摘要
This paper presents a capillary electrophoresis poly(methyl methacrylate) (PMMA) based microchip for electrochemical detection applications featuring embedded gold nanoelectrode ensemble (GNEE) working and decoupler electrodes. In fabricating the microchip, the GNEE films are pressed directly onto the metallic electrode structures using a hot embossing technique, and the microfluidic channels are then sealed using a low-temperature azeotropic solvent bonding method. The detection performance of the microchip is evaluated using dopamine and catechol analytes for illustration purposes. The experimental results show that the GNEE working electrode provides a significantly higher signal response than that obtained from a bulk gold electrode when applied to the detection of dopamine analyte. Compared to a conventional bulk palladium clecoupler electrode, the GNEE clecoupler electrode reduces both the amplitude of the charge current (3.5 nA vs. 18.7 nA) and the baseline drift at higher separation voltages. The measured baseline current drift for the microchip equipped the proposed GNEE clecoupler electrode is around three times smaller than the microchip with the palladium decoupler electrode under the applied separation electric field from 40 V/cm, to 240 V/cm. Finally, when detecting a mixture of 1 mM dopamine and 1 mM catechol, the calculated signal response of the microchip with a GNEE clecoupler electrode is approximately five times higher than that obtained from a microchip with a bulk I'd clecoupler electrode, resulting in the detection limit of 1 mu M for the proposed GNEE-based microchip device. Overall, the results indicate that the proposed capillary electrophoresis-electrochemical detection (CE-ED) microchip with embedded GNEE working and decoupler electrodes provides an ideal solution for sample detection in lab-on-a-chip and micro total analysis applications. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:231 / 236
页数:6
相关论文
共 46 条
[1]   Independent geometrical and electrochemical characterization of arrays of nanometer-scale electrodes [J].
Baker, WS ;
Crooks, RM .
JOURNAL OF PHYSICAL CHEMISTRY B, 1998, 102 (49) :10041-10046
[2]   Electrochemical detector for microchip electrophoresis of poly(dimethylsiloxane) with a three-dimensional adjustor [J].
Bao, N ;
Xu, JJ ;
Dou, YH ;
Cai, Y ;
Chen, HY ;
Xia, XH .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1041 (1-2) :245-248
[3]   Palladium film decoupler for amperometric detection in electrophoresis chips [J].
Chen, DC ;
Hsu, FL ;
Zhan, DZ ;
Chen, CH .
ANALYTICAL CHEMISTRY, 2001, 73 (04) :758-762
[4]   Colloid chemical approach to nanoelectrode ensembles with highly controllable active area fraction [J].
Cheng, WL ;
Dong, SJ ;
Wang, EK .
ANALYTICAL CHEMISTRY, 2002, 74 (15) :3599-3604
[5]   Electrochemical characterization of electrodes with submicrometer dimensions [J].
Conyers, JL ;
White, HS .
ANALYTICAL CHEMISTRY, 2000, 72 (18) :4441-4446
[6]   Integrated microfluidic device with an electroplated palladium decoupler for more sensitive amperometric detection of the 8-hydroxy-deoxyguanosine (8-OH-dG) DNA adduct [J].
Dawoud, Abdulilah A. ;
Kawaguchi, Toshikazo ;
Jankowiak, Ryszard .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (01) :245-252
[7]  
Dolník V, 2000, ELECTROPHORESIS, V21, P41, DOI 10.1002/(SICI)1522-2683(20000101)21:1<41::AID-ELPS41>3.0.CO
[8]  
2-7
[9]   Integrated capillary electrophoresis on flexible silicone microdevices: Analysis of DNA restriction fragments and detection of single DNA molecules on microchips [J].
Effenhauser, CS ;
Bruin, GJM ;
Paulus, A ;
Ehrat, M .
ANALYTICAL CHEMISTRY, 1997, 69 (17) :3451-3457
[10]   Ultrasensitive electrocatalytic DNA detection at two- and three-dimensional nanoelectrodes [J].
Gasparac, R ;
Taft, BJ ;
Lapierre-Devlin, MA ;
Lazareck, AD ;
Xu, JM ;
Kelley, SO .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2004, 126 (39) :12270-12271