Minimizing Stress Concentrations in the Femoral Heads of Hip Joint Prostheses: Effect of Borehole Shapes

被引:1
作者
Uddin, Mohammad Sharif [1 ]
Zhang, L. C. [1 ]
机构
[1] Univ New S Wales, Sch Mech & Mfg Engn, Sydney, NSW 2052, Australia
来源
ADVANCES IN MATERIALS PROCESSING IX | 2010年 / 443卷
关键词
Femoral head; Hip joint prosthesis; Borehole shape; Finite element; Stress distribution; Contact length; Interface friction; RELIABILITY;
D O I
10.4028/www.scientific.net/KEM.443.736
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper presents a stress analysis of the ceramic femoral heads of hip joint prostheses with different borehole shapes to evaluate their mechanical reliability in terms of stress concentration. Under the ideal loading conditions used for ceramic rupture tests specified by the ISO 7206-5 standard, a finite element (FE) modeling is performed to determine the tensile and hoop stress distributions in the ceramic femoral heads. Two borehole shapes that are currently used in the manufacturing industry for hip joint prostheses, namely the flat bottom and keyhole, were first studied. Two new borehole shapes, dome arc and dome ellipse, were then introduced by the authors in the paper to minimize the stress concentration. It was found that while the currently used borehole shapes lead to high tensile notch stresses at their critical corners causing possible fracture failure of ceramic heads, the authors' borehole designs can improve the mechanical reliability significantly. In addition, the effects of taper-bore contact length and their interface friction are investigated and discussed.
引用
收藏
页码:736 / 741
页数:6
相关论文
共 11 条
  • [1] AFFOLTER C, 2009, J ENG MED, V223, P237
  • [2] BERGMANN G, 1998, CD HIP 98 MIT BEL HU
  • [3] THE SAFETY OF CERAMIC BALLS ON METAL STEMS IN HIP-ARTHROPLASTY
    HEIMKE, G
    [J]. ADVANCED MATERIALS, 1994, 6 (02) : 165 - 170
  • [4] *ISO, 1992, 72065 ISO
  • [5] Friction in orthopaedic zirconia taper assemblies
    Macdonald, W
    Aspenberg, A
    Jacobsson, CM
    Carlsson, LV
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) : 685 - 692
  • [6] INVESTIGATIONS INTO THE MECHANICAL RELIABILITY OF CERAMIC FEMORAL HEADS FOR HIP JOINTS
    MIDDLETON, J
    PANDE, GN
    RICHTER, HG
    WILLMANN, G
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (6-7) : 503 - 506
  • [7] Early failure of a zirconia femoral head prosthesis: Fracture or fatigue?
    Panagiotopoulos, Elias C.
    Kallivokas, Alkiviadis G.
    Koulioumpas, Ioannis
    Mouzakis, Dionysios E.
    [J]. CLINICAL BIOMECHANICS, 2007, 22 (07) : 856 - 860
  • [8] Panagopoulos A, 2006, EUR J ORTHOP SURG TR, V16, P55
  • [9] Improvement of the reliability of ceramic hip joint implants
    Weisse, B
    Zahner, M
    Weber, W
    Rieger, W
    [J]. JOURNAL OF BIOMECHANICS, 2003, 36 (11) : 1633 - 1639
  • [10] WILLIAM G, 2003, CERAMICS ORTHOPAEDIC