A Monte Carlo investigation of cumulative dose measurements for cone beam computed tomography (CBCT) dosimetry

被引:28
|
作者
Abuhaimed, Abdullah [1 ,2 ,4 ]
Martin, Colin J. [2 ]
Sankaralingam, Marimuthu [1 ]
Gentle, David J. [3 ]
机构
[1] Beatson West Scotland Canc Ctr, Dept Clin Phys & Bioengn, Glasgow, Lanark, Scotland
[2] Univ Glasgow, Dept Clin Phys, Glasgow, Lanark, Scotland
[3] Gartnavel Royal Hosp, Dept Clin Phys & Bioengn, Glasgow G12 0YN, Lanark, Scotland
[4] King Abdulaziz City Sci & Technol, Dept Appl Phys, Riyadh, Saudi Arabia
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 04期
关键词
cumulative dose; CBCT; CTDI; EGSnrc/BEAMnrc; EGSnrc/DOSXYZnrc; AAPM TG-111; IMAGE QUALITY; EFFICIENCY IMPROVEMENTS; CT DOSIMETRY; RADIATION; RADIOTHERAPY; EGSNRC; SIMULATIONS; PROFILES; SYSTEM; INDEX;
D O I
10.1088/0031-9155/60/4/1519
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Many studies have shown that the computed tomography dose index (CTDI100) which is considered as a main dose descriptor for CT dosimetry fails to provide a realistic reflection of the dose involved in cone beam computed tomography (CBCT) scans. Several practical approaches have been proposed to overcome drawbacks of the CTDI100. One of these is the cumulative dose concept. The purpose of this study was to investigate four different approaches based on the cumulative dose concept: the cumulative dose (1) f(0,150) and (2) f(0,8) with a small ionization chamber 20 mm long, and the cumulative dose (3) f(100)(150) and (4) f(100)(infinity) with a standard 100 mm pencil ionization chamber. The study also aimed to investigate the influence of using the 20 and 100 mm chambers and the standard and the infinitely long phantoms on cumulative dose measurements. Monte Carlo EGSnrc/BEAMnrc and EGSnrc/DOSXYZnrc codes were used to simulate a kV imaging system integrated with a TrueBeam linear accelerator and to calculate doses within cylindrical head and body PMMA phantoms with diameters of 16 cm and 32 cm, respectively, and lengths of 150, 600, 900 mm. f(0,150) and f(100)(150) approaches were studied within the standard PMMA phantoms (150 mm), while the other approaches f(0,8) and f(100)(8) were within infinitely long head (600 mm) and body (900 mm) phantoms. CTDI infinity values were used as a standard to compare the dose values for the approaches studied at the centre and periphery of the phantoms and for the weighted values. Four scanning protocols and beams of width 20-300 mm were used. It has been shown that the f(0,infinity) approach gave the highest dose values which were comparable to CTDI infinity values for wide beams. The differences between the weighted dose values obtained with the 20 and 100 mm chambers were significant for the beam widths <120 mm, but these differences declined with increasing beam widths to be within 4%. The weighted dose values calculated within the infinitely long phantoms with both the chambers for the beam widths <= 140 were within 3% of those within the standard phantoms, but the differences rose to be within 15% at wider beams. By comparing the approaches studied in this investigation with other methodologies taking into account the efficiency of the approach as a dose descriptor and the simplicity of the implementation in the clinical environment, the f(0,150) method may be the best for CBCT dosimetry combined with the use of correction factors.
引用
收藏
页码:1519 / 1542
页数:24
相关论文
共 50 条
  • [31] Radiation doses in cone-beam breast computed tomography: A Monte Carlo simulation study
    Yi, Ying
    Lai, Chao-Jen
    Han, Tao
    Zhong, Yuncheng
    Shen, Youtao
    Liu, Xinming
    Ge, Shuaiping
    You, Zhicheng
    Wang, Tianpeng
    Shaw, Chris C.
    MEDICAL PHYSICS, 2011, 38 (02) : 589 - 597
  • [32] Experiences with image quality and radiation dose of cone beam computed tomography (CBCT) and multidetector computed tomography (MDCT) in pediatric extremity trauma
    Tschauner, Sebastian
    Marterer, Robert
    Nagy, Eszter
    Singer, Georg
    Riccabona, Michael
    Sorantin, Erich
    SKELETAL RADIOLOGY, 2020, 49 (12) : 1939 - 1949
  • [33] Experiences with image quality and radiation dose of cone beam computed tomography (CBCT) and multidetector computed tomography (MDCT) in pediatric extremity trauma
    Sebastian Tschauner
    Robert Marterer
    Eszter Nagy
    Georg Singer
    Michael Riccabona
    Erich Sorantin
    Skeletal Radiology, 2020, 49 : 1939 - 1949
  • [34] A Method to Convert Cone-Beam Computed Tomography (CBCT) Image for Dose Calculation and the Phantom Evaluation
    Zhang, Guang-shun
    Huang, Shao-min
    Chen, Cui
    Zhang, Dan-dan
    Deng, Xiao-wu
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 372 - 372
  • [35] A system to track skin dose for neuro-interventional cone-beam computed tomography (CBCT)
    Vijayan, Sarath
    Xiong, Zhenyu
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [36] Monte Carlo dose index estimation in computed tomography
    Ati, Moncef
    Bouamrane, Rachid
    Addi, Djamel
    Maroc, Fatima Zohra
    Mecheret, Fatima Zohra
    AIMS BIOENGINEERING, 2020, 7 (04): : 224 - 235
  • [37] Cone-beam computed tomography (CBCT) dose optimization technique and image quality assessment scoring
    Trivedi, Gaurav
    Singh, Pushpendra P.
    Oinam, Arun S.
    Singh, Ranjit
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2024, 20 (01) : 71 - 78
  • [38] Evaluation of enamel pearls by cone-beam computed tomography (CBCT)
    Akgul, Nilgun
    Caglayan, Fatma
    Durna, Nurhan
    Sumbullu, Muhammed-Akif
    Akgul, Hayati-Murat
    Durna, Dogan
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2012, 17 (02): : E218 - E222
  • [39] Cone Beam Computed Tomography (CBCT) in the Field of Interventional Oncology of the Liver
    Bapst, Blanche
    Lagadec, Matthieu
    Breguet, Romain
    Vilgrain, Valerie
    Ronot, Maxime
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2016, 39 (01) : 8 - 20
  • [40] Cone Beam Computed Tomography (CBCT) in the Field of Interventional Oncology of the Liver
    Blanche Bapst
    Matthieu Lagadec
    Romain Breguet
    Valérie Vilgrain
    Maxime Ronot
    CardioVascular and Interventional Radiology, 2016, 39 : 8 - 20