Dual-probe radiofrequency ablation - An in vitro experimental study in bovine liver

被引:16
作者
Lee, JM
Rhim, H
Han, JK
Youn, BJ
Kim, SH
Choi, BI
机构
[1] Seoul Natl Univ Hosp, Dept Diagnost Radiol, Clin Res Inst, Seoul 110744, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul 151, South Korea
[3] Seoul Natl Univ, Med Res Ctr, Inst Radiat Med, Seoul 151, South Korea
[4] Hanyang Univ, Coll Med, Dept Radiol, Seoul 133791, South Korea
关键词
liver; interventional procedures; radiofrequency ablation; experimental study;
D O I
10.1097/01.rli.0000105041.12347.4b
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To determine whether alternative radiofrequency (RF) application with dual probes has advantages over sequential or simultaneous RF application for creating larger coagulation necrosis. Materials and Methods: Using 2 17-gauge internally cooled electrodes and a 200-watt generator (CC-3 model, Radionics), RF energy was applied sequentially (group A, n = 20), simultaneously (group 13, n = 20), or alternately (group C, n = 20) to explanted bovine liver. After preliminary experiments with a probe distance of 2-5 cm, a distance of 2 cm and 3 cm were chosen for main experiments. Total duration of the RF application was 10 minutes. In the alternative mode, the energy was applied alternately to both RF electrodes in 2-second intervals. Changes occurring in tissue impedance, current, power output, and temperature at the midpoint between the 2 electrodes were measured. The technical parameters, temperatures, and dimension of the ablated area were compared in the 3 groups using analysis of variance or Kruskal-Wallis test. Results: In the groups B and C, the impedance was gradually decreased during RF application and the mean current flow of both groups was 1550 +/- 130 mA and 1375 +/- 175 mA, respectively. However, in the group A, the impedance was markedly increased during RF application, and the mean current was 940 +/- 484 mA (P < 0.05, between groups A and B, and groups A and Q. With 3-cm spacing, the alternative RF application created oval-shaped ablation zones with larger shortest axis diameter at the midpoint than either the sequential or simultaneous RF application: 12.0 +/- 7.1 mm in group A; 27.0 +/- 3.1 mm in group 13; 34.9 +/- 3.4 mm in group C (P < 0.05). With 2 cm spacing, the corresponding figures were 25.3 +/- 17.8 mm in group A; 34.9 7.8 mm in group 13; 41.5 +/- 1.8 mm in group C (P < 0.05): the differences between groups A and B, and between groups A and C were statistically significant (P < 0.05). With 3-cm spacing, the temperature at the midpoint between the 2 probes was higher in group C (102degreesC) than in either the group A (62degreesC) or the group B (78degreesC; P < 0.05). With 2-cm spacing, groups B and C showed higher temperature than group A (P < 0.05), but the difference between groups B and C was not significant (P > 0.05). Conclusion: The alternative RF application creates larger, more regular ablation zones than either the sequential or simultaneous RF application. This result suggests that use of an alternating RF power application will improve the results of RF ablation for the treatment of larger tumors.
引用
收藏
页码:89 / 96
页数:8
相关论文
共 32 条
[1]   Hepatic lesion ablation with bipolar saline-enhanced radiofrequency in the audible spectrum [J].
Burdío, F ;
Güemes, A ;
Burdío, JM ;
Castiella, T ;
De Gregorio, MA ;
Lozano, R ;
Livraghi, T .
ACADEMIC RADIOLOGY, 1999, 6 (11) :680-686
[2]   CT versus sonography for monitoring radiofrequency ablation in a porcine liver [J].
Cha, CH ;
Lee, FT ;
Gurney, JM ;
Markhardt, BK ;
Warner, TF ;
Kelcz, F ;
Mahvi, DM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (03) :705-711
[3]   Overlapping ablation using a coaxial radiofrequency electrode and multiple cannulae system:: Experimental study in ex-vivo bovine liver [J].
Choi, D ;
Lim, HK ;
Kim, MJ ;
Lee, J ;
Kim, SK ;
Kim, EY ;
Kim, S ;
Kim, SH .
KOREAN JOURNAL OF RADIOLOGY, 2003, 4 (02) :117-123
[4]   Laparoscopically guided bipolar radiofrequency ablation of areas of porcine liver [J].
Curley, SA ;
Davidson, BS ;
Fleming, RY ;
Izzo, F ;
Stephens, LC ;
Tinkey, P ;
Cromeens, D .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1997, 11 (07) :729-733
[5]   Radiofrequency ablation of 100 hepatic metastases with a mean follow-up of more than 1 year [J].
de Baere, T ;
Elias, D ;
Dromain, C ;
El Din, MG ;
Kuoch, V ;
Ducreux, M ;
Boige, V ;
Lassau, N ;
Marteau, V ;
Lasser, P ;
Roche, A .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (06) :1619-1625
[6]   Radiofrequency thermal ablation: Computer analysis of the size of the thermal injury created by overlapping ablations [J].
Dodd, GD ;
Frank, MS ;
Aribandi, M ;
Chopra, S ;
Chintapalli, KN .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (04) :777-782
[7]   Tumor ablation with radio-frequency energy [J].
Gazelle, GS ;
Goldberg, SN ;
Solbiati, L ;
Livraghi, T .
RADIOLOGY, 2000, 217 (03) :633-646
[8]  
GOLDBERG SN, 1995, ACAD RADIOL, V2, P670
[9]   Large-volume tissue ablation with radio frequency by using a clustered, internally cooled electrode technique: Laboratory and clinical experience in liver metastases [J].
Goldberg, SN ;
Solbiati, L ;
Hahn, PF ;
Cosman, E ;
Conrad, JE ;
Fogle, R ;
Gazelle, GS .
RADIOLOGY, 1998, 209 (02) :371-379
[10]   Thermal ablation therapy for focal malignancy: A unified approach to underlying principles, techniques, and diagnostic imaging guidance [J].
Goldberg, SN ;
Gazelle, GS ;
Mueller, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :323-331