Energy-resolved X-ray radiography with controlled-drift detectors at Sincrotrone Trieste

被引:5
作者
Castoldi, A
Galimberti, A
Guazzoni, C
Rehak, P
Strüder, L
Menk, RH
机构
[1] Politecn Milan, Dipartimento Ingn Nucl, I-20133 Milan, Italy
[2] Ist Nazl Fis Nucl, Sez Milano, Milan, Italy
[3] Brookhaven Natl Lab, Upton, NY 11973 USA
[4] Halbleiterlabor Max Planck Inst, D-81279 Munich, Germany
[5] Sincrotrone Trieste, I-34012 Trieste, Italy
关键词
controlled drift detector; X-ray spectroscopy X-ray imaging; time-resolved X-ray imaging; energy-resolved X-ray imaging;
D O I
10.1016/S0168-9002(03)01679-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Controlled-Drift Detector (CDD) is a fully depleted silicon detector that allows 2-D position sensing and energy spectroscopy of X-rays in the range 1-20 keV with excellent time resolution. Its distinctive feature is the simultaneous readout of the charge packets stored in the detector by means of a uniform electrostatic field leading to readout times of few microseconds. The advantage of this readout mechanism is twofold: (i) a higher frame rate/better time resolution with respect to the charge-coupled device which represents the reference X-ray spectroscopic imager and (ii) a lower contribution of the thermal noise due to a shorter integration time, leading to an excellent energy resolution also at room temperature. In this work we present the first experimental characterization of the CDD with synchrotron light in the range 8-30 keV carried out at Sincrotrone Trieste. Two-dimensional energy-resolved radiographic images carried out at a frame frequency up to 100 kHz are shown. Application of the CDD to elemental absorption contrast imaging is also presented. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:57 / 62
页数:6
相关论文
共 9 条
[1]   Radiation detectors in medical and biological applications [J].
Besch, HJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 419 (2-3) :202-216
[2]   Spectroscopic-grade X-ray imaging up to 100-kHz frame rate with controlled-drift detectors [J].
Castoldi, A ;
Guazzoni, C ;
Rehak, P ;
Strüder, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (04) :982-986
[3]   The controlled-drift detector [J].
Castoldi, A ;
Gatti, E ;
Guazzoni, C ;
Longoni, A ;
Rehak, P ;
Strüder, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 439 (2-3) :519-528
[4]   Status of the digital pixel array detector for protein crystallography [J].
Datte, P ;
Birkbeck, A ;
Beuville, E ;
Endres, N ;
Druillole, F ;
Luo, L ;
Millaud, J ;
Xuong, NH .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 421 (03) :576-590
[5]   TIME-RESOLVED X-RAY-DIFFRACTION OF BIOLOGICAL-MATERIALS [J].
GRUNER, SM .
SCIENCE, 1987, 238 (4825) :305-312
[6]   X-RAY INTERACTIONS - PHOTOABSORPTION, SCATTERING, TRANSMISSION, AND REFLECTION AT E=50-30,000 EV, Z=1-92 [J].
HENKE, BL ;
GULLIKSON, EM ;
DAVIS, JC .
ATOMIC DATA AND NUCLEAR DATA TABLES, 1993, 54 (02) :181-342
[7]   Identification of materials in integrated circuit interconnects using x-ray absorption near-edge spectroscopy [J].
Levine, ZH ;
Ravel, B .
JOURNAL OF APPLIED PHYSICS, 1999, 85 (01) :558-564
[8]  
0862226
[9]  
6249033