High-Pitch Coronary CT Angiography at 70 kVp Adopting a Protocol of Low Injection Speed and Low Volume of Contrast Medium

被引:21
作者
Feng, Ruiqi [1 ]
Tong, Jiajie [2 ]
Liu, Xiaofei [1 ]
Zhao, Yu [1 ]
Zhang, Liang [1 ]
机构
[1] China Med Univ, Affiliated Hosp 1, Dept Radiol, 155 Nanjingbei St, Shenyang 110001, Liaoning, Peoples R China
[2] Hebei Gen Hosp, Dept Radiol, Shijiazhuang 050051, Hebei, Peoples R China
关键词
Low tube voltage; Contrast medium; Coronary CT angiography; Iodine delivery rate; CT attenuation enhancement; Radiation reduction; Low dose contrast agent; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; DUAL-SOURCE CT; TUBE VOLTAGE; ITERATIVE-RECONSTRUCTION; IMAGE QUALITY; CARDIAC CT; RADIATION; FEASIBILITY; MULTICENTER; ACQUISITION;
D O I
10.3348/kjr.2017.18.5.763
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the feasibility and image quality (IQ) of prospectively high-pitch coronary CT angiography (CCTA) with low contrast medium injection rate at 70 kVp. Materials and Methods: One hundred and four patients with suspected coronary artery disease (body mass index < 26 kg/m(2), sinus rhythm and heart rate < 70 beats/min) were prospectively enrolled and randomly divided into two groups. In group A and group B, 28 mL and 40 mL of 370 mgI/mL iodinated contrast media was administrated at a flow rate of 3.5 and 5 mL/s, respectively. CT values, noise, signal-to-noise ratio, contrast-to-noise ratio (CNR) of the proximal segments of coronary arteries and subjective IQ were evaluated. Results: The CT values and noise in group A were significantly lower than those in group B (434-485 Hounsfield units [HU] vs. 772-851 HU, all p < 0.001; 17.8-22.3 vs. 23.3-26.4, all p < 0.005). The CNRs of the right coronary artery and left main artery showed no statistical difference between the two groups (42.1 +/- 13.8 vs. 36.8 +/- 16.0, p = 0.074; 38.7 +/- 10.6 vs. 38.1 +/- 17.0, p = 0.819). No statistical difference was observed between the two groups in IQ scores (3.04 +/- 0.75 vs. 3.0 +/- 0.79, p = 0.526) and diagnostic ratio (96.1% [50/52] vs. 94.2% [49/52], p = 0.647). Conclusion: Prospective high-pitch CCTA at 70 kVp with 28 mL of contrast media and injection rate of 3.5 mL/s could provide diagnostic IQ for normal-weight patients with heart rate of < 70 beats/min.
引用
收藏
页码:763 / 772
页数:10
相关论文
共 34 条
[1]   Influence of slice thickness and reconstruction kernel on the computed tomographic attenuation of coronary atherosclerotic plaque [J].
Achenbach, Stephan ;
Boehmer, Kerstin ;
Pflederer, Tobias ;
Ropers, Dieter ;
Seltmann, Martin ;
Lell, Michael ;
Anders, Katharina ;
Kuettner, Axel ;
Uder, Michael ;
Daniel, Werner G. ;
Marwan, Mohamed .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2010, 4 (02) :110-115
[2]   Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition [J].
Achenbach, Stephan ;
Marwan, Mohamed ;
Ropers, Dieter ;
Schepis, Tiziano ;
Pflederer, Tobias ;
Anders, Katharina ;
Kuettner, Axel ;
Daniel, Werner G. ;
Uder, Michael ;
Lell, Michael M. .
EUROPEAN HEART JOURNAL, 2010, 31 (03) :340-346
[3]  
Austen W G, 1975, Circulation, V51, P5
[4]  
Bongartz G, FIGMCT200020078CTTIP
[5]   CT coronary angiography at an ultra-low radiation dose (&lt;0.1 mSv): feasible and viable in times of constraint on healthcare costs [J].
Cademartiri, Filippo ;
Maffei, Erica ;
Arcadi, Teresa ;
Catalano, Onofrio ;
Midiri, Massimo .
EUROPEAN RADIOLOGY, 2013, 23 (03) :607-613
[6]   Radiation and contrast agent doses reductions by using 80-kV tube voltage in coronary computed tomographic angiography: A comparative study [J].
Cao, Jian-xin ;
Wang, Yi-min ;
Lu, Jin-guo ;
Zhang, Yu ;
Wang, Peng ;
Yang, Cheng .
EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) :309-314
[7]   Submillisievert Median Radiation Dose for Coronary Angiography with a Second-Generation 320-Detector Row CT Scanner in 107 Consecutive Patients [J].
Chen, Marcus Y. ;
Shanbhag, Sujata M. ;
Arai, Andrew E. .
RADIOLOGY, 2013, 267 (01) :76-85
[8]   Dose reduction with iterative reconstruction for coronary CT angiography: a systematic review and meta-analysis [J].
Den Harder, Annemarie M. ;
Willemink, Martin J. ;
De Ruiter, Quirina M. B. ;
De Jong, Pim A. ;
Schilham, Arnold M. R. ;
Krestin, Gabriel P. ;
Leiner, Tim ;
Budde, Ricardo P. J. .
BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1058)
[9]   High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate [J].
Gordic, Sonja ;
Husarik, Daniela B. ;
Desbiolles, Lotus ;
Leschka, Sebastian ;
Frauenfelder, Thomas ;
Alkadhi, Hatem .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 (06) :1173-1179
[10]   SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT [J].
Halliburton, Sandra S. ;
Abbara, Suhny ;
Chen, Marcus Y. ;
Gentry, Ralph ;
Mahesh, Mahadevappa ;
Raff, Gilbert L. ;
Shaw, Leslee J. ;
Hausleiter, Joerg .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2011, 5 (04) :198-224