Determination of rare-earth elements in Arima spring water by ICP-MS with ion-exchange separation

被引:0
|
作者
Tsuji, H
Fujiwara, S
Chayama, K
Teranishi, K
Isomura, K
机构
[1] Konan Univ, Higashinada Ku, Kobe, Hyogo 6588501, Japan
[2] Hyogo Prefectural Inst Publ Hlth, Hyogo Ku, Kobe, Hyogo 6520032, Japan
关键词
rare-earth elements; chelating resin; preconcentration; Arima spring water; ICP-MS;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
ICP-MS was applied to the determination of rare-earth elements(REE) in Arima spring water. For the preconcentration of REE from matrix elements, an ion-exchange method using a chelating resin having an iminodiacetic acid functional group was applied. The analytical procedure was as follows: the dissolved iron in the sample, which forms hydroxide precipitates by pH adjustment and intefer the analysis of REE, was oxidized by HNO3 and then removed by an anion-exchange resin as a chloride complex under 4 M HCL acidity. The solution was adjusted to have a pH value of 5.5 with an acetate buffer and an ammonia solution, and was then stirred for 2 h after the addition of a chelating resin. The resin was collected by a glass filter and washed with a magnesium chloride solution, an ammonium acetate buffer solution and water to remove any alkali earth elements (Ca, Mg, Ba). REE adsorbed on the resin was eluted with 5 ml of 2 M HNO3. The Cd as an internal standard element was added to the eluent and diluted to 20 ml for determination by ICP-MS. The recovery of each REE was more than 90%. The total REE concentration and Eu anomalies of Na-Cl-Fe high-temperature spring water differed from Na-Cl-HCO3 low-temperature spring water in Arima.
引用
收藏
页码:699 / 705
页数:7
相关论文
共 50 条
  • [1] Determination of rare-earth elements in environmental water using chelating resin disk separation/ICP-MS
    Takaku, Y
    Kudo, Y
    Kimura, J
    Hayashi, T
    Ota, I
    Hasegawa, H
    Ueda, S
    BUNSEKI KAGAKU, 2002, 51 (07) : 539 - 544
  • [2] Accurate Measurement of Rare Earth Elements by ICP-MS after Ion-Exchange Separation: Application to Ultra-Depleted Samples
    Ulrich, Marc
    Bureau, Sarah
    Chauvel, Catherine
    Picard, Christian
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2012, 36 (01) : 7 - 20
  • [3] Determination of rare earth elements in petroleum by ICP-MS
    Takeda, K
    Arikawa, Y
    BUNSEKI KAGAKU, 2005, 54 (10) : 939 - 943
  • [4] SEPARATION OF TUNGSTEN FROM RARE-EARTH ELEMENTS BY ION-EXCHANGE CHROMATOGRAPHY
    SHIPUNOVA, NT
    CHECHUSHKOVA, NE
    JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 48 (07) : 808 - 810
  • [5] Determination of rare earth elements in sixteen silicate reference samples by ICP-MS after Tm addition and ion exchange separation
    Barrat, JA
    Keller, F
    Amosse, J
    Taylor, RN
    Nesbitt, RW
    Hirata, T
    GEOSTANDARDS NEWSLETTER-THE JOURNAL OF GEOSTANDARDS AND GEOANALYSIS, 1996, 20 (01): : 133 - 139
  • [6] Determination of trace arsenic in silicon wafers by ion-exchange separation/ICP-MS
    Tachibe, T
    Kozuka, S
    Yamada, Y
    Tomita, M
    Takenaka, M
    BUNSEKI KAGAKU, 2003, 52 (12) : 1181 - 1185
  • [7] Determination of uranium, thorium and rare-earth elements in zircon samples using ICP-MS
    A. K. Singh
    V. Padmasubashini
    Leela Gopal
    Journal of Radioanalytical and Nuclear Chemistry, 2012, 294 : 19 - 25
  • [8] Determination of uranium, thorium and rare-earth elements in zircon samples using ICP-MS
    Singh, A. K.
    Padmasubashini, V.
    Gopal, Leela
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2012, 294 (01) : 19 - 25
  • [9] Method validation for determination of some rare-earth elements in rice using ICP-MS
    Kukusamude, C.
    Kongsri, S.
    INTERNATIONAL NUCLEAR SCIENCE AND TECHNOLOGY CONFERENCE, 2019, 1285
  • [10] Determination of rare earth elements in crude oil by ICP-MS
    Luo, Chuan
    Dai, Shujun
    Ye, Jiamei
    ADVANCES IN ENERGY SCIENCE AND EQUIPMENT ENGINEERING, 2015, : 211 - 214