Alumina-on-alumina total hip arthroplasty - A five-year minimum follow-up study

被引:135
作者
Yoo, JJ [1 ]
Kim, YM [1 ]
Yoon, KS [1 ]
Koo, KH [1 ]
Song, WS [1 ]
Kim, HJ [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Orthopaed Surg, Seoul, South Korea
关键词
D O I
10.2106/JBJS.D.01753
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Ceramic-on-ceramic couplings are attractive alternative bearing surfaces that have been reported to eliminate or reduce problems related to polyethylene wear debris. Disappointing experiences with alumina ceramic bearings in the past have led to many improvements in the manufacture and design of ceramic implants. The purpose of the present study was to report the results of contemporary alumina-on-alumina total hip arthroplasties with regard to wear, osteolysis, and fracture of the ceramic after a minimum duration of follow-up of five years. Methods: We evaluated the results of a consecutive series of 100 primary alumina-on-alumina total hip arthroplasties that had been performed with use of a metal-backed socket and a cementless stem in eighty-four patients. All of the patients were sixty-five years of age or younger (mean age, forty-one years), and a single surgeon performed all of the procedures. After a minimum duration of follow-up of sixty months, one patient (one hip) had died and four patients (six hips) had been lost to follow-up, leaving a total of seventy-nine patients (ninety-three hips) available for study. All of these patients were evaluated clinically and radiographically with special attention to wear, periprosthetic osteolysis, and ceramic failure. Results: The mean Harris hip score was 97 points at the time of the latest follow-up evaluation. All prostheses demonstrated radiographic evidence of bone ingrowth. No implant was loose radiographically, and no implant was revised. Ceramic wear was not detectable in the thirty-seven hips in which the femoral head could be differentiated from the cup on radiographs. Periprosthetic osteolysis was not observed in any hip. A fracture of the alumina femoral head and a peripheral chip fracture of the alumina insert occurred in one hip following a motor-vehicle accident. Conclusions: The results of contemporary alumina-on-alumina total hip arthroplasty with a metal-backed socket and a cementless stem were encouraging after a minimum duration of follow-up of five years. We believe that these improved alumina-on-alumina bearing implants offer a promising option for younger, active patients.
引用
收藏
页码:530 / 535
页数:6
相关论文
共 30 条
[1]   Revision total hip arthroplasty performed after fracture of a ceramic femoral head - A multicenter survivorship study [J].
Allain, J ;
Roudot-Thoraval, F ;
Delecrin, J ;
Anract, P ;
Migaud, H ;
Goutallier, D .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (05) :825-830
[2]   Failure of a stainless-steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head - A case report [J].
Allain, J ;
Goutallier, D ;
Voisin, MC ;
Lemouel, S .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (09) :1355-1360
[3]  
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[4]  
BIERBAUM BE, 2002, CLIN ORTHOP RELAT R, V405, P158
[5]   Hybrid alumina total hip arthroplasty using a press-fit metal-backed socket in patients younger than 55 years - A six- to 11-year evaluation [J].
Bizot, P ;
Hannouche, D ;
Nizard, R ;
Witvoet, J ;
Sedel, L .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (02) :190-194
[6]  
BIZOT P, 2000, CLIN ORTHOP RELAT R, V379, P134
[7]  
Callaway GH, 1995, J ARTHROPLASTY, V10, P855
[8]   New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty [J].
D'Antonio, J ;
Capello, W ;
Manley, M ;
Bierbaum, B .
JOURNAL OF ARTHROPLASTY, 2002, 17 (04) :390-397
[9]   POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS [J].
ENGH, CA ;
BOBYN, JD ;
GLASSMAN, AH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01) :45-55
[10]  
Engh CA., 1990, CLIN ORTHOP RELAT R, V257, P107