An interlaboratory comparison of sample presentation methods for the analysis of aqueous solutions using Fourier transform infrared spectroscopy

被引:0
|
作者
Bartl, F
Delgadillo, I
Davies, AN
Huvenne, JP
Meurens, M
Volka, K
Wilson, RH
机构
[1] INST FOOD RES, INST PHYS CHEM, NORWICH NR4 7UA, NORFOLK, ENGLAND
[2] UNIV MUNICH, DEPT QUIM, D-80333 MUNICH, GERMANY
[3] UNIV AVEIRO, P-3800 AVEIRO, PORTUGAL
[4] INST SPEKTROCHEM, CNRS, D-44013 DORTMUND, GERMANY
[5] SPECTROCHIM INFRAROUGE & RAMAN LAB, F-59655 VILLENEUVE DASCQ, FRANCE
[6] UNIV CATHOLIQUE LOUVAIN, DEPT ANALYT CHEM, B-1348 LOUVAIN, BELGIUM
[7] INST CHEM TECHNOL, CR-16628 PRAGUE 6, CZECH REPUBLIC
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1996年 / 354卷 / 01期
关键词
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Fourier transform infrared (FTIR) spectroscopy is becoming much more widely used as an analytical technique. It appears to be particularly useful for quantitative analysis, especially of aqueous solutions. However, there is a considerable variation in the cost and quality of instrumentation as well as a large number of accessories for sample presentation employing attenuated total reflectance (ATR). The objectives of the work reported in this paper were to investigate the effect of the choice of instrumentation on the quality of quantitative analyses as well as the effect of variable acquisition parameters. A series of experiments was designed in which samples were distributed to a number of workers using a range of FTIR instruments from different manufacturers and of differing quality. The workers were allowed to use the accessory of their own choice and to carry out a quantitative analysis of glucose solutions using their normal operating parameters. In following experiments the effects of instrumental and purge were considered and quantitative analysis was performed using a set of prescribed acquisition conditions. The results showed that there were significant differences in the quality of the spectra produced on the different combinations. However, although each worker produced quite different linear calibrations (in terms of slope and intercept), they all predicted the concentration of an unknown with a similar accuracy. However,later experiments showed that the low-cost bench-top instruments were not always so successful when the peaks used in the analysis were affected by residual water vapour interferences. Instrumental drift was noticeably worse for the low-cost instruments but this was an effect that could easily be eliminated by suitable baseline correction.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 50 条
  • [41] Spectroscopic analysis of bladder cancer tissues using Fourier transform infrared spectroscopy
    Nafie A. Al-Muslet
    Essam E. Ali
    Journal of Applied Spectroscopy, 2012, 79 : 139 - 142
  • [42] Hydrocarbons in water: Analysis using on-line aqueous supercritical fluid extraction-Fourier transform infrared spectroscopy
    Minty, B
    Ramsey, ED
    Lewis, R
    ANALYTICAL COMMUNICATIONS, 1996, 33 (06): : 203 - 204
  • [43] Fourier-transform infrared spectroscopy for characterization of liquid protein solutions: A comparison of two sampling techniques
    Kafle, Bijay
    Wubshet, Sileshi Gizachew
    Dankel, Katinka
    Mage, Ingrid
    OFarrell, Marion
    Afseth, Nils Kristian
    VIBRATIONAL SPECTROSCOPY, 2023, 124
  • [44] Analysis of mebendazole polymorphs by Fourier transform infrared spectrometry using chemometric methods
    Bunaciu, AA
    Fleschin, S
    Aboul-Enein, HY
    SPECTROSCOPY LETTERS, 2001, 34 (05) : 527 - 536
  • [45] Measurements of urea and glucose in aqueous solutions with dual-beam near-infrared Fourier transform spectroscopy
    Jensen, PS
    Bak, J
    APPLIED SPECTROSCOPY, 2002, 56 (12) : 1593 - 1599
  • [46] SAMPLE ROTATION IN DIFFUSE REFLECTANCE FOURIER-TRANSFORM INFRARED-SPECTROSCOPY
    VELAPOLDI, RA
    TVEDT, JE
    CHRISTY, AA
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (06) : 1126 - 1128
  • [47] INFRARED REFLECTION-ABSORPTION SPECTROSCOPY OF SURFACE SPECIES - A COMPARISON OF FOURIER-TRANSFORM AND DISPERSION METHODS
    GOLDEN, WG
    SAPERSTEIN, DD
    SEVERSON, M
    OVEREND, J
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1983, 186 (AUG): : 112 - PHYS
  • [48] COMPARISON OF FOURIER-TRANSFORM INFRARED-PHOTOACOUSTIC SPECTROSCOPY (FTIR-PAS) AND CONVENTIONAL METHODS FOR ANALYSIS OF COAL OXIDATION
    CHIEN, PL
    MARKUSZEWSKI, R
    MCCLELLAND, JF
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1985, 189 (APR-): : 2 - FUEL
  • [49] Identification and discrimination of bacteria using Fourier transform infrared spectroscopy
    Maity, Jyoti Prakash
    Kar, Sandeep
    Lin, Chao-Ming
    Chen, Chen-Yen
    Chang, Young-Fo
    Jean, Jiin-Shuh
    Kulp, Thomas R.
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2013, 116 : 478 - 484
  • [50] Identification of species of Brucella using fourier transform infrared spectroscopy
    Gómez, MAM
    Pérez, MAB
    Gil, FJM
    Díez, AD
    Rodríguez, JFM
    Rodríguez, PG
    Domingo, AO
    Torres, AR
    JOURNAL OF MICROBIOLOGICAL METHODS, 2003, 55 (01) : 121 - 131