Fresnel-Based Beamforming for Low-Cost Portable Ultrasound

被引:8
|
作者
Man Minh Nguyen [1 ]
Mung, Jay [1 ]
Yen, Jesse T. [1 ]
机构
[1] Univ So Calif, Dept Biomed Engn, Los Angeles, CA 90089 USA
关键词
D O I
10.1109/TUFFC.2011.1778
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In this paper, we propose a modified electronic Fresnel-based beamforming method for low-cost portable ultrasound systems. This method uses a unique combination of analog and digital beamforming methods. Two versions of Fresnel beamforming are presented in this paper: 4-phase (4 different time delays or phase shifts) and 8-phase (8 different time delays or phase shifts). The advantage of this method is that a system with 4 to 8 transmit channels and 2 receive channels with a network of switches can be used to focus an array with 64 to 128 elements. The simulation and experimental results show that Fresnel beamforming image quality is comparable to traditional delay-and-sum (DAS) beamforming in terms of spatial resolution and contrast-to-noise ratio (CNR) under certain system parameters. With an f-number of 2 and 50% signal bandwidth, the experimental lateral beamwidths are 0.54, 0.67, and 0.66 mm and the axial pulse lengths are 0.50, 0.51, and 0.50 mm for DAS, 8-phase, and 4-phase Fresnel beamforming, respectively. The experimental CNRs are 4.66, 4.42, and 3.98, respectively. These experimental results are in good agreement with simulation results.
引用
收藏
页码:112 / 121
页数:10
相关论文
共 50 条
  • [21] A PORTABLE AIR QUALITY MONITOR BASED ON LOW-COST SENSORS
    Tasic, Visa M.
    Bozilov, Aca, V
    Misic, Nikola Z.
    Topalovic, Dusan B.
    Randjelovic, Ivana S.
    Mirkov, Nikola S.
    Kamenovic, Vladan V.
    THERMAL SCIENCE, 2023, 27 (3B): : 2309 - 2319
  • [22] A portable low-cost fluorimeter based on LEDs and a smart phone
    Khoshmaram, Leila
    Mohammadi, Maryam
    Babadi, Adel Nazemi
    MICROCHEMICAL JOURNAL, 2021, 171
  • [23] Direct sampled I/Q beamforming for compact and very low-cost ultrasound Imaging
    Ranganathan, K
    Santy, MK
    Blalock, TN
    Hossack, JA
    Walker, WF
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (09) : 1082 - 1094
  • [24] Performance Comparison of Fresnel-based Concentrator Arrays
    Mohedano, Ruben
    Cvetkovic, Aleksandra
    Benitez, Pablo
    Zamora, Pablo
    Minano, Juan C.
    Chaves, Julio
    Hernandez, Maikel
    Buljan, Marina
    Vilaplana, Juan
    NONIMAGING OPTICS: EFFICIENT DESIGN FOR ILLUMINATION AND SOLAR CONCENTRATION VII, 2010, 7785
  • [25] High performance Fresnel-based photovoltaic concentrator
    Benitez, Pablo
    Minano, Juan C.
    Zamora, Pablo
    Mohedano, Ruben
    Cvetkovic, Aleksandra
    Buljan, Marina
    Chaves, Julio
    Hernandez, Maikel
    OPTICS EXPRESS, 2010, 18 (09): : A25 - A40
  • [26] Evaluation of a Fractional Filter-based Receive Beamforming Method for Low-cost Ultrasound Color Doppler Imaging
    Yang, Hana
    Kang, Jeeun
    Chang, Jin Ho
    Yoo, Yangmo
    MEDICAL IMAGING 2012: ULTRASONIC IMAGING, TOMOGRAPHY, AND THERAPY, 2012, 8320
  • [27] Low-Cost Beamforming Employing Reconfigurable Antennas
    Guo, Y. Jay
    Qin, Pei-Yuan
    Ding, Can
    2014 INTERNATIONAL WORKSHOP ON ANTENNA TECHNOLOGY: "SMALL ANTENNAS, NOVEL EM STRUCTURES AND MATERIALS, AND APPLICATIONS" (IWAT), 2014, : 155 - 158
  • [28] High performance Fresnel-based photovoltaic concentrator
    Zamora, Pablo
    Benitez, Pablo
    Minano, Juan C.
    Mohedano, Ruben
    Cvetkovic, Aleksandra
    Buljan, Marina
    Chaves, Julio
    NONIMAGING OPTICS: EFFICIENT DESIGN FOR ILLUMINATION AND SOLAR CONCENTRATION VII, 2010, 7785
  • [29] A portable, low-cost, highly integrated, 3D medical ultrasound system
    Fuller, MI
    Blalock, TN
    Hossack, JA
    Walker, WF
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 38 - 41
  • [30] The takeaway Frenzel goggles A Fresnel-based device
    Strupp, Michael
    Fischer, Carolin
    Hanss, Lena
    Bayer, Otmar
    NEUROLOGY, 2014, 83 (14) : 1241 - 1245