Initial pulse-echo imaging results with one-dimensional capacitive micromachined ultrasonic transducer arrays

被引:0
作者
Oralkan, Ö [1 ]
Jin, XC [1 ]
Kaviani, K [1 ]
Ergun, AS [1 ]
Degertekin, FL [1 ]
Karaman, M [1 ]
Khuri-Yakub, BT [1 ]
机构
[1] Stanford Univ, Edward L Ginzton Lab, Stanford, CA 94305 USA
来源
2000 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2 | 2000年
关键词
ultrasonic transducer; 1-D array; capacitive micromachined transducer; beamforming; medical ultrasound imaging;
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In this paper, the first experimental results of pulse-echo imaging employing a 1-D capacitive micromachined ultrasonic transducer array are presented. A 1-D array consisting of 16 elements is used in the experiment. A wire phantom consisting of seven steel wires immersed in vegetable oil is used as the imaging target. A B-scan image with a sector angle of 90 degrees and an image depth of 150 mm is reconstructed by employing RF beamforming and synthetic phase array approaches. The reconstructed image is displayed at different display dynamic ranges. The measured near and far side lobe levels are around -25 dB and -35 dB, respectively. The noise floor of the image is below -50 dB.
引用
收藏
页码:959 / 962
页数:4
相关论文
empty
未找到相关数据