Measurements of the internal magnetic field using the B-Stark motional Stark effect diagnostic on DIII-D (inivited)

被引:19
|
作者
Pablant, N. A. [1 ]
Burrell, K. H. [2 ]
Groebner, R. J. [2 ]
Holcomb, C. T. [3 ]
Kaplan, D. H. [2 ]
机构
[1] Univ Calif San Diego, La Jolla, CA 92093 USA
[2] Gen Atom Co, San Diego, CA 92186 USA
[3] Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2010年 / 81卷 / 10期
关键词
EXCHANGE RECOMBINATION SPECTROSCOPY; RADIAL ELECTRIC-FIELD; ITER; RECONSTRUCTION; PROFILES;
D O I
10.1063/1.3491209
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Results are presented from the B-Stark diagnostic installed on the DIII-D tokamak. This diagnostic provides measurements of the magnitude and direction of the internal magnetic field. The B-Stark system is a version of a motional Stark effect (MSE) diagnostic based on the relative line intensities and spacing of the Stark split D-alpha emission from injected neutral beams. This technique may have advantages over MSE polarimetry based diagnostics in future devices, such as the ITER. The B-Stark diagnostic technique and calibration procedures are discussed. The system is shown to provide accurate measurements of B-theta/B-T and vertical bar B vertical bar over a range of plasma conditions. Measurements have been made with toroidal fields in the range of 1.2-2.1 T, plasma currents in the range 0.5-2.0 MA, densities between 1.7 and 9.0 X 10(19) m(-3), and neutral beam voltages between 50 and 81 keV. The viewing direction and polarization dependent transmission properties of the collection optics are found using an in situ beam into gas calibration. These results are compared to values found from plasma equilibrium reconstructions and the MSE polarimetry system on DIII-D. (C) 2010 American Institute of Physics. [doi:10.1063/1.3491209]
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Measurements of the internal magnetic field on DIII-D using intensity and spacing of the motional Stark multiplet
    Pablant, N. A.
    Burrell, K. H.
    Groebner, R. J.
    Kaplan, D. H.
    Holcomb, C. T.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10):
  • [2] q profile measurements with the motional Stark effect diagnostic in the DIII-D tokamak
    Rice, BW
    FUSION ENGINEERING AND DESIGN, 1997, 34-35 : 135 - 142
  • [3] q profile measurements with the motional Stark effect diagnostic in the DIII-D tokamak
    Lawrence Livermore Natl Lab, Livermore, United States
    Fusion Eng Des, (135-142):
  • [4] Motional stark effect upgrades on DIII-D
    Rice, B.W.
    Nilson, D.G.
    Wroblewski, D.
    Review of Scientific Instruments, 1995, 66 (1 pt 2):
  • [5] MOTIONAL STARK-EFFECT UPGRADES ON DIII-D
    RICE, BW
    NILSON, DG
    WROBLEWSKI, D
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01): : 373 - 375
  • [6] Motional Stark effect diagnostic expansion on DIII-D for enhanced current and Er profile measurements
    Holcomb, C. T.
    Makowski, M. A.
    Jayakumar, R. J.
    Allen, S. A.
    Ellis, R. M.
    Geer, R.
    Behne, D.
    Morris, K. L.
    Seppala, L. G.
    Moller, J. M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (10):
  • [7] MOTIONAL STARK-EFFECT POLARIMETRY FOR A CURRENT PROFILE DIAGNOSTIC IN DIII-D
    WROBLEWSKI, D
    BURRELL, KH
    LAO, LL
    POLITZER, P
    WEST, WP
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10): : 2988 - 2988
  • [8] Magnetohydrodynamic interference with the edge pedestal motional Stark effect diagnostic on DIII-D
    King, J. D.
    Makowski, M. A.
    Holcomb, C. T.
    Allen, S. L.
    Hill, D. N.
    La Haye, R. J.
    Turco, F.
    Petty, C. C.
    Van Zeeland, M. A.
    Rhodes, T. L.
    Meyer, W. H.
    Geer, R.
    Morse, E. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (03):
  • [9] A digital lock-in upgrade of the motional Stark effect diagnostic on DIII-D
    King, J. D.
    Makowski, M. A.
    Holcomb, C. T.
    Allen, S. L.
    Geer, R.
    Meyer, W. H.
    Hill, D. N.
    Pham, D.
    Morse, E. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (10):
  • [10] MOTIONAL STARK-EFFECT POLARIMETRY FOR A CURRENT PROFILE DIAGNOSTIC IN DIII-D
    WROBLEWSKI, D
    BURRELL, KH
    LAO, LL
    POLITZER, P
    WEST, WP
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (11): : 3552 - 3556