Inductively coupled plasma mass spectrometry; Polyamide; Post-printing treatment; Three-dimensional printing; Sample preparation; Solid phase extraction; Trace metals; PLASMA-MASS-SPECTROMETRY; ICP-MS DETERMINATION; SAMPLE PREPARATION; SEQUENTIAL INJECTION; ANALYTICAL-CHEMISTRY; ATOMIC SPECTROSCOPY; POROUS POLYMER; FLOW-INJECTION; HEAVY-METALS; ELEMENTS;
D O I
10.1016/j.talanta.2022.123237
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Post-printing treatment is a promising means of enhancing the performance of devices manufactured using conventional three-dimensional printing (3DP) technologies. In this study we developed a post-printing solution foaming process-involving respective treatment with formic acid (60%, v/v) and sodium bicarbonate (5%, w/v) solutions to generate CO2 as a foaming agent-to increase the surface roughness and porosity of the polyamide 6 (PA6) monolithic packing in a multimaterial fused deposition modeling 3D-printed solid phase extraction (SPE) column, thereby enhancing the extraction of Mn, Co, Ni, Cu, Zn, Cd, and Pb ions from complicated real samples prior to their determination using inductively coupled plasma mass spectrometry. After optimizing the column fabrication process, the solution foaming treatment process, the extraction conditions, and the automatic analytical system, the 3D-printed SPE column incorporating the solution foaming-treated PA6 monolithic packing extracted these metal ions with 18.9- to 42.0-fold enhancements, relative to those of the as-printed column, with absolute extraction efficiencies all greater than 94.3% and method detection limits ranging from 0.2 to 7.7 ng L-1. We verified the reliability and applicability of this method through analyses of the tested metal ions in several reference materials (CASS-4, SLEW-3, 1643f, and 2670a) and spike analyses of seawater, river water, ground water, and urine samples. We conclude that post-printing treatment can dramatically improve the performance of 3D-printed analytical devices.
引用
收藏
页数:9
相关论文
共 65 条
[1]   Matrix effects in inductively coupled plasma mass spectrometry: A review [J].
Agatemor, Christian ;
Beauchemin, Diane .
ANALYTICA CHIMICA ACTA, 2011, 706 (01) :66-83
[2]   ICP-MS and trace element analysis as tools for better understanding medical conditions [J].
Amais, Renata S. ;
Donati, George L. ;
Zezzi Arruda, Marco A. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2020, 133
[3]   Sample preparation for the determination of metals in food samples using spectroanalytical methods - A review [J].
Andrade Korn, Maria Das Gracas ;
da Boa Morte, Elane Santos ;
Muniz Batista dos Santos, Daniele Cristina ;
Castro, Jacira Teixeira ;
Pereira Barbosa, Jose Tiago ;
Teixeira, Alete Paixao ;
Fernandes, Andrea Pires ;
Welz, Bernhard ;
Carvalho dos Santos, Wagna Piler ;
Guimaraes Nunes dos Santos, Eduardo Batista ;
Korn, Mauro .
APPLIED SPECTROSCOPY REVIEWS, 2008, 43 (02) :67-92
[4]   Fused Deposition Modeling Enables the Low-Cost Fabrication of Porous, Customized-Shape Sorbents for Small-Molecule Extraction [J].
Belka, Mariusz ;
Ulenberg, Szymon ;
Baczek, Tomasz .
ANALYTICAL CHEMISTRY, 2017, 89 (08) :4373-4376
[5]   3D printed device for the automated preconcentration and determination of chromium (VI) [J].
Calderilla, Carlos ;
Maya, Fernando ;
Cerda, Victor ;
Leal, Luz O. .
TALANTA, 2018, 184 :15-22
[6]   Solid phase extraction of trace elements [J].
Camel, V .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2003, 58 (07) :1177-1233
[7]   Additive-manufactured (3D-printed) electrochemical sensors: A critical review [J].
Cardoso, Rafael M. ;
Kalinke, Cristiane ;
Rocha, Raquel G. ;
dos Santos, Pamyla L. ;
Rocha, Diego P. ;
Oliveira, Paulo R. ;
Janegitz, Bruno C. ;
Bonacin, Juliano A. ;
Richter, Eduardo M. ;
Munoz, Rodrigo A. A. .
ANALYTICA CHIMICA ACTA, 2020, 1118 :73-91
[8]   Magnetic metal-organic framework composites for dual-column solid-phase microextraction combined with ICP-MS for speciation of trace levels of arsenic [J].
Chen, Zhenna ;
Chen, Beibei ;
He, Man ;
Hu, Bin .
MICROCHIMICA ACTA, 2020, 187 (01)
[9]   Opportunities for 3D printed millifluidic platforms incorporating on-line sample handling and separation [J].
Cocovi-Solberg, David J. ;
Worsfold, Paul J. ;
Miro, Manuel .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2018, 108 :13-22
[10]   3D-printed miniaturized fluidic tools in chemistry and biology [J].
Dixit, C. K. ;
Kadimisetty, K. ;
Rusling, J. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2018, 106 :37-52