Validation of the mobil-O-Graph: 24 h-blood pressure measurement device

被引:236
作者
Weiss, Wolfgang [1 ]
Toelle, Markus [1 ]
Zidek, Walter [1 ]
van der Giet, Markus [1 ]
机构
[1] Charite, Med Klin 4, Dept Nephrol, D-12200 Berlin, Germany
关键词
ambulatory blood pressure measurement; British Hypertension Society protocol; 24 h-blood pressure; validation; HYPERTENSION-SOCIETY PROTOCOL; IMPACT;
D O I
10.1097/MBP.0b013e328338892f
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Twenty-four-hour blood pressure measurement is of importance not only in the detection of hypertension but also in the detection of blood pressure changes in hypertensive and nonhypertensives over the day to identify, for example, nondipper hypertensives. This study describes the validation of the mobil-O-Graph according to the criteria of the British Hypertension Society (BHS). Methods For each patient three readings obtained by the mobil-O-Graph were compared with auscultatory sphygmomanometric readings obtained by two trained clinicians. The sphygmomanometric reference measurements were alternated with the readings obtained by the device. Eighty-five patients (mean age 53.4 +/- 18.4 years) were recruited for the BHS protocol. Differences between blood pressure values of the test device and the mercury reading were calculated for each measurement. Results In the BHS validation procedure the mean differences of the observer readings and the test device were -2.2 +/- 6.7 (systolic) and -0.6 +/- 5.6 mmHg (diastolic) for observer 1 and -2.2 +/- 7.3 mmHg (systolic) and -0.4 +/- 6.1 mmHg (diastolic) for observer 2. The device achieved grade A for systolic and diastolic blood pressure for both the observers 1 and 2 leading to a final grade A/A. According to the BHS protocol the measurements of the device have to be considered 'very accurate and with no error of clinical relevance'. Conclusion The device met the accuracy requirements of the BHS standard and can be recommended for clinical use. Blood Press Monit 15: 225-228 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:225 / 228
页数:4
相关论文
共 7 条
[1]  
*ASS ADV MED INSTR, 2002, AM NAT STAND MAN EL
[2]  
Mancia G, 2007, EUR HEART J, V28, P1462, DOI [10.1093/eurheartj/ehm236, 10.1093/eurheartj/eht151]
[3]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[4]   AN OUTLINE OF THE REVISED BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF BLOOD-PRESSURE MEASURING DEVICES [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
ALTMAN, DG ;
BLAND, M ;
COATS, A ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1993, 11 (06) :677-679
[5]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[6]   Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes - Principal results of the Conduit Artery Function Evaluation (CAFE) study [J].
Williams, B ;
Lacy, PS ;
Thom, SM ;
Cruickshank, K ;
Stanton, A ;
Collier, D ;
Hughes, AD ;
Thurston, H .
CIRCULATION, 2006, 113 (09) :1213-1225
[7]   Impact of Statin Therapy on Central Aortic Pressures and Hemodynamics Principal Results of the Conduit Artery Function Evaluation-Lipid-Lowering Arm (CAFE-LLA) Study [J].
Williams, Bryan ;
Lacy, Peter S. ;
Cruickshank, J. Kennedy ;
Collier, David ;
Hughes, Alun D. ;
Stanton, Alice ;
Thom, Simon ;
Thurston, Herbert .
CIRCULATION, 2009, 119 (01) :53-U98