Bone-hydroxyapatite interface in retrieved hydroxyapatite-coated titanium implants: A clinical and histologic report

被引:0
|
作者
Piattelli, A
Scarano, A
Di Alberti, L
Piattelli, M
机构
[1] Univ G dAnnunzio, Sch Dent, Chieti, Italy
[2] UCL Eastman Dent Hosp Inst Oral Hlth Care Sci, Dept Oral Med, London WC1X 8LD, England
关键词
cathodoluminescence; hydroxyapatite; interface; osseointegration; titanium implants;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Two hydroxyapatite-coated implants were retrieved after 12 months of loading because of a fracture of the abutments. The specimens were treated to obtain thin ground sections, and a microprobe chemical analysis was done under a scanning electron microscope equipped with an energy-dispersive x-ray analysis and cathodoluminescence system. Under light microscopy, close contact between the bone and the hydroxyapatite coating was seen, with no gaps at the interface. In some areas of the coating a reduction of the coating thickness could be observed, along with the presence of some detached hydroxyapatite particles embedded in newly formed bone. The chemical analysis of the cathodoluminescent areas at the interface showed a reduced calcium:phosphorus ratio in this region.
引用
收藏
页码:233 / 238
页数:6
相关论文
共 50 条
  • [1] Histologic evaluation of hydroxyapatite-coated root-form implants retrieved after 7 years in function: A case report
    Proussaefs, PT
    Tatakis, DN
    Lozada, J
    Caplanis, N
    Rohrer, MD
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2000, 15 (03) : 438 - 443
  • [2] Postmortem histologic evaluation of mandibular titanium and maxillary hydroxyapatite-coated implants from 1 patient
    Rohrer, MD
    Sobczak, RR
    Prasad, HS
    Morris, HF
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1999, 14 (04) : 579 - 586
  • [3] Histomorphometric Analysis of Hydroxyapatite-Coated Implants in Rabbit Cortical Bone Using Longitudinal and Transverse Histologic Sections
    Rebollal, Julio
    Soares, Gloria de Almeida
    Vidigal, Guaracilei Maciel, Jr.
    IMPLANT DENTISTRY, 2010, 19 (02) : 137 - 144
  • [4] Melatonin improves the osseointegration of hydroxyapatite-coated titanium implants in senile female rats
    Sun, Tao
    Li, Jian
    Xing, Hai-Lin
    Tao, Zhou-Shan
    Yang, Min
    ZEITSCHRIFT FUR GERONTOLOGIE UND GERIATRIE, 2020, 53 (08): : 770 - 777
  • [5] Effects of guided bone regeneration around commercially pure titanium and hydroxyapatite-coated dental implants .2. Histologic analysis
    Stentz, WC
    Mealey, BL
    Gunsolley, JC
    Waldrop, TC
    JOURNAL OF PERIODONTOLOGY, 1997, 68 (10) : 933 - 949
  • [6] Clinical and histologic evaluation of submerged and nonsubmerged hydroxyapatite-coated implants:: A preliminary study in dogs
    Kohal, RJ
    De LaRosa, M
    Patrick, D
    Hürzeler, MB
    Caffesse, RG
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1999, 14 (06) : 824 - 834
  • [7] Bone properties surrounding hydroxyapatite-coated custom osseous integrated dental implants
    Baker, M. I.
    Eberhardt, A. W.
    Martin, D. M.
    McGwin, G.
    Lemons, J. E.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2010, 95B (01) : 218 - 224
  • [8] Human histologic and histomorphometric analyses of hydroxyapatite-coated implants after 10 years of function: A case report
    Trisi, P
    Keith, DJ
    Rocco, S
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2005, 20 (01) : 124 - 130
  • [9] Analysis of retrieved hydroxyapatite-coated hip prostheses
    Gross, KA
    Walsh, W
    Swarts, E
    JOURNAL OF THERMAL SPRAY TECHNOLOGY, 2004, 13 (02) : 190 - 199
  • [10] Analysis of retrieved hydroxyapatite-coated hip prostheses
    K. A. Gross
    W. Walsh
    E. Swarts
    Journal of Thermal Spray Technology, 2004, 13 : 190 - 199