Different platform and tube geometries and atomization temperatures in graphite furnace atomic absorption spectrometry: Cadmium determination in whole blood as a case study

被引:15
作者
Moreira, Fatima R. [1 ]
Mello, Monique G. [1 ,2 ]
Campos, Reinaldo C. [2 ]
机构
[1] CESTEH ENSP Oswaldo Cruz Fdn, Toxicol Lab, BR-21041210 Rio De Janeiro, RJ, Brazil
[2] Pontificia Univ Catolica Rio de Janeiro, Dept Chem, BR-22453900 Rio De Janeiro, RJ, Brazil
关键词
cadmium; blood; interferences; graphite furnace atomic absorption spectrometry;
D O I
10.1016/j.sab.2007.09.002
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In the present work the performance of different platform and tube geometries and atomization temperatures in graphite furnace atomic absorption spectrometry was investigated, using the determination of Cd in whole blood as an example. Grooved, integrated and fork platforms as well as atomization temperatures between 1200 degrees C and 2200 degrees C were investigated in a longitudinally heated graphite atomizer and compared with the performance of a transversely heated furnace. In the longitudinally heated furnace the increase of the atomization temperature in the studied range resulted in an increase of matrix effects for all platform geometries. The integrated platform exhibited slightly lower sensitivity and increased multiplicative interferences in comparison to the other two platform designs. Interference-free Cd determination was possible with all types of platforms and 1200 degrees C as the atomization temperature as well as with grooved and fork platforms at 1700 degrees C. On the other hand, lower atomization temperatures resulted in poorer limits of detection, due to the longer integration time needed. No matrix effect was observed at any atomization temperature using the transversely heated atomizer; in addition, limits of detection were better than those observed with the longitudinally heated atomizer. Best values were around 0.02 mu g L-1 with the latter atomizer compared to values around 0.02 mu g L-1 with the former one. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:1273 / 1277
页数:5
相关论文
共 19 条
[1]  
Afridi HI, 2007, J AOAC INT, V90, P470
[2]  
ATSDR, 1998, TOX PROF CADM
[3]   Rapid determination of lead and cadmium in biological fluids by electrothermal atomic absorption spectrometry using Zeeman correction [J].
Campillo, N ;
Viñas, P ;
López-García, I ;
Hernández-Córdoba, M .
ANALYTICA CHIMICA ACTA, 1999, 390 (1-3) :207-215
[4]   Low-volume, high-sensitivity assay for cadmium in blood and urine using conventional atomic absorption spectrophotometry [J].
Cerny, EA ;
Bhattacharyya, MH .
ANALYTICAL BIOCHEMISTRY, 2003, 314 (02) :180-193
[5]  
Elinder C.G., 1994, BIOL MONITORING META, P44
[6]  
GARDINER WP, 1997, STAT ANAL METHODS CH, P168
[7]   Electrothermal atomic absorption spectrometric determination of cadmium and lead in blood using flow injection on-line sorption preconcentration in a knotted reactor [J].
Ivanova, E ;
Van Mol, W ;
Adams, F .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (6-8) :1041-1048
[8]   TRACE-ELEMENT REFERENCE VALUES IN TISSUES FROM INHABITANTS OF THE EUROPEAN COMMUNITY .1. A STUDY OF 46 ELEMENTS IN URINE, BLOOD AND SERUM OF ITALIAN SUBJECTS [J].
MINOIA, C ;
SABBIONI, E ;
APOSTOLI, P ;
PIETRA, R ;
POZZOLI, L ;
GALLORINI, M ;
NICOLAOU, G ;
ALESSIO, L ;
CAPODAGLIO, E .
SCIENCE OF THE TOTAL ENVIRONMENT, 1990, 95 :89-105
[9]  
MOREIRA MR, 1995, ANALYST, V120, P947
[10]   Determination of lead, cadmium and mercury in blood for assessment of environmental exposure: A comparison between inductively coupled plasma-mass spectrometry and atomic absorption spectrometry [J].
Palmer, Christopher D. ;
Lewis, Miles E., Jr. ;
Geraghty, Ciaran M. ;
Barbosa, Fernando, Jr. ;
Parsons, Patnck J. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (08) :980-990