Experimental validation of a flat punch indentation methodology calibrated against unconfined compression tests for determination of soft tissue biomechanics

被引:59
作者
Delaine-Smith, R. M. [1 ,2 ]
Burney, S. [1 ]
Balkwill, F. R. [2 ]
Knight, M. M. [1 ]
机构
[1] Queen Mary Univ London, Inst Bioengn, Sch Engn & Mat Sci, London E1 4NS, England
[2] Queen Mary Univ London, Barts Canc Inst, Ctr Canc & Inflammat, Charterhouse Sq, London EC1M 6BQ, England
基金
欧洲研究理事会;
关键词
Tissue mechanics; Viscoelasticity; Flat-ended indentation; Cartilage; Agarose hydrogel; FINITE-ELEMENT-ANALYSIS; ARTICULAR-CARTILAGE; STRESS-RELAXATION; ELASTIC-MODULUS; YOUNGS MODULUS; POISSONS RATIO;
D O I
10.1016/j.jmbbm.2016.02.019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mechanical characterisation of soft biological tissues using standard compression or tensile testing presents a significant challenge due to specimen geometrical irregularities, difficulties in cutting intact and appropriately sized test samples, and issues with slippage or damage at the grips. Indentation can overcome these problems but requires fitting a model to the resulting load-displacement data in order to calculate moduli. Despite the widespread use of this technique, few studies experimentally validate their chosen model or compensate for boundary effects. In this study, viscoelastic hydrogels of different concentrations and dimensions were used to calibrate an indentation technique performed at large specimen-strain deformation (20%) and analysed with a range of routinely used mathematical models. A rigid, flat-ended cylindrical indenter was applied to each specimen from which 'indentation moduli' and relaxation properties were calculated and compared against values obtained from unconfined compression. Only one indentation model showed good agreement (<10% difference) with all moduli values obtained from compression. A sample thickness to indenter diameter ratio >= 1:1 and sample diameter to indenter diameter ratio >= 4:1 was necessary to achieve the greatest accuracy. However, it is not always possible to use biological samples within these limits, therefore we developed a series of correction factors. The approach was validated using human diseased omentum and bovine articular cartilage resulting in mechanical properties closely matching compression values. We therefore present a widely useable indentation analysis method to allow more accurate calculation of material mechanics which is important in the study of soft tissue development, ageing, health and disease. (C) 2016 The Authors. Published by Elsevier Ltd.
引用
收藏
页码:401 / 415
页数:15
相关论文
共 40 条
[1]   The mechanical properties of human adipose tissues and their relationships to the structure and composition of the extracellular matrix [J].
Alkhouli, Nadia ;
Mansfield, Jessica ;
Green, Ellen ;
Bell, James ;
Knight, Beatrice ;
Liversedge, Neil ;
Tham, Ji Chung ;
Welbourn, Richard ;
Shore, Angela C. ;
Kos, Katarina ;
Winlove, C. Peter .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2013, 305 (12) :E1427-E1435
[2]   The effect of concentration, thermal history and cell seeding density on the initial mechanical properties of agarose hydrogels [J].
Buckley, Conor T. ;
Thorpe, Stephen D. ;
O'Brien, Fergal J. ;
Robinson, Anthony J. ;
Kelly, Daniel J. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2009, 2 (05) :512-521
[3]  
Bueckle H., 1973, The Science of Hardness Testing and its Research Applications
[4]   Nanoindentation of polydimethylsiloxane elastomers: Effect of crosslinking, work of adhesion, and fluid environment on elastic modulus [J].
Carrillo, F ;
Gupta, S ;
Balooch, M ;
Marshall, SJ ;
Marshall, GW ;
Pruitt, L ;
Puttlitz, CM .
JOURNAL OF MATERIALS RESEARCH, 2005, 20 (10) :2820-2830
[5]   Material characterization of ex vivo prostate tissue via spherical indentation in the clinic [J].
Carson, William C. ;
Gerling, Gregory J. ;
Krupski, Tracey L. ;
Kowalik, Casey G. ;
Harper, Jeffrey C. ;
Moskaluk, Christopher A. .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (03) :302-309
[6]  
Chen X., 2011, J MATER RES, V16, P2974
[7]   The compressive response of porcine adipose tissue from low to high strain rate [J].
Comley, Kerstyn ;
Fleck, Norman .
INTERNATIONAL JOURNAL OF IMPACT ENGINEERING, 2012, 46 :1-10
[8]   Soft tissue elastometer [J].
Egorov, V. ;
Tsyuryupa, S. ;
Kanilo, S. ;
Kogit, M. ;
Sarvazyan, A. .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (02) :206-212
[9]   Non-ideal effects in indentation testing of soft tissues [J].
Finan, John D. ;
Fox, Patrick M. ;
Morrison, Barclay, III .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2014, 13 (03) :573-584
[10]   A review of analysis methods for sub-micron indentation testing [J].
Fischer-Cripps, AC .
VACUUM, 2000, 58 (04) :569-585