A new less toxic polymer gel dosimeter: Radiological characteristics and dosimetry properties

被引:24
作者
Abtahi, Seyed Mohammad Mahdi [1 ]
Pourghanbari, Mohammad [2 ]
机构
[1] Imam Khomeini Int Univ, Dept Phys, Qazvin, Iran
[2] Kermanshah Univ Med Sci, Imam Reza Hosp, Med Imaging Ctr, Kermanshah, Iran
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2018年 / 53卷
关键词
Polymer gel dosimeter; PAMPSGAT; MRI; Radiological characteristics; RADIATION-DOSE DISTRIBUTIONS; RESOLUTION; NMR; OPTIMIZATION; PARAMETERS; SCANNER; NUMBER;
D O I
10.1016/j.ejmp.2018.08.018
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A new polymer gel dosimeter recipe was investigated that may be more suitable for widespread applications than polyacrylamide gel dosimeters, since the extremely toxic acrylamide has been replaced with the less harmful monomer 2-Acrylamido 2-Methyl Propane Sulfonic acid (AMPS). Methods: The new formulation was named PAMPSGAT. The MRI response (R2) of the dosimeters was analyzed for conditions of varying dose, dose rate, and temperature during scanning. Radiological properties of the PAMPSGAT polymer gel dosimeter were investigated. Results: The dose-response (R2) of AMPS/Bis appears to be linear over a dose range 10-40 Gy. The percentage of difference between the R2 values for imaging at 15 degrees C and MRI room temperature is about 4.6% for vial with 40 Gy absorbed dose which decreased to less than 1% for imaging at 20 degrees C. The percentage difference of Zeff of PAMPSGAT gel and soft tissue was less than 1% in the practical energy range (100 KeV-100 MeV). The electron density of the PAMPSGAT polymer gel was 2.9% higher than that of muscle. Results showed that the sensitivity of PAMPSGAT polymer gel dosimeter irradiated by Co-60 (energy = 1.25 MeV) is about 27.7% higher than that of irradiated using a 6 MeV Linac system. Conclusions: Temperature during MRI scanning has a small effect on the R2 response of the PAMPSGAT polymer gel dosimeter. Results confirmed tissue equivalency of the PAMPSGAT polymer gel dosimeter in most practical energy range. The PAMPSGAT polymer gel dosimeter response depends on energy and dose rate.
引用
收藏
页码:137 / 144
页数:8
相关论文
共 47 条
[1]   An investigation into the effect of magnetic resonance imaging (MRI) echo time spacing and number of echoes on the sensitivity and dose resolution of PAGATUG polymer-gel dosimeter [J].
Abtahi, S. M. ;
Khalilabadi, R. Jafari ;
Aftabi, S. .
INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2017, 15 (02) :185-196
[2]   Characteristics of a novel polymer gel dosimeter formula for MRI scanning: Dosimetry, toxicity and temporal stability of response [J].
Abtahi, S. M. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (09) :1156-1161
[3]   Optical and MRI investigations of an optimized acrylamide-based polymer gel dosimeter [J].
Abtahi, S. M. ;
Aghamiri, S. M. R. ;
Khalafi, H. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2014, 300 (01) :287-301
[4]  
[Anonymous], 1981, Heat and Thermodynamics
[5]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[6]   Dose resolution in radiotherapy polymer gel dosimetry:: effect of echo spacing in MRI pulse sequence [J].
Baldock, C ;
Lepage, M ;
Bäck, SÅJ ;
Murry, PJ ;
Jayasekera, PM ;
Porter, D ;
Kron, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02) :449-460
[7]   Development of a 3D optical scanner for evaluating patient-specific dose distributions [J].
Chang, Kyung Hwan ;
Lee, Suk ;
Jung, Hong ;
Choo, Yeon-Wook ;
Cao, Yuan Jie ;
Shim, Jang Bo ;
Kim, Kwang Hyeon ;
Lee, Nam Kwon ;
Park, Young Je ;
Kim, Chul Yong ;
Cho, Sam Ju ;
Lee, Sang Hoon ;
Min, Chul Kee ;
Kim, Woo Chul ;
Cho, Kwang Hwan ;
Huh, Hyun Do ;
Lim, Sangwook .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (05) :553-559
[8]   A systematic approach to determine optimal composition of gel used in radiation therapy [J].
Chang, Yuan-Jen ;
Hsieh, Bor-Tsung ;
Liang, Ji-An .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01) :783-785
[9]   A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis [J].
Cheng, Kai-Yuan ;
Hsieh, Ling-Ling ;
Shih, Cheng-Ting .
PLOS ONE, 2016, 11 (05)
[10]   Introducing gel dosimetry in a clinical environment: Customization of polymer gel composition and magnetic resonance imaging parameters used for 3D dose verifications in radiosurgery and intensity modulated radiotherapy [J].
Crescenti, Remo A. ;
Scheib, Stefan G. ;
Schneider, Uwe ;
Gianolini, Stefano .
MEDICAL PHYSICS, 2007, 34 (04) :1286-1297