Minimally invasive versus conventional fixation of tracer in robot-assisted pedicle screw insertion surgery: a randomized control trial

被引:8
作者
Li Yongqi [1 ]
Zhang Dehua [1 ]
Wu Hongzi [1 ]
Zhang Ke [1 ]
Yang Rui [1 ]
Fang Zhou [1 ]
Wang Shaobo [1 ]
Liao Yi [1 ]
机构
[1] Karamay Cent Hosp, Dept Orthoped, Karamay 834000, Peoples R China
关键词
Modified; Minimally invasive; Tracer fixation; Robot; Pedicle screw placement; PLACEMENT; ACCURACY; FLUOROSCOPY; SPINE; SYSTEM;
D O I
10.1186/s12891-020-03239-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: This study evaluated the minimal invasiveness, safety, and accuracy of robot-assisted pedicle screw placement procedure using a modified tracer fixation device. Methods: Patients were randomly assigned to conventional fixation group (25 patients) and modified fixation group (27 patients). Results: No baseline statistical difference was observed between the groups (P > 0.05). The length of unnecessary incision, amount of bleeding, and fixation duration for tracer fixation respectively were 6.08 +/- 1.02 mm, 1.46 +/- 0.84 ml, and 1.56 +/- 0.32 min in the modified fixation group and 40.28 +/- 8.52 mm, 12.02 +/- 2.24 ml, and 5.08 +/- 1.06 min in the conventional group. The difference between both groups was significant (P < 0.05). However, no significant difference between the two groups was observed in terms of the accuracy of pedicle screw placement (P > 0.05). Conclusions: The modified minimally invasive procedure for tracer fixation results in minimal trauma and is simple, reliable, and highly safe. Additionally, the procedure does not compromise the accuracy of pedicle screw placement. Thus, it has great clinical applicable value.
引用
收藏
页数:9
相关论文
共 50 条
[31]   Unilateral versus bilateral percutaneous pedicle screw fixation in minimally invasive transforaminal lumbar interbody fusion [J].
Choi, Un Yong ;
Park, Jeong Yoon ;
Kim, Kyung Hyun ;
Kuh, Sung Uk ;
Chin, Dong Kyu ;
Kim, Keun Su ;
Cho, Yong Eun .
NEUROSURGICAL FOCUS, 2013, 35 (02)
[32]   Comparison of robot versus fluoroscopy-assisted pedicle screw instrumentation in adolescent idiopathic scoliosis surgery: A retrospective study [J].
Hou, Canglong ;
Yang, Huan ;
Chen, Yu ;
Yang, Yilin ;
Zhang, Beichen ;
Chen, Kai ;
Li, Ming ;
Yang, Mingyuan ;
Chen, Kai .
FRONTIERS IN SURGERY, 2023, 9
[33]   Robot-assisted Percutaneous Transfacet Screw Fixation Supplementing Oblique Lateral Interbody Fusion Procedure: Accuracy and Safety Evaluation of This Novel Minimally Invasive Technique [J].
Wu, Jing-ye ;
Yuan, Qiang ;
Liu, Ya-jun ;
Sun, Yu-qing ;
Zhang, Yong ;
Tian, Wei .
ORTHOPAEDIC SURGERY, 2019, 11 (01) :25-33
[34]   Safety and risk factors of TINAVI robot-assisted percutaneous pedicle screw placement in spinal surgery [J].
Ren-Jie Zhang ;
Lu-Ping Zhou ;
Lai Zhang ;
Hua-Qing Zhang ;
Jian-Xiang Zhang ;
Cai-Liang Shen .
Journal of Orthopaedic Surgery and Research, 17
[35]   Robot-assisted versus laparoscopic minimally invasive surgery for the treatment of stage I endometrial cancer [J].
Argenta, Peter A. ;
Mattson, Jordan ;
Rivard, Colleen L. ;
Luther, Elizabeth ;
Schefter, Alexandra ;
Vogel, Rachel, I .
GYNECOLOGIC ONCOLOGY, 2022, 165 (02) :347-352
[36]   Effectiveness and safety of robot-assisted versus fluoroscopy-assisted pedicle screw implantation in scoliosis surgery: a systematic review and meta-analysis [J].
Wang, Xu ;
Li, Hao-xuan ;
Zhu, Qing-san ;
Zhu, Yu-hang .
NEUROSURGICAL REVIEW, 2024, 47 (01)
[37]   Robot-assisted and conventional freehand pedicle screw placement: a systematic review and meta-analysis of randomized controlled trials [J].
Gao, Shutao ;
Lv, Zhengtao ;
Fang, Huang .
EUROPEAN SPINE JOURNAL, 2018, 27 (04) :921-930
[38]   Robot-assisted Minimally Invasive Thoracolaparoscopic Esophagectomy Versus Open Transthoracic Esophagectomy for Resectable Esophageal Cancer A Randomized Controlled Trial [J].
van der Sluis, Pieter C. ;
van der Horst, Sylvia ;
May, Anne M. ;
Schippers, Carlo ;
Brosens, Lodewijk A. A. ;
Joore, Hans C. A. ;
Kroese, Christiaan C. ;
Mohammad, Nadia Haj ;
Mook, Stella ;
Vleggaar, Frank P. ;
Rinkes, Inne H. M. Borel ;
Ruurda, Jelle P. ;
van Hillegersberg, Richard .
ANNALS OF SURGERY, 2019, 269 (04) :621-630
[39]   Robot-assisted percutaneous pars-pedicle screw fixation for treating Hangman's fracture [J].
Zhao, Jingwei ;
Liu, Yajun ;
Zhang, Qi ;
Fan, Mingxing ;
Han, Xiaoguang ;
He, Da ;
Liu, Bo ;
Tian, Wei .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2023, 18 (01)
[40]   Robot-Assisted Minimally Invasive Surgery: First Report from the Caribbean [J].
Cawich, Shamir O. ;
Arulampalam, Tan ;
Senasi, Ramdas ;
Naraynsingh, Vijay .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (10)