A Cube-Based Dual-GPU List-Mode Reconstruction Algorithm for PET Imaging

被引:4
作者
Teimoorisichani, Mohammadreza [1 ]
Goertzen, Andrew L. [2 ,3 ,4 ]
机构
[1] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
[2] Univ Manitoba, Dept Radiol, Winnipeg, MB R3E 3P4, Canada
[3] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3E 3P4, Canada
[4] Hlth Sci Ctr, Sect Nucl Med, Winnipeg, MB R3E 3P4, Canada
关键词
Graphics processing unit (GPU); list-mode maximum-likelihood expectation-maximization (ML-EM); positron emission tomography (PET); tomographic reconstruction; EM ALGORITHM; ACCURATE; DETECTOR;
D O I
10.1109/TRPMS.2021.3077012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Graphics processing units (GPUs) have been a popular choice for the acceleration of iterative tomographic reconstruction algorithms. In this work, two dual-GPU approaches to iterative list-mode reconstruction of PET data are studied. The developed techniques split the image into equal-size cubes and distribute them between the GPUs. Benefiting from caching a plane of the image into the fast shared memory, line-driven forward and back projection kernels are implemented incorporating a shift-invariant point spread function (PSF). In the first approach, lines were individually convolved with the PSF kernel whereas the second approach convolves the reconstructed image with the PSF before the forward projection and after the back projection. The algorithms were implemented using two CUDA-compatible devices with no CPU/GPU data transfer during the reconstruction process. List-mode data obtained from Monte Carlo simulations were used to evaluate the methods. One complete iteration of 10(7) events on an image of 220 x 220 x 110 voxels is done in less than 510 ms for a non-PSF reconstruction and in 520 to 5220 ms for a PSF reconstruction depending on the PSF and the reconstruction parameters. Dual-GPU memory layout, implementation details, sample images, and performance metrics of the developed algorithms are discussed.
引用
收藏
页码:463 / 474
页数:12
相关论文
共 59 条
  • [51] Strong Vivian E, 2009, Ann Surg Innov Res, V3, P2, DOI 10.1186/1750-1164-3-2
  • [52] Impact of image-space resolution modeling for studies with the high-resolution research tomograph
    Sureau, Florent C.
    Reader, Andrew J.
    Comtat, Claude
    Leroy, Claire
    Ribeiro, Maria-Joao
    Buvat, Irene
    Trebossen, Regine
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (06) : 1000 - 1008
  • [53] Multiple Importance Sampling for PET
    Szirmay-Kalos, Laszlo
    Magdics, Milan
    Toth, Balazs
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (04) : 970 - 978
  • [54] Geometry Optimization of a Dual-Layer Offset Detector for Use in Simultaneous PET/MR Neuroimaging
    Teimoorisichani, Mohammadreza
    Goertzen, Andrew L.
    [J]. IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2019, 3 (03) : 275 - 284
  • [55] A study of inter-crystal scatter in dual-layer offset scintillator arrays for brain-dedicated PET scanners
    Teimoorisichani, Mohammadreza
    Goertzen, Andrew L.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11)
  • [56] Teimoorisichani M, 2016, IEEE NUCL SCI CONF R
  • [57] Torres F, 2011, IEEE NUCL SCI S
  • [58] GPU Accelerated Stochastic Origin Ensemble Method With List-Mode Data for Compton Camera Imaging in Proton Therapy
    Zheng, Anran
    Yao, Zhiyang
    Xiao, Yongshun
    [J]. IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (02) : 243 - 252
  • [59] Fast and efficient fully 3D PET image reconstruction using sparse system matrix factorization with GPU acceleration
    Zhou, Jian
    Qi, Jinyi
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (20) : 6739 - 6757