Simultaneous determination of cobalt and nickel in catalyst by microwave digestion-dual wavelength equal absorption spectrophotometry

被引:0
作者
Li, LH [1 ]
Zhang, JS [1 ]
Gao, CL [1 ]
机构
[1] Fushan Petr Inst, Dept Appl Chem, Fushun 113001, Peoples R China
关键词
microwave digestion; dual wavelength equal absorption spectrophotometry; catalyst; cobalt; nickel;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Catalyst samples are digested in a microwave digestion system. The optimum parameters for microwave digestion are selected. Cobalt and nickel in the mixture of Co2+ and Ni2+ with 4-(2-pyridylazo) resorcinol (PAR) can be determined simultaneously by dual-wavelength equal absorption spectrophotometry. By means of the combination of two methods, Co2+ and Ni2+ in catalyst can be determined rapidly, accurately, and contamination problems avoided. The linear ranges are 0-30 mug(.)25 mL(-1) for Co2+ and 0-25 mug(.)25 mL(-1) for Ni2+. The recoveries of Co2+ and Ni2+ in synthetic samples are between 98.2 % - 103.6 % and between 97.9 % - 103.7 %, respectively. The relative standard deviations of analytical results in catalyst samples are less than 2.2 % for Co2+ and less than 1.8 % for Ni2+, and relative errors are less than +/- 2.5 % for Co2+ and Ni2+
引用
收藏
页码:531 / 533
页数:3
相关论文
共 1 条
[1]  
WU SH, 1978, PHYS TEST CHEM ANA B, P39