Comparing the fracture limits of the proximal femur under impact and quasi-static conditions in simulation of a sideways fall

被引:12
作者
Jazinizadeh, Fatemeh [1 ]
Mohammadi, Hojjat [2 ]
Quenneville, Cheryl E. [1 ,3 ]
机构
[1] McMaster Univ, Dept Mech Engn, Hamilton, ON, Canada
[2] McMaster Univ, Dept Civil Engn, Hamilton, ON, Canada
[3] McMaster Univ, Sch Biomed Engn, Hamilton, ON, Canada
关键词
Proximal femur; Fracture risk; Sideways fall; Impact loading; Quasi-static loading; Rate; Hip; FINITE-ELEMENT MODELS; BONE-MINERAL DENSITY; HIP FRACTURE; LOADING RATE; PREDICTION; BEHAVIOR; STRAIN; STRENGTH; RISK; VALIDATION;
D O I
10.1016/j.jmbbm.2019.103593
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Sideways falls onto the hip are responsible for a great number of fractures in older adults. One of the possible ways to prevent these fractures is through early identification of people at greatest risk so that preventive measures can be properly implemented. Many numerical techniques that are designed to predict the femur fracture risk are validated through performing quasi-static (QS) mechanical tests on isolated cadaveric femurs, whereas the real hip fracture is a result of an impact (IM) incident. The goal of this study was to compare the fracture limits of the proximal femur under IM and QS conditions in the simulation of a sideways fall to identify any possible relationship between them. Eight pairs of fresh frozen cadaveric femurs were divided into two groups of QS and IM (left and right randomized). All femurs were scanned with a Hologic DXA scanner and then cut and potted in a cylindrical tube. To measure the stiffness in two conditions of the single-leg stance (SLS) and sideways fall (SWF), non-destructive tests at a QS displacement rate were performed on the two groups. For the destructive tests, the QS group was tested in SWF configuration with the rate of 0.017 mm/s using a material testing machine, and the IM group was tested in the same configuration inside a pneumatic IM device with the projectile target displacement rate of 3 m/s. One of the IM specimens was excluded due to multiple strikes. The result of this study showed that there were no significant differences in the SLS and SWF stiffnesses between the two groups (P = 0.15 and P = 0.64, respectively). The destructive test results indicated that there was a significant difference in the fracture loads of the two groups (P < 0.00001) with the impact ones being higher; however, they were moderately correlated (R-2 = 0.45). Also, the comparison of the fracture location showed a qualitatively good agreement between the two groups. Using the relationship developed herein, results from another study were extrapolated with errors of less than 12%, showing that meaningful predictions for the impact scenario can be made based on the quasi-static tests. The result of this study suggests that there is a potential to replace IM tests with QS displacement rate tests, and this will provide important information that can be used for future studies evaluating clinical factors related to fracture risk.
引用
收藏
页数:8
相关论文
共 42 条
  • [1] Comparison of explicit finite element and mechanical simulation of the proximal femur during dynamic drop-tower testing
    Ariza, O.
    Gilchrist, S.
    Widmer, R. P.
    Guy, P.
    Ferguson, S. J.
    Cripton, P. A.
    Helgason, B.
    [J]. JOURNAL OF BIOMECHANICS, 2015, 48 (02) : 224 - 232
  • [2] Effects of loading rate on the of mechanical behavior of the femur in falling condition
    Askarinejad, Sina
    Johnson, Joshua E.
    Rahbar, Nima
    Troy, Karen L.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 96 : 269 - 278
  • [3] Prediction of strength and strain of the proximal femur by a CT-based finite element method
    Bessho, Masahiko
    Ohnishi, Isao
    Matsuyama, Juntaro
    Matsumoto, Takuya
    Imai, Kazuhiro
    Nakamura, Kozo
    [J]. JOURNAL OF BIOMECHANICS, 2007, 40 (08) : 1745 - 1753
  • [4] Energy absorption during impact on the proximal femur is affected by body mass index and flooring surface
    Bhan, Shivam
    Levine, Iris C.
    Laing, Andrew C.
    [J]. JOURNAL OF BIOMECHANICS, 2014, 47 (10) : 2391 - 2397
  • [5] Prediction of the fracture load of whole proximal femur specimens by topological analysis of the mineral distribution in DXA-scan images
    Boehm, Holger F.
    Horng, Annie
    Notohamiprodjo, Mike
    Eckstein, Felix
    Burklein, Dominik
    Panteleon, Alexandra
    Lutz, Juergen
    Reiser, Maximilian
    [J]. BONE, 2008, 43 (05) : 826 - 831
  • [6] COMPRESSIVE BEHAVIOR OF BONE AS A 2-PHASE POROUS STRUCTURE
    CARTER, DR
    HAYES, WC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (07) : 954 - 962
  • [7] The Injury Tolerance of the Tibia Under Off-Axis Impact Loading
    Chakravarty, Avery B.
    Martinez, Alberto A.
    Quenneville, Cheryl E.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2017, 45 (06) : 1534 - 1542
  • [8] Prediction of Femoral Fracture Load: Cross-sectional Study of Texture Analysis and Geometric Measurements on Plain Radiographs versus Bone Mineral Density
    Chappard, Christine
    Bousson, Valerie
    Bergot, Catherine
    Mitton, David
    Marchadier, Arnaud
    Moser, Thomas
    Benhamou, C. Laurent
    Laredo, Jean-Denis
    [J]. RADIOLOGY, 2010, 255 (02) : 536 - 543
  • [9] EFFECTS OF LOADING RATE ON STRENGTH OF THE PROXIMAL FEMUR
    COURTNEY, AC
    WACHTEL, EF
    MYERS, ER
    HAYES, WC
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1994, 55 (01) : 53 - 58
  • [10] Experimental validation of DXA-based finite element models for prediction of femoral strength
    Dall'Ara, E.
    Eastell, R.
    Viceconti, M.
    Pahr, D.
    Yang, L.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 63 : 17 - 25