4D XCAT phantom for multimodality imaging research

被引:967
|
作者
Segars, W. P. [1 ]
Sturgeon, G. [1 ,2 ]
Mendonca, S. [1 ]
Grimes, Jason [1 ]
Tsui, B. M. W. [3 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Carl E Ravin Adv Imaging Labs, Durham, NC 27705 USA
[2] Univ N Carolina, Dept Biomed Engn, Chapel Hill, NC 27599 USA
[3] Johns Hopkins Med Inst, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21287 USA
基金
美国国家卫生研究院;
关键词
medical imaging simulation; computer phantom; SPECT; PET; CT; NCAT PHANTOM; SIMULATION; SPECT; MODELS; ATTENUATION; PERFORMANCE; CT;
D O I
10.1118/1.3480985
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors develop the 4D extended cardiac-torso (XCAT) phantom for multimodality imaging research. Methods: Highly detailed whole-body anatomies for the adult male and female were defined in the XCAT using nonuniform rational B-spline (NURBS) and subdivision surfaces based on segmentation of the Visible Male and Female anatomical datasets from the National Library of Medicine as well as patient datasets. Using the flexibility of these surfaces, the Visible Human anatomies were transformed to match body measurements and organ volumes for a 50th percentile (height and weight) male and female. The desired body measurements for the models were obtained using the PEOPLESIZE program that contains anthropometric dimensions categorized from 1st to the 99th percentile for US adults. The desired organ volumes were determined from ICRP Publication 89 [ICRP, "Basic anatomical and physiological data for use in radiological protection: reference values," ICRP Publication 89 (International Commission on Radiological Protection, New York, NY, 2002)]. The male and female anatomies serve as standard templates upon which anatomical variations may be modeled in the XCAT through user-defined parameters. Parametrized models for the cardiac and respiratory motions were also incorporated into the XCAT based on high-resolution cardiac-and respiratory-gated multislice CT data. To demonstrate the usefulness of the phantom, the authors show example simulation studies in PET, SPECT, and CT using publicly available simulation packages. Results: As demonstrated in the pilot studies, the 4D XCAT (which includes thousands of anatomical structures) can produce realistic imaging data when combined with accurate models of the imaging process. With the flexibility of the NURBS surface primitives, any number of different anatomies, cardiac or respiratory motions or patterns, and spatial resolutions can be simulated to perform imaging research. Conclusions: With the ability to produce realistic, predictive 3D and 4D imaging data from populations of normal and abnormal patients under various imaging parameters, the authors conclude that the XCAT provides an important tool in imaging research to evaluate and improve imaging devices and techniques. In the field of x-ray CT, the phantom may also provide the necessary foundation with which to optimize clinical CT applications in terms of image quality versus radiation dose, an area of research that is becoming more significant with the growing use of CT. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3480985]
引用
收藏
页码:4902 / 4915
页数:14
相关论文
共 50 条
  • [31] 4D MR imaging using robust internal respiratory signal
    Hui, CheukKai
    Wen, Zhifei
    Stemkens, Bjorn
    Tijssen, R. H. N.
    van den Berg, C. A. T.
    Hwang, Ken-Pin
    Beddar, Sam
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (09) : 3472 - 3487
  • [32] Development of the 4D Phantom for patient-specific, end-to-end radiation therapy QA
    Malinowski, K.
    Noel, C.
    Lu, W.
    Lechleiter, K.
    Hubenschmidt, J.
    Low, D.
    Parikh, P.
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [33] Radiolucent 4D Ultrasound Imaging: System Design and Application to Radiotherapy Guidance
    Schlosser, Jeffrey
    Hristov, Dimitre
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (10) : 2292 - 2300
  • [34] OPEN-SOURCE 4D STATISTICAL SHAPE MODEL OF THE HEART FOR X-RAY PROJECTION IMAGING
    Unberath, Mathias
    Maier, Andreas
    Fleischmann, Dominik
    Hornegger, Joachim
    Fahrig, Rebecca
    2015 IEEE 12th International Symposium on Biomedical Imaging (ISBI), 2015, : 739 - 742
  • [35] 4D deformable models with temporal constraints: application to 4D cardiac image segmentation
    Montagnat, J
    Delingette, H
    MEDICAL IMAGE ANALYSIS, 2005, 9 (01) : 87 - 100
  • [36] Quantifying the impact of respiratory-gated 4D CT acquisition on thoracic image quality: A digital phantom study
    Bernatowicz, K.
    Keall, P.
    Mishra, P.
    Knopf, A.
    Lomax, A.
    Kipritidis, J.
    MEDICAL PHYSICS, 2015, 42 (01) : 324 - 334
  • [37] 4D PET iterative deconvolution with spatiotemporal regularization for quantitative dynamic PET imaging
    Reilhac, Anthonin
    Charil, Arnaud
    Wimberley, Catriona
    Angelis, Georgios
    Hamze, Hasar
    Callaghan, Paul
    Garcia, Marie-Paule
    Boisson, Frederic
    Ryder, Will
    Meikle, Steven R.
    Gregoire, Marie-Claude
    NEUROIMAGE, 2015, 118 : 484 - 493
  • [38] Extension of the NCAT phantom for the investigation of intra-fraction respiratory motion in IMRT using 4D Monte Carlo
    McGurk, Ross
    Seco, Joao
    Riboldi, Marco
    Wolfgang, John
    Segars, Paul
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) : 1475 - 1490
  • [39] A proto-type design of a real-tissue phantom for the validation of deformation algorithms and 4D dose calculations
    Szegedi, M.
    Rassiah-Szegedi, P.
    Fullerton, G.
    Wang, B.
    Salter, B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (13) : 3685 - 3699
  • [40] Validation of the 4D NCAT simulation tools for use in high-resolution x-ray CT research
    Segars, WP
    Mahesh, M
    Beck, T
    Frey, EC
    Tsui, BMW
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 828 - 834