Matrix effects of concomitant species on the determination of nickel by inductively coupled plasma atomic emission spectrometry

被引:0
作者
Akba, O [1 ]
Baysal, A [1 ]
Merdivan, M [1 ]
Hamamci, C [1 ]
Gumgum, B [1 ]
机构
[1] Dicle Univ, Fac Sci & Arts, Dept Chem, TR-21280 Diyarbakir, Turkey
关键词
matrix effect; concomitant species; nickel; ICP-AES;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The interference on the determination of nickel by the salts LiCl, NaCl, KCl, CsCl, NH4Cl, Na2SO4, NaNO3, NaC2H3O2 in an inductively coupled plasma-atomic emission spectrometry has been carried out. The salt concentration was in the 100-10000 mg L-1 range. Nickel emission was enhanced at lower matrix concentrations (100 mg L-1) and was depressed at higher matrix concentrations (10000 mg L-1). Except cesium, above 100 mg L-1, all studied ions caused a decrease in the signal intensity in the range of 0-15%. The sequence order of the investigated ions for the enhancement and depression was unstable. The sequence order in enhancement was found as Li+ < Na+ < K+ < Cs+ < NH4+ for cation; NO3- < SO42- < Cl- < OAc- for anions and Cs+ < NH4+ < K+ < Li+ < Na+ for cat-ions and OAc- < SO42- Cl- < NO3- for anions in depression at the studied lines.
引用
收藏
页码:1518 / 1526
页数:9
相关论文
共 38 条
[1]  
ABDALLAH MH, 1976, ANAL CHIM ACTA, V84, P238
[2]   REMOVAL OF A MATRIX EFFECT IN ICP-AES MULTIELEMENT ANALYSIS BY SIMPLEX OPTIMIZATION [J].
BELCHAMBER, RM ;
BETTERIDGE, D ;
WADE, AP ;
CRUICKSHANK, AJ ;
DAVISON, P .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (05) :503-505
[3]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[4]  
BOUNMANS PWJ, 1976, SPECTROCHIM ACTA B, V31, P355
[5]   Compensation of Ca and Na interference effects in axially and radially viewed inductively coupled plasmas [J].
Brenner, IB ;
Le Marchand, A ;
Daraed, C ;
Chauvet, L .
MICROCHEMICAL JOURNAL, 1999, 63 (03) :344-355
[6]   Comparison of axially and radially viewed inductively coupled plasmas for multi-element analysis: Effect of sodium and calcium [J].
Brenner, IB ;
Zander, A ;
Cole, M ;
Wiseman, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (09) :897-906
[7]   Matrix interferences in the determination of trace elements in waste waters by inductively coupled plasma atomic emission spectrometry with ultrasonic nebulization [J].
Budic, B .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 368 (04) :371-377
[8]   Elimination of nitric acid interference in ICP-AES by using a cyclonic spray chamber/Nafion membrane-based desolvation system [J].
Canals, A ;
Gras, L ;
Contreras, H .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (03) :219-226
[9]   PLASMA SOURCE-MASS SPECTROMETRY USING AN INDUCTIVELY COUPLED PLASMA AND A HIGH-RESOLUTION QUADRUPOLE MASS FILTER [J].
DATE, AR ;
GRAY, AL .
ANALYST, 1981, 106 (1269) :1255-1267
[10]   Comparison of ionic line-based internal standardization with axially and radially viewed inductively coupled plasma atomic emission spectrometry to compensate for sodium effects on accuracy [J].
Dubuisson, C ;
Poussel, E ;
Mermet, JM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (11) :1265-1269