Application of refocused steady-state free-precession methods at 1.5 and 3 T to in vivo high-resolution MRI of trabecular bone: Simulations and experiments
被引:40
作者:
Banerjee, S
论文数: 0引用数: 0
h-index: 0
机构:Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
Banerjee, S
Han, ET
论文数: 0引用数: 0
h-index: 0
机构:Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
Han, ET
Krug, R
论文数: 0引用数: 0
h-index: 0
机构:Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
Krug, R
Newitt, DC
论文数: 0引用数: 0
h-index: 0
机构:Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
Newitt, DC
Majumdar, S
论文数: 0引用数: 0
h-index: 0
机构:Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
Majumdar, S
机构:
[1] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
[2] Univ Calif Berkeley, Joint Grad Grp Bioengn, San Francisco, CA 94107 USA
Purpose: To evaluate the potential of fully-balanced steady-state free-precession (SSFP) sequences in in vivo high-resolution (HR) MRI of trabecular bone at field strengths of 1.5 and 3 T by simulation and experimental methods. Materials and Methods: Using simulation studies, refocused SSFP acquisition was optimized for our imaging purposes with a focus on signal-to-noise ratio (SNR) and SNR efficiency. The signal behavior in trabecular bone was estimated using a magnetostatic model of the trabecular bone and marrow. Eight normal volunteers were imaged at the proximal femur, calcaneus, and the distal tibia on a GE Signa scanner at 1.5 and at 3 T with an optimized single-acquisition SSFP sequence (three-dimensional FIESTA) and an optimized multiple-acquisition SSFP sequence (three-dimensional FIESTA-c). Images were also acquired with a fast gradient echo (FGRE) sequence for evaluation of the SNR performance of SSFP methods. Results: Refocused SSFP images outperformed FGRE acquisitions in both SNR and SNR efficiency at both field strengths. At 3 T, susceptibility effects were visible in FIESTA and FGRE images and much reduced in FIESTA-c images. The magnitude of SNR boost at 3 T was closely predicted by simulations. Conclusion: Single-acquisition SSFP (at 1.5 T) and multiple-acquisition SSFP (at 3 T) hold great potential for HR-MRI of trabecular bone. (c) 2005 Wiley-Liss, Inc.