Enzyme sensor for simultaneous determination of cholesterol, triglyceride and HDL cholesterol in serum

被引:7
作者
Uemura, Takeshi [1 ,3 ]
Yamazaki, Hiroki [1 ]
Itoh, Tetsuji [2 ]
Nishizawa, Seiichi [3 ]
机构
[1] Techno Med Co Ltd, Engn Planning Off, Yokohama, Kanagawa, Japan
[2] Natl Inst Adv Ind Sci & Technol, Sendai, Miyagi, Japan
[3] Tohoku Univ, Grad Sch Sci, Dept Chem, Sendai, Miyagi, Japan
关键词
Enzyme sensor; Cholesterol; Triglyceride; HDL cholesterol; HIGH-DENSITY-LIPOPROTEINS; ELECTRODE; BIOSENSOR;
D O I
10.1007/s44211-022-00148-w
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A three-electrode lipid biosensor that simultaneously measures the total cholesterol (CHO), triglyceride (TG), and HDL cholesterol (HDL-C) in standard serum has been developed. The lipid biosensor is designed for clinical use where production cost, low sample requirement, portability, stability, and speed are high priorities. The device design filters out blood cells and lipoproteins from the serum, where the target molecules are catalyzed by enzymes encapsulated in carboxymethyl cellulose (CMC) to produce electron mediators, potassium ferrocyanide. These electron mediators were subsequently detected by amperometric determination. The sensor exhibit high selectivity towards the targets and can measure the target lipids in 4 min with 10 mu L of serum. Lastly, the device can be stored up to 18 months with a minimal decrease in catalytic efficiency.
引用
收藏
页码:1189 / 1197
页数:9
相关论文
共 17 条
[11]   ANTIOXIDATIVE ACTIVITY OF HIGH-DENSITY-LIPOPROTEINS INVIVO [J].
KLIMOV, AN ;
GUREVICH, VS ;
NIKIFOROVA, AA ;
SHATILINA, LV ;
KUZMIN, AA ;
PLAVINSKY, SL ;
TERYUKOVA, NP .
ATHEROSCLEROSIS, 1993, 100 (01) :13-18
[12]   Reverse cholesterol transport and cholesterol efflux in atherosclerosis [J].
Ohashi, R ;
Mu, H ;
Wang, X ;
Yao, Q ;
Chen, C .
QJM-AN INTERNATIONAL JOURNAL OF MEDICINE, 2005, 98 (12) :845-856
[13]   ENZYME ELECTRODE FOR FREE CHOLESTEROL [J].
SATOH, I ;
KARUBE, I ;
SUZUKI, S .
BIOTECHNOLOGY AND BIOENGINEERING, 1977, 19 (07) :1095-1099
[14]   Amperometric L-lactate biosensor based on screen-printed carbon electrode containing cobalt phthalocyanine, coated with lactate oxidase-mesoporous silica conjugate layer [J].
Shimomura, Takeshi ;
Sumiya, Touru ;
Ono, Masatoshi ;
Ito, Tetsuji ;
Hanaoka, Taka-aki .
ANALYTICA CHIMICA ACTA, 2012, 714 :114-120
[15]   Electrochemical biosensor for the detection of formaldehyde based on enzyme immobilization in mesoporous silica materials [J].
Shimomura, Takeshi ;
Itoh, Tetsuji ;
Sumiya, Touru ;
Mizukami, Fujio ;
Ono, Masatoshi .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 135 (01) :268-275
[16]   ENZYME ELECTRODE [J].
UPDIKE, SJ ;
HICKS, GP .
NATURE, 1967, 214 (5092) :986-&
[17]   The metabolic syndrome: A global public health problem and a new definition [J].
Zimmet, Paul ;
Magliano, Dianna ;
Matsuzawa, Yuji ;
Alberti, George ;
Shaw, Jonathan .
JOURNAL OF ATHEROSCLEROSIS AND THROMBOSIS, 2005, 12 (06) :295-300