In vivo demonstration of magnetic guidewire steerability in a MRI system with additional gradient coils

被引:19
作者
Lalande, Viviane [1 ]
Gosselin, Frederick P. [2 ]
Vonthron, Manuel [1 ]
Conan, Benjamin [1 ]
Tremblay, Charles [1 ]
Beaudoin, Gilles [3 ]
Soulez, Gilles [3 ]
Martel, Sylvain [1 ]
机构
[1] Ecole Polytech Montreal, Dept Genie Informat & Genie Logiciel, NanoRobot Lab, Montreal, PQ H3T 1J4, Canada
[2] Ecole Polytech Montreal, Dept Genie Mecan, Montreal, PQ H3T 1J4, Canada
[3] CHUM Hop Notre Dame, Dept Radiol, Montreal, PQ H2L 4M1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
guidewire; MRI; steering; in vivo; magnetic; magnetic gradient; NAVIGATION; CATHETER;
D O I
10.1118/1.4906194
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the ability to control the steering of a modified guidewire actuated by the magnetic force of a magnetic resonance imaging system with additional gradient coils for selective arterial catheterization in rabbits. Methods: Selective catheterizations of the right renal artery, left renal artery, superior mesenteric artery, and iliac artery were performed on two rabbits. A 3D magnetic force was applied onto a magnetic bead placed at the tip of a guidewire. The ability of the guidewire to advance in the aorta without entering the side branches when the magnetic force was not applied was also evaluated. Steering of the guidewire was combined with a dedicated tracking system and its position was registered on the 3D model of a magnetic resonance angiography (MRA). Results: The magnetic catheterization of the renal arteries was successful and showed reproducibility. Superior mesenteric artery and iliac artery showed that the catheterization was feasible. These two arteries were difficult to visualize on MRA, making catheterization and setting the direction of the force more difficult. There was no inadvertent catheterization of side vessels when the guidewire was advanced with magnetic steering despite the hook shape at the tip of the guidewire caused by the alignment of the bead anisotropy with the permanent magnetic field. Conclusions: This first evaluation of selective catheterization of aortic branches with a magnetic guidewire provided a successful steering in the less angled side branches and this modified guidewire was advanced in the aorta without inadvertent selective catheterization when manipulated without magnetic actuation. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:969 / 976
页数:8
相关论文
共 21 条
  • [11] Mathieu J.B., 2008, Journal of Vascular and Interventional Radiology, V19, pS31
  • [12] Method of propulsion of a ferromagnetic core in the cardiovascular system through magnetic gradients generated by an MRI system
    Mathieu, JB
    Beaudoin, G
    Martel, S
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (02) : 292 - 299
  • [13] Moore W., 2011, Endovascular Surgery, V4th
  • [14] MRI visualization of a single 15 μm navigable imaging agent and future microrobot
    Olamaei, N.
    Cheriet, F.
    Beaudoin, G.
    Martel, S.
    [J]. 2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 4355 - 4358
  • [15] Novel robotic catheter remote control system: Feasibility and safety of transseptal puncture and endocardial catheter navigation
    Saliba, Walid
    Cummings, Jennifer E.
    Oh, Seil
    Zhang, Youhua
    Mazgalev, Todor N.
    Schweikert, Robert A.
    Burkhardt, J. David
    Natale, Andrea
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (10) : 1102 - 1105
  • [16] Vascular guide wire navigation with a magnetic guidance system: Experimental results in a phantom
    Schiemann, M
    Killmann, R
    Kleen, M
    Abolmaali, N
    Finney, J
    Vogl, TJ
    [J]. RADIOLOGY, 2004, 232 (02) : 475 - 481
  • [17] Schneider P.A., 2003, Endovascular skills : guidewire and catheter skills for endovascular surgery
  • [18] Magnetically-assisted remote control (MARC) steering of endovascular catheters for interventional MRI: A model for deflection and design implications
    Settecase, Fabio
    Sussman, Marshall S.
    Wilson, Mark W.
    Hetts, Steven
    Arenson, Ronald L.
    Malba, Vincent
    Bernhardt, Anthony F.
    Kucharczyk, Walter
    Roberts, Timothy P. L.
    [J]. MEDICAL PHYSICS, 2007, 34 (08) : 3135 - 3142
  • [19] RF Heating of MRI-Assisted Catheter Steering Coils for Interventional MRI
    Settecase, Fabio
    Hetts, Steven W.
    Martin, Alastair J.
    Roberts, Timothy P. L.
    Bernhardt, Anthony F.
    Evans, Lee
    Malba, Vincent
    Saeed, Maythem
    Arenson, Ronald L.
    Kucharzyk, Walter
    Wilson, Mark W.
    [J]. ACADEMIC RADIOLOGY, 2011, 18 (03) : 277 - 285
  • [20] Vonthron M, 2011, IEEE ENG MED BIO, P5392, DOI 10.1109/IEMBS.2011.6091333