A Cross-Calibration Study of GE Lunar iDXA and GE Lunar DPX Pro for Body Composition Measurements in Children and Adults

被引:15
作者
Khadilkar, Anuradha [1 ]
Chiplonkar, Shashi [1 ]
Sanwalka, Neha [2 ]
Khadilkar, Vaman [1 ]
Mandlik, Rubina [1 ]
Ekbote, Veena [1 ]
机构
[1] Jehangir Hosp, Hirabai Cowasji Jehangir Med Res Inst, Block 5,Lower Ground Floor,32 Sassoon Rd, Pune 411001, Maharashtra, India
[2] NutriCanvas, Mumbai, Maharashtra, India
关键词
Cross calibration; GE Lunar iDXA; GE Lunar DPX-Pro; phantom scan; X-RAY ABSORPTIOMETRY; PENCIL-BEAM; PRODIGY; DXA; PRECISION;
D O I
10.1016/j.jocd.2019.03.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To cross-calibrate dual energy X-ray absorptiometry machines when replacing GE Lunar DPX-Pro with GE Lunar iDXA. Methods: A cross-sectional study was conducted in 126 children (3-19 years) and 135 adults (20-66 years). Phantom cross calibration was carried out using aluminum phantom provided with each of the machines on both machines. Total body less head (TBLH), lumbar spine (L2-L4) and left femoral neck bone mineral density (BMD), bone mineral content (BMC), and bone area were assessed for each patient on both machines. TBLH lean and fat mass were also measured. Bland-Altman analysis, linear regressions, and independent sample t test were performed to evaluate consistency of measurements and to establish cross-calibration equations. Results: iDXA measured 0.33% lower BMD and 0.64% lower BMC with iDXA phantom as compared to DPX-Pro phantom (p < 0.001). In children, TBLH-BMC, femoral BMC and area were measured 10% -14% lesser, TBLH area was higher by 1%-2% and L2-L4 area by 10%-14% by iDXA as compared to DPX-Pro. iDXA measured higher TBLH fat [15% (girls), 31% (boys)] than DPX-Pro. In adults, TBLH-BMD (1.7%-3.4%), BMC (6.0%-10.9%) and area (4.2%-7.6%) were measured lesser by iDXA than DPX-Pro. L2-L4 BMD was higher [2.7% (men), 1.8% (women)] by iDXA than DPX-Pro. Femoral BMC was 2.11% higher in men and 4.1% lower in women by iDXA as compared to DPX-Pro. In children, R 2 of cross-calibration equations, ranged from 0.91 to 0.96; in adults, it ranged from 0.93 to 0.99 (p < 0.01). After the regression equations were applied, differences in BMD values between both machines were negligible. Conclusion: A strong agreement for bone mass and body composition was established between both machines. Cross-calibration equations need to be applied to transform DPX-Pro measurements into iDXA measurements to avoid errors in assessment. This study documents a need for use of cross-calibration equations to transform DPX-Pro body composition data into iDXA values for clinical diagnosis.
引用
收藏
页码:128 / 137
页数:10
相关论文
共 18 条
[1]   Official positions of the International Society for Clinical Densitometry and executive summary of the 2007 ISCD Position Development Conference [J].
Baim, Sanford ;
Binkley, Neil ;
Bilezikian, John R. ;
Kendler, David L. ;
Hans, Didier B. ;
Lewiecki, E. Michael ;
Silverman, Stuart .
JOURNAL OF CLINICAL DENSITOMETRY, 2008, 11 (01) :75-91
[2]  
Bhudhikanok GS, 1996, J BONE MINER RES, V11, P1545
[3]   Replacing DXA scanners: Cross-calibration with phantoms may be misleading [J].
Blake, GM .
CALCIFIED TISSUE INTERNATIONAL, 1996, 59 (01) :1-5
[4]   Applying the right statistics: analyses of measurement studies [J].
Bland, JM ;
Altman, DG .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 22 (01) :85-93
[5]   Cross-Calibration of iDXA and Prodigy on Spine and Femur Scans in Korean Adults [J].
Choi, Yong Jun ;
Lee, Byung Joo ;
Lim, Hyun Chae ;
Chung, Yoon-Sok .
JOURNAL OF CLINICAL DENSITOMETRY, 2009, 12 (04) :450-455
[6]   Pediatric in vivo cross-calibration between the GE Lunar Prodigy and DPX-L bone densitometers [J].
Crabtree, NJ ;
Shaw, NJ ;
Boivin, CM ;
Oldroyd, B ;
Truscott, JG .
OSTEOPOROSIS INTERNATIONAL, 2005, 16 (12) :2157-2167
[7]   Dual energy X-ray absorptiometry interpretation and reporting in children and adolescents: The 2007 ISCD Pediatric Official Positions [J].
Gordon, Catherine M. ;
Bachrach, Laura K. ;
Carpenter, Thomas O. ;
Crabtree, Nicola ;
El-Hajj Fuleihan, Ghada ;
Kutilek, Stepan ;
Lorenc, Roman S. ;
Tosi, Laura L. ;
Ward, Katherine A. ;
Ward, Leanne M. ;
Kalkwarf, Heidi J. .
JOURNAL OF CLINICAL DENSITOMETRY, 2008, 11 (01) :43-58
[8]   iDXA, Prodigy, and DPXL Dual-Energy X-ray Absorptiometry Whole-Body Scans: A Cross-Calibration Study [J].
Hull, Holly ;
He, Qing ;
Thornton, John ;
Javed, Fahad ;
Allen, Lynn ;
Wang, Jack ;
Pierson, Richard N., Jr. ;
Gallagher, Dympna .
JOURNAL OF CLINICAL DENSITOMETRY, 2009, 12 (01) :95-102
[9]   Revised IAP Growth Charts for Height, Weight and Body Mass Index for 5- to 18-year-old Indian Children [J].
Khadilkar, Vaman ;
Yadav, Sangeeta ;
Agrawal, K. K. ;
Tamboli, Suchit ;
Banerjee, Monidipa ;
Cherian, Alice ;
Goyal, Jagdish P. ;
Khadilkar, Anuradha ;
Kumaravel, V. ;
Mohan, V. ;
Narayanapa, D. ;
Ray, I. ;
Yewale, Vijay .
INDIAN PEDIATRICS, 2015, 52 (01) :47-55
[10]   Body composition results by DXA differ with manufacturer, instrument generation and software version [J].
Kistorp, CN ;
Svendsen, OL .
APPLIED RADIATION AND ISOTOPES, 1998, 49 (5-6) :515-516