Nature of sensitive high-order resonant modes in piezoelectric excited millimeter sized cantilever (PEMC) sensors

被引:9
作者
Lakshmanan, Ramji S. [1 ]
Mutharasan, Raj [1 ]
机构
[1] Drexel Univ, Dept Chem & Biol Engn, Philadelphia, PA 19104 USA
关键词
Resonance; Density; In-liquid; Damping; High-order modes; Sensitivity; MASS-CHANGE SENSITIVITY; BACILLUS-ANTHRACIS; MICROCANTILEVERS; ALKANETHIOLS; GOLD;
D O I
10.1016/j.sna.2011.07.003
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
High-order resonant modes of piezoelectric-excited cantilever (PEMC) sensors were previously shown to be very highly sensitive at 0.3-2 fg/Hz for in-liquid applications. The purpose of this work is to show experimentally and with finite element model (FEM) simulations that such sensitive modes are strongly influenced by the sensor width, suggesting that the sensitive modes are torsional or buckling modes. From experimental observations the resonant frequency of high-order modes had a strong dependence on width, where a sensor with a smaller width had resonant modes at a higher frequency. Also the FEM simulations indicate that in this frequency range there are resonant modes with a buckling nature that change for a decrease in width, consistent with experimental observations. In order to establish that the width-dependent modes are mass-sensitive in liquids, resonant frequency change to density changes in flow experiments under fully liquid-immersed conditions were determined. Average frequency shifts of 475 +/- 49 Hz (n = 5), 533 +/- 31 Hz (n=5), 715 +/- 103 Hz (n = 5) and 725 +/- 37 Hz (n = 5) were obtained for the four designs investigated in response to a density change of 0.0118 g/cm(3). The results show that the resonant frequency response to variations in the geometry provides insightful data on the role of width in PEMC sensor design. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:79 / 86
页数:8
相关论文
共 21 条
[1]   Use of piezoelectric-excited millimeter-sized cantilever sensors to measure albumin interaction with self-assembled monolayers of alkanethiols having different functional headgroups [J].
Campbell, GA ;
Mutharasan, R .
ANALYTICAL CHEMISTRY, 2006, 78 (07) :2328-2334
[2]   Piezoelectric-excited millimeter-sized cantilever (PEMC) sensors detect Bacillus anthracis at 300 spores/mL [J].
Campbell, GA ;
Mutharasan, R .
BIOSENSORS & BIOELECTRONICS, 2006, 21 (09) :1684-1692
[3]   Monitoring of the self-assembled monolayer of 1-hexadecanethiol on a gold surface at nanomolar concentration using a piezo-excited millimeter-sized cantilever sensor [J].
Campbell, GA ;
Mutharasan, R .
LANGMUIR, 2005, 21 (25) :11568-11573
[4]   Detection of pathogen Escherichia coli O157:H7 using self-excited PZT-glass microcantilevers [J].
Campbell, GA ;
Mutharasan, R .
BIOSENSORS & BIOELECTRONICS, 2005, 21 (03) :462-473
[5]   Sensing of liquid level at micron resolution using self-excited millimeter-sized PZT-cantilever [J].
Campbell, GA ;
Mutharasan, R .
SENSORS AND ACTUATORS A-PHYSICAL, 2005, 122 (02) :326-334
[6]   Detection of Bacillus anthracis spores and a model protein using PEMC sensors in a flow cell at 1 mL/min [J].
Campbell, Gossett A. ;
Mutharasan, Raj .
BIOSENSORS & BIOELECTRONICS, 2006, 22 (01) :78-85
[7]   PEMC sensor's mass change sensitivity is 20 pg/Hz under liquid immersion [J].
Campbell, Gossett A. ;
Mutharasan, Raj .
BIOSENSORS & BIOELECTRONICS, 2006, 22 (01) :35-41
[8]   Enhanced functionality of cantilever based mass sensors using higher modes [J].
Dohn, S ;
Sandberg, R ;
Svendsen, W ;
Boisen, A .
APPLIED PHYSICS LETTERS, 2005, 86 (23) :1-3
[9]   Integrated cantilever sensors with a torsional resonance mode for ultraresoluble on-the-spot bio/chemical detection [J].
Jin, Dazhong ;
Li, Xinxin ;
Bao, Hanhan ;
Zhang, Zhixiang ;
Wang, Yuelin ;
Yu, Haitao ;
Zuo, Guomin .
APPLIED PHYSICS LETTERS, 2007, 90 (04)
[10]   Use of thermally induced higher order modes of a microcantilever for mercury vapor detection [J].
Kadam, Ashwini R. ;
Nordin, Gregory P. ;
George, Michael A. .
JOURNAL OF APPLIED PHYSICS, 2006, 99 (09)