Diagnostic accuracy of the real-time PCR cobas® Liat® Influenza A/B assay and the Alere i Influenza A & B NEAR isothermal nucleic acid amplification assay for the detection of influenza using adult nasopharyngeal specimens

被引:32
作者
Young, Stephen [1 ]
Illescas, Patrick [1 ]
Nicasio, Joclin [1 ]
Sickler, Joanna Jackson [2 ]
机构
[1] TriCore Reference Labs, 1001 Woodward Pl NE, Albuquerque, NM 87102 USA
[2] Roche Mol Syst, Pleasanton, CA USA
关键词
Influenza virus infection; Point-of-care; Real-time polymerase chain reaction cobas (R) Liat (R) Influenza A/B assay; Alere i Influenza A & B assay; UNITED-STATES; RAPID DETECTION; PERFORMANCE; VIRUSES; PREVENTION; SEASON;
D O I
10.1016/j.jcv.2017.07.012
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Accurate detection of influenza requires diagnostic testing; however, methods such as RADTs and central laboratory-based tests are limited by low sensitivity and time constraints, respectively. Objective: To compare the performances of the cobas (R) Liat (R) Influenza A/B and Alere (TM) i Influenza A & B point-of-care (POC) assays for detecting influenza A and B viruses using fresh nasopharyngeal specimens with the GenMark Dx (R) Respiratory Viral Panel as the reference method, a FDA cleared IVD PCR test. Study design: A total of 87 samples collected in viral transport medium from adults >= 18 years of age were re-tested on both POC assays (based on the reference PCR method, 29 were influenza A and 18 were influenza B virus positive). Results: The overall sensitivity and specificity of the cobas Influenza A/B for the detection of influenza A and B relative to reference PCR was 97.9% (95% confidence interval [CI] 88.9%, 99.6%) and 97.5% (95% CI: 87.1%, 99.6%), respectively, while the sensitivity of the Alere i Influenza A & B assay relative to the reference PCR method was 63.8% (95% CI: 49.5%, 76.0%) and the specificity was 97.5% (95% CI: 87.1%, 99.6%). The individual sensitivities and specificities of the cobas Influenza A/B assay for influenza A alone and influenza B alone were comparable to those of the reference PCR method (influenza A: sensitivity of 100% [95% CI: 88.3%, 100.0%] and specificity of 98.3% [95% CI: 90.9%, 99.7%]; influenza B: sensitivity of 94.4% [95% CI: 74.2%, 99.0%] and specificity of 100% [95% CI: 94.7%, 100.0%]). For the Alere i Influenza A & B assay, the individual specificities for influenza A and B were comparable to those of the reference PCR method (98.3% [95% CI: 90.9%, 99.7%] and 97.1% [95% CI: 90.0%, 99.2%], respectively), while the individual sensitivities were low relative to reference PCR (55.2% [95% CI: 37.5%, 71.6%] and 72.2% [95% CI: 49.1%, 87.5%], respectively). Conclusion: The cobas Influenza A/B assay demonstrated performance equivalent to laboratory-based PCR, and could replace rapid antigen tests.
引用
收藏
页码:86 / 90
页数:5
相关论文
共 31 条
  • [21] Sensitive Detection and Simultaneous Discrimination of Influenza A and B Viruses in Nasopharyngeal Swabs in a Single Assay Using Next-Generation Sequencing-Based Diagnostics
    Zhao, Jiangqin
    Liu, Jikun
    Vemula, Sai Vikram
    Lin, Corinna
    Tan, Jiying
    Ragupathy, Viswanath
    Wang, Xue
    Mbondji-Wonje, Christelle
    Ye, Zhiping
    Landry, Marie L.
    Hewlett, Indira
    PLOS ONE, 2016, 11 (09):
  • [22] Evaluation of Nanogen MGB Alert® Detection Reagents in a multiplex real-time PCR for influenza virus types A and B and respiratory syncytial virus
    Hymas, Weston C.
    Hillyard, David R.
