Early pacemaker twiddler syndrome

被引:37
作者
Fahraeus, T [1 ]
Höijer, CJ [1 ]
机构
[1] Univ Lund Hosp, Pacemaker Dept, Heart & Lung Div, S-22185 Lund, Sweden
来源
EUROPACE | 2003年 / 5卷 / 03期
关键词
twiddler's syndrome; pacemaker lead dislodgment;
D O I
10.1016/S1099-5129(03)00032-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Twiddler's syndrome is a well-known complication of pacemaker treatment. It was first described by Bayliss et al. when a patient manipulated and rotated the pulse generator in the pocket so many turns that it resulted in lead dislodgment, diaphragmatic stimulation and loss of capture. In this case report we present a patient who managed to rotate her dual chamber pulse generator so quickly after implantation that exit block occurred within 17 h. She had wound the two leads as far as their tips in a perfect formation around the pulse generator. (C) 2003 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.