A preliminary dosimetric characterization of chemical vapor deposition diamond detector prototypes in photon and electron radiotherapy beams

被引:28
作者
Fidanzio, A
Azario, L
Kalish, R
Avigal, Y
Conte, G
Ascarelli, P
机构
[1] Univ Cattolica Sacro Cuore, Ist Fis, I-00168 Rome, Italy
[2] Technion Israel Inst Technol, Dept Phys, IL-32000 Haifa, Israel
[3] Technion Israel Inst Technol, Inst Solid State, IL-32000 Haifa, Israel
[4] INFM, I-00146 Rome, Italy
[5] Univ Rome Roma Tre, Dept Elect Engn, I-00146 Rome, Italy
[6] CNR, Inst Inorgan Methodol & Plasma, IMIP, I-00016 Monterotondo, Rome, Italy
关键词
CVD; synthetic diamond; radiotherapy dosimetry;
D O I
10.1118/1.1851887
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Three radiation detectors based on polycrystalline diamond films with different thickness and resistivity, obtained by microwave chemical vapor deposition, were tested to assess their suitability for relative dosimetry of photon and electron beams supplied by clinical linear accelerators. All samples showed a linear response as a function of the absorbed dose. The sensitivity per unit of detector sensitive volume spanned between 7 and 43 nC Gy(-1) mm(-3) with an applied electric field of 40 kV/cm. The dose rate dependence was evaluated following the Fowler theory and A coefficient values between 0.95 and 1.00 were found for the three samples when polarized at 40 kV/cm. Percentage depth dose curves, output factors, and normalized dose profiles were determined for 6 and 10 MV photon beams and for 6 and 15 MeV electron beams. The results obtained with the diamond detectors were in good agreement with those obtained by reference detector measurements [all the data were within the experimental uncertainty of 1% (1sigma)]. (C) 2005 American Association Physicists in Medicine.
引用
收藏
页码:389 / 395
页数:7
相关论文
共 24 条
  • [1] Recent results from the RD42 Diamond Detector Collaboration
    Bauer, C
    Baumann, I
    Colledani, C
    Conway, J
    Delpierre, P
    Djama, F
    Dulinski, W
    Fallou, A
    Gan, KK
    Gilmore, RS
    Grigoriev, E
    Hallewell, G
    Han, S
    Hessing, T
    Honscheid, K
    Hrubec, J
    Husson, D
    James, R
    Kagan, H
    Kania, D
    Kass, R
    Knopfle, KT
    Krammer, M
    Llewellyn, TJ
    Manfredi, PF
    Meier, D
    Pan, LS
    Pernegger, H
    Pernicka, M
    Re, V
    Roe, S
    Roff, D
    Rudge, A
    Schaeffer, M
    Schieber, M
    Schnetzer, S
    Somalwar, S
    Speziali, V
    Stone, R
    Tapper, RJ
    Tesarek, R
    Trischuk, W
    Turchetta, R
    Thomson, GB
    Wagner, R
    Weilhammer, P
    White, C
    Ziock, HJ
    Zoeller, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 383 (01) : 64 - 74
  • [2] Development of diamond films for particle detector applications
    Behnke, T
    Oh, A
    Wagner, A
    Zeuner, W
    Bluhm, A
    Klages, CP
    Paul, M
    Schafer, L
    [J]. DIAMOND AND RELATED MATERIALS, 1998, 7 (10) : 1553 - 1557
  • [3] Characterisation of CVD diamond dosimeters in on-line configuration
    Bruzzi, M
    Bucciolini, M
    Cirrone, GAP
    Cuttone, G
    Mazzocchi, S
    Pirollo, S
    Sciortino, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 454 (01) : 142 - 146
  • [4] PTW-diamond detector: Dose rate and particle type dependence
    Fidanzio, A
    Azario, L
    Miceli, R
    Russo, A
    Piermattei, A
    [J]. MEDICAL PHYSICS, 2000, 27 (11) : 2589 - 2593
  • [5] Fidanzio A, 2002, PHYS MEDICA, V18, P129
  • [6] FOWLER JF, 1966, RAD DOSIMETRY, P291
  • [7] EVALUATION OF A PTW DIAMOND DETECTOR FOR ELECTRON-BEAM MEASUREMENTS
    HEYDARIAN, M
    HOBAN, PW
    BECKHAM, WA
    BORCHARDT, IM
    BEDDOE, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (08) : 1035 - 1042
  • [8] DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM
    HOBAN, PW
    HEYDARIAN, M
    BECKHAM, WA
    BEDDOE, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) : 1219 - 1229
  • [9] Hubbell J. H., 1995, 5632 NISTIR
  • [10] *IAEA TRS NO 398, 2000, ABS DOS DET EXT BEAM