    JOURNAL OF VIROLOGICAL METHODS, 2009, 156 (1-2) : 124 - 128
  • [23] Comparison of Becton Dickinson Directigen EZ Flu A plus B Test against the CDC Real-Time PCR Assay for Detection of 2009 Pandemic Influenza A/H1N1 Virus
    Karre, Tess
    Maguire, Hugh F.
    Butcher, David
    Graepler, Amy
    Weed, Diane
    Wilson, Michael L.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (01) : 343 - 344
  • [24] A real-time RT-PCR for detection of clade 1 and 2 H5N1 Influenza A virus using Locked Nucleic Acid (LNA) TaqMan probes
    Thanh Tran Tan
    Hana Apsari Pawestri
    Ngoc Nghiem My
    Hien Vo Minh
    Harun Syahrial
    Trung Nguyen Vu
    H Rogier van Doorn
    Heiman FL Wertheim
    Chau Nguyen Van Vinh
    Ha Do Quang
    Jeremy J Farrar
    Hien Tran Tinh
    Endang R Sedyaningsih
    Menno D de Jong
    Virology Journal, 7
  • [25] A multiplex TaqMan real-time RT-PCR assay for the simultaneous detection of H4, H6, and H10 avian influenza viruses
    Yang, Fan
    Yan, Sijing
    Zhu, Linwei
    Yao, Hangping
    Dong, Dalu
    Wu, Danna
    Wu, Nanping
    Ye, Chunsheng
    Wu, Haibo
    HELIYON, 2023, 9 (05)
  • [26] Development and application of a triplex real-time PCR assay for the simultaneous detection of avian influenza virus subtype H5, H7 and H9
    Liu, Junli
    Yao, Lu
    Zhai, Feifei
    Chen, Yuqing
    Lei, Jing
    Bi, Zhenwei
    Hu, Jianhua
    Xiao, Qian
    Song, Suquan
    Yan, Liping
    Zhou, Jiyong
    JOURNAL OF VIROLOGICAL METHODS, 2018, 252 : 49 - 56
  • [27] Multiplex loop-mediated isothermal amplification (M-LAMP) assay for the detection of influenza A/H1, A/H3 and influenza B can provide a specimen-to-result diagnosis in 40 min with single genome copy sensitivity
    Mahony, James
    Chong, Sylvia
    Bulir, David
    Ruyter, Alexandra
    Mwawasi, Ken
    Waltho, Daniel
    JOURNAL OF CLINICAL VIROLOGY, 2013, 58 (01) : 127 - 131
  • [28] Polymeric LabChip Real-Time PCR as a Point-of-Care-Potential Diagnostic Tool for Rapid Detection of Influenza A/H1N1 Virus in Human Clinical Specimens
    Song, Hyun-Ok
    Kim, Je-Hyoung
    Ryu, Ho-Sun
    Lee, Dong-Hoon
    Kim, Sun-Jin
    Kim, Deog-Joong
    Suh, In Bum
    Choi, Du Young
    In, Kwang-Ho
    Kim, Sung-Woo
    Park, Hyun
    PLOS ONE, 2012, 7 (12):
  • [29] Automated multiplex nucleic acid tests for rapid detection of SARS-CoV-2, influenza A and B infection with direct reverse-transcription quantitative PCR (dirRT-qPCR) assay in a centrifugal microfluidic platform
    Ji, Minghui
    Xia, Yun
    Loo, Jacky Fong-Chuen
    Li, Lang
    Ho, Ho-Pui
    He, Jianan
    Gu, Dayong
    RSC ADVANCES, 2020, 10 (56) : 34088 - 34098
  • [30] Evaluation of the Kaira COVID-19/Flu/RSV Detection Kit for detection of SARS-CoV-2, influenza A/B, and respiratory syncytial virus: A comparative study with the PowerChek SARS-CoV-2, influenza A&B, RSV Multiplex Real-time PCR Kit
    Kim, Tae Yeul
    Bae, Go Eun
    Kim, Ji-Youn
    Kang, Minhee
    Jang, Ja-Hyun
    Huh, Hee Jae
    Chung, Doo Ryeon
    Lee, Nam Yong
    PLOS ONE, 2022, 17 (12):