Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators

被引:92
作者
Hurkmans, CW
Scheepers, E
Springorum, BGF
Uiterwaal, H
机构
[1] Catharina Hosp Eindhoven, Dept Radiotherapy, NL-5602 ZA Eindhoven, Netherlands
[2] Eindhoven Univ Technol, Dept Biomed Engn, NL-5600 MB Eindhoven, Netherlands
[3] Onze Lieve Vrouw Hosp, Dept Cardiol, Amsterdam, Netherlands
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2005年 / 63卷 / 01期
关键词
cardioverter-defibrillator; pacemaker; radiotherapy; radiation damage;
D O I
10.1016/j.ijrobp.2005.04.047
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Radiotherapy can influence the functioning of pacemakers and implantable cardioverter-defibrillators (ICDs). ICDs offer the same functionality as pacemakers, but are also able to deliver a high-voltage shock to the heart if needed. Guidelines for radiotherapy treatment of patients with an implanted rhythm device have been published in 1994 by The American Association of Physicists in Medicine, and are based only on experience with pacemakers. Data on the influence of radiotherapy on ICDs are limited. The objective of our study is to determine the influence of radiotherapy on the latest generation of ICDs. Methods and Materials: Eleven modern ICDs have been irradiated in our department. The irradiation was performed with a 6-MV photon beam. The given dose was fractionated up to a cumulative dose of 120 Gy. Two to 5 days passed between consecutive irradiations. Frequency, output, sensing, telemetry, and shock energy were monitored. Results: Sensing interference by ionizing radiation on all ICDs has been demonstrated. For four ICDs, this would have caused the inappropriate delivery of a shock because of interference. At the end of the irradiation sessions, all devices had reached their point of failure. Complete loss of function was observed for four ICDs at dose levels between 0.5 Gy and 1.5 Gy. Conclusions: The effect of radiation therapy on the newest generation of ICDs varies widely. If tachycardia monitoring and therapy are functional (programmed on) during irradiation, the ICD might inappropriately give antitachycardia therapy, often resulting in a shock. Although most ICDs did not fail below 80 Gy, some devices had already failed at doses below 1.5 Gy. Guidelines are formulated for the treatment of patients with an ICD. (c) 2005 Elsevier Inc.
引用
收藏
页码:282 / 289
页数:8
相关论文
共 17 条
[1]   DAMAGING EFFECT OF THERAPEUTIC RADIATION ON PROGRAMMABLE PACEMAKERS [J].
ADAMEC, R ;
HAEFLIGER, JM ;
KILLISCH, JP ;
NIEDERER, J ;
JAQUET, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :146-150
[2]   THERAPEUTIC RADIATION AND PACEMAKERS [J].
CALFEE, RV .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :160-161
[3]   Implantable cardiac defibrillators may be damaged by radiation therapy [J].
Hoecht, S ;
Rosenthal, P ;
Sancar, D ;
Behrens, S ;
Hinkelbein, W ;
Hoeller, U .
JOURNAL OF CLINICAL ONCOLOGY, 2002, 20 (08) :2212-2213
[4]   Radiotherapy in patients with cardiac pacemakers [J].
Last, A .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841) :4-10
[5]   MANAGEMENT OF RADIATION ONCOLOGY PATIENTS WITH IMPLANTED CARDIAC-PACEMAKERS - REPORT OF AAPM TASK GROUP NO-34 [J].
MARBACH, JR ;
SONTAG, MR ;
VANDYK, J ;
WOLBARST, AB .
MEDICAL PHYSICS, 1994, 21 (01) :85-90
[6]   Influence of high-energy photon beam irradiation on pacemaker operation [J].
Mouton, J ;
Haug, R ;
Bridier, A ;
Dodinot, B ;
Eschwege, F .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :2879-2893
[7]  
Ngu S L, 1993, Australas Radiol, V37, P105, DOI 10.1111/j.1440-1673.1993.tb00027.x
[8]  
Niehaus M, 2001, HEART, V86, P246
[9]   RADIATION-INDUCED EFFECTS IN MULTIPROGRAMMABLE PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS [J].
RODRIGUEZ, F ;
FILIMONOV, A ;
HENNING, A ;
COUGHLIN, C ;
GREENBERG, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (12) :2143-2153
[10]  
SALMI J, 1990, STRAHLENTHER ONKOL, V166, P153