A preliminary dosimetric characterization of chemical vapor deposition diamond detector prototypes in photon and electron radiotherapy beams

被引:28
作者
Fidanzio, A
Azario, L
Kalish, R
Avigal, Y
Conte, G
Ascarelli, P
机构
[1] Univ Cattolica Sacro Cuore, Ist Fis, I-00168 Rome, Italy
[2] Technion Israel Inst Technol, Dept Phys, IL-32000 Haifa, Israel
[3] Technion Israel Inst Technol, Inst Solid State, IL-32000 Haifa, Israel
[4] INFM, I-00146 Rome, Italy
[5] Univ Rome Roma Tre, Dept Elect Engn, I-00146 Rome, Italy
[6] CNR, Inst Inorgan Methodol & Plasma, IMIP, I-00016 Monterotondo, Rome, Italy
关键词
CVD; synthetic diamond; radiotherapy dosimetry;
D O I
10.1118/1.1851887
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Three radiation detectors based on polycrystalline diamond films with different thickness and resistivity, obtained by microwave chemical vapor deposition, were tested to assess their suitability for relative dosimetry of photon and electron beams supplied by clinical linear accelerators. All samples showed a linear response as a function of the absorbed dose. The sensitivity per unit of detector sensitive volume spanned between 7 and 43 nC Gy(-1) mm(-3) with an applied electric field of 40 kV/cm. The dose rate dependence was evaluated following the Fowler theory and A coefficient values between 0.95 and 1.00 were found for the three samples when polarized at 40 kV/cm. Percentage depth dose curves, output factors, and normalized dose profiles were determined for 6 and 10 MV photon beams and for 6 and 15 MeV electron beams. The results obtained with the diamond detectors were in good agreement with those obtained by reference detector measurements [all the data were within the experimental uncertainty of 1% (1sigma)]. (C) 2005 American Association Physicists in Medicine.
引用
收藏
页码:389 / 395
页数:7
相关论文
共 24 条
  • [11] DIAMOND RADIATION DETECTORS
    KANIA, DR
    LANDSTRASS, MI
    PLANO, MA
    PAN, LS
    HAN, S
    [J]. DIAMOND AND RELATED MATERIALS, 1993, 2 (5-7) : 1012 - 1019
  • [12] SYNTHETIC DIAMONDS AS IONIZATION-CHAMBER RADIATION DETECTORS IN BIOLOGICAL ENVIRONMENTS
    KEDDY, RJ
    NAM, TL
    BURNS, RC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06) : 751 - 759
  • [13] KONOROVA EA, 1971, SOV PHYS SEMICOND+, V4, P1600
  • [14] DIAMOND DOSIMETER FOR X-RAY AND GAMMA-RADIATION
    KOZLOV, SF
    KONOROVA, EA
    KUZNETSOV, YA
    SALIKOV, YA
    REDKO, VI
    GRINBERG, VR
    MEILMAN, ML
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) : 235 - 237
  • [15] PREPARATION AND CHARACTERISTICS OF NATURAL DIAMOND NUCLEAR RADIATION DETECTORS
    KOZLOV, SF
    STUCK, R
    HAGEALI, M
    SIFFERT, P
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1975, NS22 (01) : 160 - 170
  • [16] Energy and dose rate dependence of a diamond detector in the dosimetry of 4-25 MV photon beams
    Laub, WU
    Kaulich, TW
    Nusslin, F
    [J]. MEDICAL PHYSICS, 1997, 24 (04) : 535 - 536
  • [17] Growth and characterisation of CVD diamond wires for x-ray detection
    Manfredotti, C
    Fizzotti, F
    LoGiudice, A
    Mucera, G
    Polesello, P
    Vittone, E
    Mankelevich, YA
    Suetin, NV
    [J]. DIAMOND AND RELATED MATERIALS, 1997, 6 (08) : 1051 - 1056
  • [18] The value of the LA48 linear ion chamber array for characterization of intensity-modulated beams
    Martens, C
    De Wagter, C
    De Neve, W
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04) : 1131 - 1148
  • [19] Piermattei A, 1998, PHYS MEDICA, V14, P9
  • [20] EVALUATION OF DIAMOND RADIATION DOSIMETERS
    PLANSKOY, B