Two highly sensitive microwave cavity spectrometers

被引:21
作者
Dumesh, BS
Surin, LA
机构
[1] Institute of Spectroscopy, Russian Academy of Sciences
关键词
D O I
10.1063/1.1147159
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A description is presented for two unconventional highly sensitive spectrometers operating in the decimeter-wave region (90-450 MHz) and millimeter-wave region (90-150 GHz), respectively. In both cases a resonant cavity is used as an absorption cell. Either design combines high sensitivity with high resolution and wide spectral range. The radio frequency spectrometer allows measuring of extremely low absorption quantity 3x10(-12) cm(-1) in combination with high spectral resolution of order 1 kHz with the radio frequency-microwave double resonance method. The gas cell is a coaxial cavity (Q approximate to 10(3)) with a diameter of 40 cm and length about a wavelength. The millimeter-wave spectrometer utilizes an orotron oscillator as the tunable, coherent source of radiation from 90 to 150 GHz. A gas cell is placed into a high quality (Q approximate to 10(4)) Fabry-Perot resonator of the orotron, and the absorption signal is detected by variation of the orotron electron current in the collector circuit. The sensitivity limit of (2-5)x10(-10) cm(-1) is achieved with Stark modulation as well as source frequency modulation. Some results of successful application of both designs in molecular spectroscopy are presented, and potential use of the millimeter-wave cavity spectrometer as a gas analyzer with a detection limit better than 1 ppm is briefly discussed. (C) 1996 American Institute of Physics.
引用
收藏
页码:3458 / 3464
页数:7
相关论文
共 30 条
[1]   AN AUTOMATIC MOLECULAR-BEAM MICROWAVE FOURIER-TRANSFORM SPECTROMETER [J].
ANDERSEN, U ;
DREIZLER, H ;
GRABOW, JU ;
STAHL, W .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (12) :3694-3699
[2]   FABRY-PEROT CAVITY PULSED FOURIER-TRANSFORM MICROWAVE SPECTROMETER WITH A PULSED NOZZLE PARTICLE SOURCE [J].
BALLE, TJ ;
FLYGARE, WH .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (01) :33-45
[3]  
DUMESH BS, 1989, INSTRUM EXP TECH+, V32, P634
[4]   HIGHLY SENSITIVE MILLIMETER-WAVE SPECTROMETER BASED ON AN OROTRON [J].
DUMESH, BS ;
KOSTROMIN, VP ;
RUSIN, FS ;
SURIN, LA .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (09) :873-878
[5]   HIGH-Q STARK CAVITY ABSORPTION CELL FOR MICROWAVE SPECTROMETERS [J].
DYMANUS, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1959, 30 (03) :191-195
[6]  
Gershenzon E. M., 1972, Instruments and Experimental Techniques, V15, P436
[8]  
GRABOIS MN, 1986, PRIB TEKH EKSP, V1, P133
[9]   FORBIDDEN ROTATIONAL SPECTRA OF SYMMETRIC-TOP MOLECULES - PH3 AND PD3 [J].
HELMS, DA ;
GORDY, W .
JOURNAL OF MOLECULAR SPECTROSCOPY, 1977, 66 (02) :206-218
[10]   FORBIDDEN MILLIMETER-WAVE TRANSITIONS IN ARSINE [J].
HELMS, DA ;
GORDY, W .
JOURNAL OF MOLECULAR SPECTROSCOPY, 1978, 69 (03) :473-481