Minimum and optimum light requirements for laryngoscopy in paediatric anaesthesia: a manikin study

被引:12
作者
Malan, C. A. [1 ]
Scholz, A.
Wilkes, A. R.
Hampson, M. A.
Hall, J. E.
机构
[1] Cardiff Univ, Coll Med, Dept Anaesthet & Intens Care Med, Cardiff CF14 4XN, Wales
[2] Univ Wales Hosp, Dept Anaesthet & Intens Care Med, Cardiff CF14 4XW, Wales
关键词
D O I
10.1111/j.1365-2044.2007.05259.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Direct laryngoscopy is performed regularly in paediatric anaesthetic practice for the purpose of intubating a patient's trachea. A minimum illumination of 700 lux at a distance of 20 mm has been suggested in a draft standard for laryngoscopes from the International Organization for Standardization. We investigated the minimum and optimum illumination required by anaesthetists during laryngoscopy. Fifty anaesthetists were recruited to perform laryngoscopy on a child manikin with a selection of laryngoscope blades and bulbs attached to a variable voltage supply. There was a large variation in the illumination range of different bulb and blade combinations. We demonstrated a large interindividual variability between anaesthetists for illumination requirements during laryngoscopy. Anaesthetists can see the larynx at very low light levels, but optimum illumination was significantly greater than minimum illumination. Two of the six combinations could produce more than 700 lux. In all, 78% and 54% of anaesthetists considered 700 lux too bright with a halogen and xenon bulb, respectively. Our study suggests that the proposed standard of 700 lux may possibly be too bright.
引用
收藏
页码:65 / 70
页数:6
相关论文
共 12 条
  • [1] An assessment of the luminance and light field characteristics of used direct laryngoscopes
    Crosby, E
    Cleland, M
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1999, 46 (08): : 792 - 796
  • [2] Flexibility and light emission of disposable paediatric Miller 1 laryngoscope blades
    Goodwin, N.
    Wilkes, A. R.
    Hall, J. E.
    [J]. ANAESTHESIA, 2006, 61 (08) : 792 - 799
  • [3] *HLTH SAF EX, 1998, LIGHT WORK HSG 38 19
  • [4] The reliability of laryngoscope lights
    Howes, BW
    [J]. ANAESTHESIA, 2006, 61 (05) : 488 - 491
  • [5] *ISO, 2005, 121SC ISO TC
  • [6] THE MACINTOSH LARYNGOSCOPE - A HISTORICAL NOTE ON ITS CLINICAL AND COMMERCIAL DEVELOPMENT
    JEPHCOTT, A
    [J]. ANAESTHESIA, 1984, 39 (05) : 474 - 479
  • [7] The Cardiff paediatric laryngoscope and its 'non-conforming' siblings
    Jones, RM
    Jones, PL
    Gildersleve, CD
    Hall, JE
    Manners, N
    [J]. ANAESTHESIA, 2005, 60 (07) : 721 - 722
  • [8] TRACHEAL INTUBATION AND LARYNGOSCOPE DESIGN
    MCINTYRE, JWR
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1993, 40 (03): : 193 - 196
  • [9] LARYNGOSCOPE DESIGN AND THE DIFFICULT ADULT TRACHEAL INTUBATION
    MCINTYRE, JWR
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1989, 36 (01): : 94 - 98
  • [10] Minimum and optimum light output of Macintosh size 3 laryngoscopy blades: a manikin study
    Scholz, A.
    Farnum, N.
    Wilkes, A. R.
    Hampson, M. A.
    Hall, J. E.
    [J]. ANAESTHESIA, 2007, 62 (02) : 163 - 168