摘要
Background: Accurate detection of influenza requires diagnostic testing; however, methods such as RADTs and central laboratory-based tests are limited by low sensitivity and time constraints, respectively. Objective: To compare the performances of the cobas (R) Liat (R) Influenza A/B and Alere (TM) i Influenza A & B point-of-care (POC) assays for detecting influenza A and B viruses using fresh nasopharyngeal specimens with the GenMark Dx (R) Respiratory Viral Panel as the reference method, a FDA cleared IVD PCR test. Study design: A total of 87 samples collected in viral transport medium from adults >= 18 years of age were re-tested on both POC assays (based on the reference PCR method, 29 were influenza A and 18 were influenza B virus positive). Results: The overall sensitivity and specificity of the cobas Influenza A/B for the detection of influenza A and B relative to reference PCR was 97.9% (95% confidence interval [CI] 88.9%, 99.6%) and 97.5% (95% CI: 87.1%, 99.6%), respectively, while the sensitivity of the Alere i Influenza A & B assay relative to the reference PCR method was 63.8% (95% CI: 49.5%, 76.0%) and the specificity was 97.5% (95% CI: 87.1%, 99.6%). The individual sensitivities and specificities of the cobas Influenza A/B assay for influenza A alone and influenza B alone were comparable to those of the reference PCR method (influenza A: sensitivity of 100% [95% CI: 88.3%, 100.0%] and specificity of 98.3% [95% CI: 90.9%, 99.7%]; influenza B: sensitivity of 94.4% [95% CI: 74.2%, 99.0%] and specificity of 100% [95% CI: 94.7%, 100.0%]). For the Alere i Influenza A & B assay, the individual specificities for influenza A and B were comparable to those of the reference PCR method (98.3% [95% CI: 90.9%, 99.7%] and 97.1% [95% CI: 90.0%, 99.2%], respectively), while the individual sensitivities were low relative to reference PCR (55.2% [95% CI: 37.5%, 71.6%] and 72.2% [95% CI: 49.1%, 87.5%], respectively). Conclusion: The cobas Influenza A/B assay demonstrated performance equivalent to laboratory-based PCR, and could replace rapid antigen tests.
引用
收藏
页码:86 / 90
页数:5
相关论文
共 31 条
  • [1] Evaluation of a rapid isothermal nucleic acid amplification kit, Alere™ i Influenza A&B, for the detection of avian influenza viruses
    Bazarragchaa, Enkhbold
    Okamatsu, Masatoshi
    Ulaankhuu, Ankhanbaatar
    Twabela, Augustin Tshibwabwa
    Matsuno, Keita
    Kida, Hiroshi
    Sakoda, Yoshihiro
    JOURNAL OF VIROLOGICAL METHODS, 2019, 265 : 121 - 125
  • [2] Comparing the Cobas Liat Influenza A/B and respiratory syncytial virus assay with multiplex nucleic acid testing
    Gosert, Rainer
    Naegele, Klaudia
    Hirsch, Hans H.
    JOURNAL OF MEDICAL VIROLOGY, 2019, 91 (04) : 582 - 587
  • [3] Analytical performance of the Alere™ i Influenza A&B assay for the rapid detection of influenza viruses
    Riazzo, Cristina
    Perez-Ruiz, Mercedes
    Sanbonmatsu-Gamez, Sara
    Pedrosa-Corral, Irene
    Gutierrez-Fernandez, Jose
    Navarro-Mari, Jose-Maria
    ENFERMEDADES INFECCIOSAS Y MICROBIOLOGIA CLINICA, 2017, 35 (07): : 438 - 440
  • [4] A Universal Influenza A and B Duplex Real-time RT-PCR Assay
    Lee, Hong Kai
    Loh, Tze Ping
    Lee, Chun Kiat
    Tang, Julian Wei-Tze
    Chiu, Lily
    Koay, Evelyn Siew-Chuan
    JOURNAL OF MEDICAL VIROLOGY, 2012, 84 (10) : 1646 - 1651
  • [5] Evaluation of the molecular Xpert Xpress Flu/RSV assay vs. Alere i Influenza A & B assay for rapid detection of influenza viruses
    Chen, J. H.
    Lam, H. Y.
    Yip, C. C.
    Cheng, V. C.
    Chan, J. F.
    Leung, T. H.
    Sridhar, S.
    Chan, K. H.
    Tang, B. S.
    Yuen, K. Y.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2018, 90 (03) : 177 - 180
  • [6] Comparison of the ID Now Influenza A & B 2, Cobas Influenza A/B, and Xpert Xpress Flu Point-of-Care Nucleic Acid Amplification Tests for Influenza A/B Virus Detection in Children
    Kanwar, Neena
    Michael, Jeffrey
    Doran, Kathryn
    Montgomery, Emily
    Selvarangan, Rangaraj
    JOURNAL OF CLINICAL MICROBIOLOGY, 2020, 58 (03)
  • [7] Prospective evaluation of the Alere i Influenza A&B nucleic acid amplification versus Xpert Flu/RSV
    Van, J. C. Nguyen
    Camelena, F.
    Dahoun, M.
    Pilmis, B.
    Mizrahi, A.
    Lourtet, J.
    Behillil, S.
    Enouf, V.
    Le Monnier, A.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2016, 85 (01) : 19 - 22
  • [8] Multicenter evaluation of the Alere™ i influenza A&B assay using respiratory specimens collected in viral transport media
    Hassan, Ferdaus
    Crawford, Jordan
    Bonner, Aleta B.
    Ledeboer, Nathan A.
    Selvarangan, Rangaraj
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2018, 92 (04) : 294 - 298
  • [9] Diagnostic accuracy and cost analysis of the Alere™ i Influenza A&B near-patient test using throat swabs
    Davis, S.
    Allen, A. J.
    O'Leary, R.
    Power, M.
    Price, D. A.
    Simpson, A. J.
    Tunbridge, A.
    Vale, L.
    Whiteside, M.
    Evans, C.
    Raza, M.
    JOURNAL OF HOSPITAL INFECTION, 2017, 97 (03) : 301 - 309
  • [10] Lab-in-a-tube: Real-time molecular point-of-care diagnostics for influenza A and B using the cobas® Liat® system
    Melchers, Willem J. G.
    Kuijpers, Judith
    Sickler, Joanna Jackson
    Rahamat-Langendoen, Janette
    JOURNAL OF MEDICAL VIROLOGY, 2017, 89 (08) : 1382 - 1386