Comparison of estimation algorithms in single-molecule localization

被引:2
作者
Abraham, Anish V. [1 ,2 ]
Ram, Sripad [2 ]
Chao, Jerry [1 ,2 ]
SallyWard, E. [2 ]
Ober, Raimund J. [1 ,2 ]
机构
[1] Univ Texas Dallas, Dept Elect Engn, Richardson, TX 75083 USA
[2] Univ Texas Southwestern Med Ctr Dallas, Dept Immunol, Dallas, TX 75390 USA
来源
THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XVII | 2010年 / 7570卷
基金
美国国家卫生研究院;
关键词
localization; single molecule; tracking; Cramer-Rao lower bound; RESTORATION ALGORITHMS; NANOMETER LOCALIZATION;
D O I
10.1117/12.842178
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Different techniques have been advocated for estimating single molecule locations from microscopy images. The question arises as to which technique produces the most accurate results. Various factors, e. g. the stochastic nature of the photon emission/detection process, extraneous additive noise, pixelation, etc., result in the estimated single molecule location deviating from its true location. Here, we review the results presented by [Abraham et. al, Optics Express, 2009, 23352-23373], where the performance of the maximum likelihood and nonlinear least squares estimators for estimating single molecule locations are compared. Our results show that on average both estimators recover the true single molecule location in all scenarios. Comparing the standard deviations of the estimates, we find that in the absence of noise and modeling inaccuracies, the maximum likelihood estimator is more accurate than the non-linear least squares estimator, and attains the best achievable accuracy for the sets of experimental and imaging conditions tested. In the presence of noise and modeling inaccuracies, the maximum likelihood estimator produces results with consistent accuracy across various model mismatches and misspecifications. At high noise levels, neither estimator has an accuracy advantage over the other. We also present new results regarding the performance of the maximum likelihood estimator with respect to the objective function used to fit data containing both additive Gaussian and Poisson noise. Comparisons were also carried out between two localization accuracy measures derived previously. User-friendly software packages were developed for single molecule location estimation (EstimationTool) and localization accuracy calculations (FandPLimitTool).
引用
收藏
页数:7
相关论文
共 19 条
[11]  
Ober RJ, 2004, BIOPHYS J, V86, p514A
[12]   Nanometre localization of single ReAsH molecules [J].
Park, H ;
Hanson, GT ;
Duff, SR ;
Selvin, PR .
JOURNAL OF MICROSCOPY, 2004, 216 :199-205
[13]   A stochastic analysis of performance limits for optical microscopes [J].
Ram, S ;
Ward, ES ;
Ober, RJ .
MULTIDIMENSIONAL SYSTEMS AND SIGNAL PROCESSING, 2006, 17 (01) :27-57
[14]   Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM) [J].
Rust, Michael J. ;
Bates, Mark ;
Zhuang, Xiaowei .
NATURE METHODS, 2006, 3 (10) :793-795
[15]   Precise nanometer localization analysis for individual fluorescent probes [J].
Thompson, RE ;
Larson, DR ;
Webb, WW .
BIOPHYSICAL JOURNAL, 2002, 82 (05) :2775-2783
[16]  
Torok P., 2003, Optical Imaging and Microscopy, Vfirst
[17]   Efficient superresolution restoration algorithms using maximum a posteriori estimations with application to fluorescence microscopy [J].
Verveer, PJ ;
Jovin, TM .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (08) :1696-1706
[18]   Single-molecule approach to molecular biology in living bacterial cells [J].
Xie, X. Sunney ;
Choi, Paul J. ;
Li, Gene-Wei ;
Lee, Nam Ki ;
Lia, Giuseppe .
ANNUAL REVIEW OF BIOPHYSICS, 2008, 37 :417-444
[19]   Myosin V walks hand-over-hand: Single fluorophore imaging with 1.5-nm localization [J].
Yildiz, A ;
Forkey, JN ;
McKinney, SA ;
Ha, T ;
Goldman, YE ;
Selvin, PR .
SCIENCE, 2003, 300 (5628) :2061-2065