Method for confirmation of synthetic corticosteroids in doping urine samples by liquid chromatography-electrospray ionisation mass spectrometry

被引:59
作者
Fluri, K [1 ]
Rivier, L [1 ]
Dienes-Nagy, A [1 ]
You, CH [1 ]
Maître, A [1 ]
Schweizer, C [1 ]
Saugy, M [1 ]
Mangin, P [1 ]
机构
[1] Inst Univ Med Legale, CH-1005 Lausanne, Switzerland
关键词
doping analysis; corticosteroids; steroids;
D O I
10.1016/S0021-9673(01)01065-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this study, we report on the development of a method to confirm simultaneously nine of the most commonly abused synthetic corticosteroids in urine based on liquid chromatography-electrospray ionisation mass spectrometry. A considerable simplified sample preparation procedure, including liquid-liquid phase extraction with Extrelut-NT3 columns, provided both excellent sample purification and high overall recoveries. Complete HPLC separations were obtained on a reversed-phase column with 1 mM ammonium acetate-acetonitrile (60:40, v/v) as mobile phase. Mass spectral acquisition was done in the negative ion, and selected ion monitoring modes to identify the drugs with at least three characteristic ions. Detection limits were determined at less than or equal to1 ng/ml and the confirmation limits at 1 to 5 ng/ml. (C) 2001 Published by Elsevier Science B.V.
引用
收藏
页码:87 / 95
页数:9
相关论文
共 24 条
[1]  
Antignac JP, 2000, RAPID COMMUN MASS SP, V14, P33, DOI 10.1002/(SICI)1097-0231(20000115)14:1<33::AID-RCM829>3.0.CO
[2]  
2-R
[3]   Identification of ten corticosteroids in human hair by liquid chromatography-ionspray mass spectrometry [J].
Cirimele, V ;
Kintz, P ;
Dumestre, V ;
Goullé, JP ;
Ludes, B .
FORENSIC SCIENCE INTERNATIONAL, 2000, 107 (1-3) :381-388
[4]   Immunoaffinity chromatography combined on-line with high-performance liquid chromatography mass spectrometry for the determination of corticosteroids [J].
Creaser, CS ;
Feely, SJ ;
Houghton, E ;
Seymour, M .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :37-43
[5]  
DELEHAUT P, 1997, J CHROMATOGR B, V696, P203
[6]   Identification of main corticosteroids as illegal feed additives in milk replacers by liquid chromatography atmospheric pressure chemical ionisation mass spectrometry [J].
Fiori, M ;
Pierdominici, E ;
Longo, F ;
Brambilla, G .
JOURNAL OF CHROMATOGRAPHY A, 1998, 807 (02) :219-227
[7]   Compared interest between hair analysis and urinalysis in doping controls -: Results for amphetamines, corticosteroids and anabolic steroids in racing cyclists [J].
Gaillard, Y ;
Vayssette, F ;
Pépin, G .
FORENSIC SCIENCE INTERNATIONAL, 2000, 107 (1-3) :361-379
[8]   Simultaneous determination of anabolic and catabolic steroid hormones in meat by gas chromatography mass spectrometry [J].
Hartmann, S ;
Steinhart, H .
JOURNAL OF CHROMATOGRAPHY B, 1997, 704 (1-2) :105-117
[9]   THE USE OF COMBINED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY TO CONFIRM THE ADMINISTRATION OF SYNTHETIC CORTICOSTEROIDS TO HORSES [J].
HOUGHTON, E ;
DUMASIA, MC ;
WELLBY, JK .
BIOMEDICAL MASS SPECTROMETRY, 1981, 8 (11) :558-564
[10]  
*INT OL COMM, 1999, ANT COD OL MOV