A Microfluidics-based Pulpal Arteriole Blood Flow Phantom for Validation of Doppler Ultrasound Devices in Pulpal Blood Flow Velocity Measurement

被引:9
|
作者
Kim, Dohyun
Park, Sung-Ho [1 ,2 ]
机构
[1] Yonsei Univ, Coll Dent, Dept Conservat Dent, 50-1 Yonsei Ro, Seoul 03722, South Korea
[2] Yonsei Univ, Coll Dent, Oral Sci Res Ctr, 50-1 Yonsei Ro, Seoul 03722, South Korea
关键词
Doppler ultrasound; flow phantom; microfluidic chip; microfluidics; pulpal blood flow velocity; TEETH;
D O I
10.1016/j.joen.2016.08.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: Recently, Doppler ultrasound has been used for the measurement of pulpal blood flow in human teeth. However, the reliability of this method has not been verified. In this study, we developed a model to simulate arteriole blood flow within the dental pulp by using microfluidics. This arteriole simulator, or flow phantom, was used to determine the reliability of measurements obtained by using a Doppler ultrasound device. Methods: A microfluidic chip was fabricated by using. the soft lithography technique, and blood mimicking fluid was pumped through the channel by a microfluidic system. A Doppler ultrasound device was used for the measurement of flow velocity. The peak, mean, and minimal flow velocities obtained from the phantom and the Doppler ultrasound device were Compared by using linear regression analysis and Pear son correlation coefficient. Bland-Altman analyses were performed to evaluate the velocity differences between the flow generated by the phantom and the flow measurements made with the Doppler ultraSound device. Results: The microfluidic system was able to generate the flow profiles as intended, and the fluid flow could be monitored and controlled by the software program. There were excellent linear correlations between the peak, mean, and minimal flow velocities of the phantom and those of the Doppler ultrasound device (r = 0.94-0.996, P < .001). However, the velocities were overestimated by the Doppler ultrasound device: Conclusions: This phantom provides opportunities for research and education involving the Doppler ultrasound technique in dentistry. Although Doppler ultrasound can be an effective tool for the measurement of pulpal blood flow velocity, it is essential to validate and calibrate the device before clinical use. (J Endod 2016;42:1660-1666)
引用
收藏
页码:1660 / 1666
页数:7
相关论文
共 50 条
  • [21] ARTERIAL BLOOD-PRESSURE REGULATION OF PULPAL BLOOD-FLOW AS DETERMINED BY LASER DOPPLER
    SASANO, T
    KURIWADA, S
    SANJO, D
    JOURNAL OF DENTAL RESEARCH, 1989, 68 (05) : 791 - 795
  • [22] Pulpal blood flow measurement in human young permanent teeth.
    Komatsu, H
    Ikawa, M
    Mayanagi, H
    JOURNAL OF DENTAL RESEARCH, 2003, 82 : 437 - 437
  • [23] Pulpal blood flow responses depend on baseline blood flow levels.
    Sasano, T
    Shoji, N
    Kuriwada-Satoh, S
    Iikubo, M
    JOURNAL OF DENTAL RESEARCH, 1999, 78 (05) : 1117 - 1117
  • [24] Environmental effects on laser Doppler pulpal blood-flow measurements in man
    Hartmann, A
    Azerad, J
    Boucher, Y
    ARCHIVES OF ORAL BIOLOGY, 1996, 41 (04) : 333 - 339
  • [25] RELIABILITY OF PULPAL BLOOD-FLOW MEASUREMENTS UTILIZING LASER DOPPLER FLOWMETRY
    RAMSAY, DS
    ARTUN, J
    MARTINEN, SS
    JOURNAL OF DENTAL RESEARCH, 1991, 70 (11) : 1427 - 1430
  • [26] Effect of restorations on pulpal blood flow in molars measured by laser Doppler flowmetry
    Chandler, N. P.
    Ford, T. R. Pitt
    Monteith, B. D.
    INTERNATIONAL ENDODONTIC JOURNAL, 2010, 43 (01) : 41 - 46
  • [27] A laboratory study to detect simulated pulpal blood flow in extracted human teeth using ultrasound Doppler flowmetry
    Yoon, M. J.
    Kim, D. H.
    Jung, I. Y.
    Park, S. H.
    INTERNATIONAL ENDODONTIC JOURNAL, 2021, 54 (02) : 231 - 240
  • [28] Dependence of pulpal blood-flow responses on baseline blood-flow in the cat
    Sasano, T
    Shoji, N
    Kuriwada-Satoh, S
    Iikubo, M
    Izumi, H
    Karita, K
    ARCHIVES OF ORAL BIOLOGY, 2002, 47 (02) : 131 - 137
  • [29] PULPAL BLOOD-FLOW AFTER SUBAPICAL OSTEOTOMY
    CAVANAUGH, G
    MEYER, M
    ISAACSON, R
    JOURNAL OF DENTAL RESEARCH, 1975, 54 : L140 - L140
  • [30] Effect of limited exercise on pulpal blood flow in man
    Soo-Ampon, S
    Thiensiri, I
    Soo-Ampon, M
    Vongsavon, N
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : A62 - A62