Diagnosis of small-bowel pathology using paired capsule endoscopy with two different devices: a randomized study

被引:56
作者
Hartmann, D. [1 ,2 ]
Eickhoff, A. [2 ]
Damian, U. [2 ]
Riemann, J. F. [2 ]
机构
[1] Klinikum Stadt Ludwigshafen GmbH, Dept Med C, D-67063 Ludwigshafen, Germany
[2] Ludwigshafen Rheim, Acad Teaching Hosp, Dept Gastroenterol, Rhein, Germany
关键词
D O I
10.1055/s-2007-966943
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background and study aims: Capsule endoscopy is an established technique in the evaluation of obscure gastrointestinal bleeding. The primary aim of the study is to compare the diagnostic yield of the two different capsule endoscopes. Patients and methods: Patients with signs of mid-gastrointestinal bleeding after negative upper endoscopy and colonoscopy studies were included. Patients were randomized to undergo two capsule endoscopies using different capsule endoscopes (Given Pillcam SB and Olympus EndoCapsule) in random order. Results: Forty patients (18 women, 22 men) were included in the study. All EndoCapsules reached the colon within the mean recording time of 591 +/- 52 minutes, whereas 33 out of 40 PillCam SB reached the colon within the mean recording time of 471 +/- 27 minutes. Lesions were detected in 31 patients: with both devices in 24, only by PillCam SB in 2, and, conversely, only by EndoCapsule in 5 (not significant). With regard to lesions with high bleeding potential (P2 lesions), PillCarn SB detected them in 22 patients, EndoCapsule in 25 patients (not significant). In all four cases of intestinal P2 lesions that were not detected by the PillCam SB but were detected by the EndoCapsule, the PillCam SB had not reached the cecum. Conclusion: In this study there was a statistically nonsignificant trend for the EndoCapsule to detect more bleeding sources in patients with suspected small bowel bleeding than did the Pillcam SB, which may have been due to the longer recording times with the currently available EndoCapsule.
引用
收藏
页码:1041 / 1045
页数:5
相关论文
共 14 条
[1]   Second capsule endoscopy for patients with severe iron deficiency anemia [J].
Bar-Meir, S ;
Eliakim, R ;
Nadler, M ;
Barkay, O ;
Fireman, Z ;
Scapa, E ;
Chowers, Y ;
Bardan, E .
GASTROINTESTINAL ENDOSCOPY, 2004, 60 (05) :711-713
[2]   A multi-center randomized comparison of the endocapsule: Olympus inc and the pillcam SB: Given imaging in patients with obscure GI bleeding [J].
Cave, David R. ;
Fleischer, Dalid E. ;
Gostout, Christopher ;
Faigel, Douglas O. ;
Leighton, Jonathan A. ;
Heigh, Russell I. ;
Sharma, Virender K. ;
Mergener, Klaus ;
Bhattacharya, Knishka ;
Rajan, Elizabeth ;
Foley, Anne ;
Lee, Michad ;
Knipschield, Mary ;
Hibberd, Patricia .
GASTROINTESTINAL ENDOSCOPY, 2007, 65 (05) :AB125-AB125
[3]   Structured terminology for capsule endoscopy:: Results of retrospective testing and validation in 766 small-bowel investigations [J].
Delvaux, M ;
Friedman, S ;
Keuchel, M ;
Hagenmüller, F ;
Weinstein, M ;
Cave, D ;
de Franchis, R ;
Gay, G ;
Korman, LY .
ENDOSCOPY, 2005, 37 (10) :945-950
[4]   The first prospective controlled trial comparing wireless capsule endoscopy with push enteroscopy in chronic gastrointestinal bleeding [J].
Ell, C ;
Remke, S ;
May, A ;
Helou, L ;
Henrich, R ;
Mayer, G .
ENDOSCOPY, 2002, 34 (09) :685-689
[5]   A prospective two-center study comparing wireless capsule endoscopy with intraoperative enteroscopy in patients with obscure GI bleeding [J].
Hartmann, D ;
Schmidt, H ;
Bolz, G ;
Schilling, D ;
Kinzel, F ;
Eickhoff, A ;
Huschner, W ;
Möller, K ;
Jakobs, R ;
Reitzig, P ;
Weickert, U ;
Gellert, K ;
Schultz, H ;
Guenther, K ;
Hollerbuhl, H ;
Schoenleben, K ;
Schulz, HJ ;
Riemann, JF .
GASTROINTESTINAL ENDOSCOPY, 2005, 61 (07) :826-832
[6]  
Hartmann D, 2003, Z GASTROENTEROL, V41, P377
[7]   Yield of repeat wireless video capsule endoscopy in patients with obscure gastrointestinal bleeding [J].
Jones, BH ;
Fleischer, DE ;
Sharma, VK ;
Heigh, RI ;
Shiff, AD ;
Hernandez, JL ;
Leighton, JA .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2005, 100 (05) :1058-1064
[8]   Meta-analysis: capsule enteroscopy vs. conventional modalities in diagnosis of small bowel diseases [J].
Marmo, R ;
Rotondano, G ;
Piscopo, R ;
Bianco, MA ;
Cipolletta, L .
ALIMENTARY PHARMACOLOGY & THERAPEUTICS, 2005, 22 (07) :595-604
[9]   European Society of Gastrointestinal Endoscopy (ESGE) Video capsule endoscopy: Update to guidelines (May 2006) [J].
Rey, J. F. ;
Ladas, S. ;
Alhassani, A. ;
Kuznetsov, K. .
ENDOSCOPY, 2006, 38 (10) :1047-1053
[10]   European Society of Gastrointestinal Endoscopy guideline for video capsule endoscopy [J].
Rey, JF ;
Gay, G ;
Kruse, A ;
Lambert, R .
ENDOSCOPY, 2004, 36 (07) :656-658