Improving Low -contrast Detectability and Noise Texture Pattern for Computed Tomography Using Iterative Reconstruction Accelerated with Machine Learning Method: A Phantom Study

被引:6
作者
Funama, Yoshinori [1 ]
Takahashi, Hisashi [2 ,4 ]
Goto, Taiga [2 ,4 ]
Aoki, Yuko [2 ,4 ]
Yoshida, Ryo [2 ]
Kumagai, Yukio [2 ,4 ]
Awai, Kazuo [3 ]
机构
[1] Kumamoto Univ, Fac Life Sci, Dept Med Radiat Sci, 4-24-1 Kuhonji, Kumamoto 8620976, Japan
[2] Hitachi Ltd, Healthcare Business Unit, Kashiwa, Chiba, Japan
[3] Hiroshima Univ, Grad Sch Biomed Sci, Dept Diagnost Radiol, Hiroshima, Japan
[4] HITACHI Ltd, Kashiwa, Chiba, Japan
关键词
RADIATION-DOSE REDUCTION; FILTERED BACK-PROJECTION; MODERN DIAGNOSTIC MDCT; MODEL RECONSTRUCTION; IMAGE QUALITY; POWER SPECTRUM; CT ANGIOGRAPHY; ABDOMINAL CT; OPTIMIZATION; HYBRID;
D O I
10.1016/j.acra.2019.09.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To evaluate the performance of iterative reconstruction (IR) and filtered back projection (FBP) images in terms of low-contrast detectability at different radiation doses, IR levels, and slice thickness using the mathematical model observer with a focus on low-contrast detectability. Materials and Methods: The CCT189 MITA CT IQ Low-Contrast Phantom was used and helical scans were performed using a 64-detector CT scanner. Tube voltage was set at 120 kVp and tube current was adjusted from 45 to 600 mA. Images were reconstructed at slice thicknesses of 0.625 and 5.0 mm with FBP and five types of iterative progressive reconstruction with visual modeling (IPV) algorithms. The noise power spectrum (NPS) and normalized NPS were calculated. To evaluate low-contrast detectability, the model observer with the channelized Hotelling observer model was applied using low-contrast modules in the phantom. Results: The NPS and normalized NPS for IPV images had similar curves as that for FBP images. At a slice thickness of 0.625 mm and equivalent radiation dose level, the mean improvement of low-contrast detectability for IPV images was 1.19–2.15-fold greater than FBP images with corresponding noise reduction levels. At equivalent noise levels of 5.0–8.0 HU, low-contrast detectability of the IPVstd2 to IPVstr2 images as almost the same or better than that of the FBP images. However, the detectability of the IPVstr4 image was lower than that of the FBP image (p = 0.02). Conclusion: Low-contrast detectability of the IPV images was improved with a similar normalized NPS as with FBP images. Furthermore, a radiation reduction of >50% was achieved for the IPV images, while maintaining similar low-contrast detectability. © 2019 The Association of University Radiologists
引用
收藏
页码:929 / 936
页数:8
相关论文
共 27 条
  • [1] Barrett HH., 1998, STABILIZED ESTIMATES
  • [2] Application of the noise power spectrum in modern diagnostic MDCT: Part II. Noise power spectra and signal to noise
    Boedeker, K. L.
    McNitt-Gray, M. F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) : 4047 - 4061
  • [3] Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta
    Boedeker, K. L.
    Cooper, V. N.
    McNitt-Gray, M. F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) : 4027 - 4046
  • [4] Simulated 50 % radiation dose reduction in coronary CT angiography using adaptive iterative dose reduction in three-dimensions (AIDR3D)
    Chen, Marcus Y.
    Steigner, Michael L.
    Leung, Steve W.
    Kumamaru, Kanako K.
    Schultz, Kurt
    Mather, Richard T.
    Arai, Andrew E.
    Rybicki, Frank J.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (05) : 1167 - 1175
  • [5] Advanced Modeled Iterative Reconstruction (ADMIRE) Facilitates Radiation Dose Reduction in Abdominal CT
    Ellmann, Stephan
    Kammerer, Ferdinand
    Allmendinger, Thomas
    Hammon, Matthias
    Janka, Rolf
    Lell, Michael
    Uder, Michael
    Kramer, Manuel
    [J]. ACADEMIC RADIOLOGY, 2018, 25 (10) : 1277 - 1284
  • [6] Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages
    Euler, Andre
    Stieltjes, Bram
    Szucs-Farkas, Zsolt
    Eichenberger, Reto
    Reisinger, Clemens
    Hirschmann, Anna
    Zaehringer, Caroline
    Kircher, Achim
    Streif, Matthias
    Bucher, Sabine
    Buergler, David
    D'Errico, Luigia
    Kopp, Sebastien
    Wilhelm, Markus
    Schindera, Sebastian T.
    [J]. EUROPEAN RADIOLOGY, 2017, 27 (12) : 5252 - 5259
  • [7] Ferrero A, 2017, P SPIE INT SOC OPT E
  • [8] Observer Performance with Varying Radiation Dose and Reconstruction Methods for Detection of Hepatic Metastases
    Fletcher, Joel G.
    Fidler, Jeff L.
    Venkatesh, Sudhakar K.
    Hough, David M.
    Takahashi, Naoki
    Yu, Lifeng
    Johnson, Matthew
    Leng, Shuai
    Holmes, David R., III
    Carter, Rickey
    McCollough, Cynthia H.
    [J]. RADIOLOGY, 2018, 289 (02) : 455 - 464
  • [9] Image quality assessment of an iterative reconstruction algorithm applied to abdominal CT imaging
    Funama, Yoshinori
    Taguchi, Katsuyuki
    Utsunomiya, Daisuke
    Oda, Seitaro
    Katahira, Kazuhiro
    Tokuyasu, Shinichi
    Yamashita, Yasuyuki
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (04): : 527 - 534
  • [10] CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT
    Gervaise, Alban
    Osemont, Benoit
    Lecocq, Sophie
    Noel, Alain
    Micard, Emilien
    Felblinger, Jacques
    Blum, Alain
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (02) : 295 - 301