Numerical Simulation of Fluid-Structure Interaction Under the Condition of Pulsatile Blood Flow of Renal Artery with Radiofrequency Electrode

被引:0
作者
Nan, Qun [1 ]
Cheng, Yanyan [1 ]
Tian, Zhen [1 ]
Dong, Tong [1 ]
Gao, Xiang [1 ]
机构
[1] Beijing Univ Technol, Coll Life Sci & Bioengn, Beijing, Peoples R China
来源
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1 | 2019年 / 68卷 / 01期
基金
北京市自然科学基金; 美国国家科学基金会;
关键词
Fluid-solid interaction; Numerical simulation; Pulsating blood flow; Hemodynamics; Radiofrequency electrode;
D O I
10.1007/978-981-10-9035-6_95
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To investigate the effect of Fluid-Structure interaction of the renal artery with radiofrequency electrode under the situation of pulsatile blood flow, the ideal renal artery model that contains electrode was constructed by Solid Works. The renal arterial bifurcation vessel model with electrode and non-containing electrode was selected for comparison. Comsol software was applied to mimic the pulsatile blood flow. Six key points was selected for the analysis of hemodynamic parameters. The distribution of velocity, pressure, the equivalent stress and the deformation of the vascular wall were obtained. Compared to the pure arterial model, the speed is reduced and the pressure and equivalent stress are increased at point F When containing electrode. However, the deformation of vessel walls caused by this is very small and can be neglected. And the velocity pressure, equivalent stress and deformation of point B are relatively small at the same time. which to some extent protects the blood vessels. In combination, the insertion of RF electrode does little harm to the arterial blood vessels.
引用
收藏
页码:515 / 518
页数:4
相关论文
共 9 条
  • [1] Fluid-structure interaction within realistic three-dimensional models of the aneurysmatic aorta as a guidance to assess the risk of rupture of the aneurysm
    Di Martino, ES
    Guadagni, G
    Fumero, A
    Ballerini, G
    Spirito, R
    Biglioli, P
    Redaelli, A
    [J]. MEDICAL ENGINEERING & PHYSICS, 2001, 23 (09) : 647 - 655
  • [2] Fluid, solid and fluid-structure interaction simulations on patient-based abdominal aortic aneurysm models
    Kelly, Sinead
    O'Rourke, Malachy
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2012, 226 (H4) : 288 - 304
  • [3] Liu Ying, 2015, Applied Mathematics and Mechanics (Chinese Edition), V36, P1058, DOI 10.3879/j.issn.1000-0887.2015.10.005
  • [4] Geometrical Enhancements for Abdominal Aortic Stent-Grafts
    Molony, David S.
    Callanan, Anthony
    Morris, Liam G.
    Doyle, Barry J.
    Walsh, Michael T.
    McGloughlin, Tim M.
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2008, 15 (05) : 518 - 529
  • [5] The critical role of hemodynamics in the development of cerebral vascular disease
    Nixon, Alexander M.
    Gunel, Murat
    Sumpio, Bauer E.
    [J]. JOURNAL OF NEUROSURGERY, 2010, 112 (06) : 1240 - 1253
  • [6] Unsteady and three-dimensional simulation of blood flow in the human aortic arch
    Shahcheraghi, N
    Dwyer, HA
    Cheer, AY
    Barakat, AI
    Rutaganira, T
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2002, 124 (04): : 378 - 387
  • [7] YANG J, 2010, BIOMEDICAL ENG CLIN, V14, P390
  • [8] YANG J, 2013, CHINESE J MED PHYS, V17, P1
  • [9] YANG J, 2014, BME CLIN MED, V18, P310