P 18 SCI M INT SOC M, P5035
[4]   Automatic correction of motion artifacts in magnetic resonance images using an entropy focus criterion [J].
Atkinson, D ;
Hill, DLG ;
Stoyle, PNR ;
Summers, PE ;
Keevil, SF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (06) :903-910
[5]   Augmented generalized SENSE reconstruction to correct for rigid body motion [J].
Bammer, Roland ;
Aksoy, Murat ;
Liu, Chunlei .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (01) :90-102
[6]   Rapid gridding reconstruction with a minimal oversampling ratio [J].
Beatty, PJ ;
Nishimura, DG ;
Pauly, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (06) :799-808
[7]   ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES [J].
EHMAN, RL ;
FELMLEE, JP .
RADIOLOGY, 1989, 173 (01) :255-263
[8]  
Gelb A., 1989, Applied Optimal Estimation
[9]   Cartesian SENSE and k-t SENSE reconstruction using commodity graphics hardware [J].
Hansen, Michael S. ;
Atkinson, David ;
Sorensen, Thomas S. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (03) :463-468
[10]   Convolution Kernel Design and Efficient Algorithm for Sampling Density Correction [J].
Johnson, Kenneth O. ;
Pipe, James G. .