PET/MR brain imaging: evaluation of clinical UTE-based attenuation correction

被引:39
作者
Aasheim, Lars Birger [1 ,2 ]
Karlberg, Anna [2 ]
Goa, Pal Erik [2 ,3 ]
Haberg, Asta [4 ,8 ]
Sorhaug, Sveinung [5 ]
Fagerli, Unn-Merete [6 ,7 ]
Eikenes, Live [1 ]
机构
[1] Norwegian Univ Sci & Technol NTNU, Dept Circulat & Med Imaging, N-7489 Trondheim, Norway
[2] St Olavs Univ Hosp, Clin Radiol & Nucl Med, Trondheim, Norway
[3] NTNU, Dept Phys, Trondheim, Norway
[4] NTNU, Dept Neurosci, Trondheim, Norway
[5] St Olavs Univ Hosp, Dept Thorac Med, Trondheim, Norway
[6] St Olavs Univ Hosp, Dept Oncol, Trondheim, Norway
[7] NTNU, Dept Canc Res & Mol Med, Trondheim, Norway
[8] St Olavs Univ Hosp, Dept Med Imaging, Trondheim, Norway
关键词
PET/MR; Attenuation correction; UTE; Neuroimaging; MAGNETIC-RESONANCE; SYSTEMS;
D O I
10.1007/s00259-015-3060-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
One of the greatest challenges in PET/MR imaging is that of accurate MR-based attenuation correction (AC) of the acquired PET data, which must be solved if the PET/MR modality is to reach its full potential. The aim of this study was to investigate the performance of Siemens' most recent version (VB20P) of MR-based AC of head PET data, by comparing it to CT-based AC. Methods:F-18-FDG PET data from seven lymphoma and twelve lung cancer patients examined with a Biograph mMR PET/MR system were reconstructed with both CT-based and MR-based AC, avoiding sources of error arising when comparing PET data from different systems. The resulting images were compared quantitatively by measuring changes in mean SUV in ten different brain regions in both hemispheres, as well as the brainstem. In addition, the attenuation maps (mu maps) were compared regarding volume and localization of cranial bone. Results: The UTE mu maps clearly overestimate the amount of bone in the neck, while slightly underestimating the amount of bone in the cranium, and the localization of bone in the cranial region also differ from the CT mu maps. In air/tissue interfaces in the sinuses and ears, the MRAC method struggles to correctly classify the different tissues. The misclassification of tissue is most likely caused by a combination of artefacts and the insufficiency of the UTE method to accurately separate bone. Quantitatively, this results in a combination of overestimation (0.5-3.6 %) and underestimation (2.7-5.2 %) of PET activity throughout the brain, depending on the proximity to the inaccurate regions. Conclusions: Our results indicate that the performance of the UTE method as implemented in VB20P is close to the theoretical maximum of such an MRAC method in the brain, while it does not perform satisfactorily in the neck or face/nasal area. Further improvement of the UTE MRAC or other available methods for more accurate segmentation of bone should be incorporated.
引用
收藏
页码:1439 / 1446
页数:8
相关论文
共 27 条
[1]   Improved UTE-based attenuation correction for cranial PET-MR using dynamic magnetic field monitoring [J].
Aitken, A. P. ;
Giese, D. ;
Tsoumpas, C. ;
Schleyer, P. ;
Kozerke, S. ;
Prieto, C. ;
Schaeffter, T. .
MEDICAL PHYSICS, 2014, 41 (01)
[2]   MRI-guided attenuation correction in whole-body PET/MR: assessment of the effect of bone attenuation [J].
Akbarzadeh, A. ;
Ay, M. R. ;
Ahmadian, A. ;
Alam, N. Riahi ;
Zaidi, H. .
ANNALS OF NUCLEAR MEDICINE, 2013, 27 (02) :152-162
[3]   Combined PET/MR imaging in neurology: MR-based attenuation correction implies a strong spatial bias when ignoring bone [J].
Andersen, Flemming Littrup ;
Ladefoged, Claes Nohr ;
Beyer, Thomas ;
Keller, Sune Hogild ;
Hansen, Adam Espe ;
Hojgaard, Liselotte ;
Kjaer, Andreas ;
Law, Ian ;
Holm, Soren .
NEUROIMAGE, 2014, 84 :206-216
[4]   Whole-body PET/MRI: The effect of bone attenuation during MR-based attenuation correction in oncology imaging [J].
Aznar, M. C. ;
Sersar, R. ;
Saabye, J. ;
Ladefoged, C. N. ;
Andersen, F. L. ;
Rasmussen, J. H. ;
Lofgren, J. ;
Beyer, T. .
EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (07) :1177-1183
[5]  
Bailey D.L., 2005, POSITRON EMISSION TO, P13, DOI [10.1007/1-84628-007-9_2, DOI 10.1007/1-84628-007-9_2]
[6]   MRI-Based Attenuation Correction for Hybrid PET/MRI Systems: A 4-Class Tissue Segmentation Technique Using a Combined Ultrashort-Echo-Time/Dixon MRI Sequence [J].
Berker, Yannick ;
Franke, Jochen ;
Salomon, Andre ;
Palmowski, Moritz ;
Donker, Henk C. W. ;
Temur, Yavuz ;
Mottaghy, Felix M. ;
Kuhl, Christiane ;
Izquierdo-Garcia, David ;
Fayad, Zahi A. ;
Kiessling, Fabian ;
Schulz, Volkmar .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (05) :796-804
[7]   MR-Based PET Attenuation Correction for PET/MR Imaging [J].
Bezrukov, Ilja ;
Mantlik, Frederic ;
Schmidt, Holger ;
Schoelkopf, Bernhard ;
Pichler, Bernd J. .
SEMINARS IN NUCLEAR MEDICINE, 2013, 43 (01) :45-59
[8]  
Block KT., 2011, P INT SOC MAG RESON, V19, P2816
[9]   Toward Implementing an MRI-Based PET Attenuation-Correction Method for Neurologic Studies on the MR-PET Brain Prototype [J].
Catana, Ciprian ;
van der Kouwe, Andre ;
Benner, Thomas ;
Michel, Christian J. ;
Hamm, Michael ;
Fenchel, Matthias ;
Fischl, Bruce ;
Rosen, Bruce ;
Schmand, Matthias ;
Sorensen, A. Gregory .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (09) :1431-1438
[10]  
CHEETHAM A H, 1969, Journal of Paleontology, V43, P1130