Intra-observer and inter-observer reliability and reproducibility of slit-lamp-adapted optical coherence tomography for evaluation of anterior chamber depth and central corneal thickness

被引:10
作者
Sandler, Shlomit F. [1 ]
Zelefsky, Joseph R. [2 ,3 ]
Dorairaj, Syril [4 ]
Arthur, Stella N. [4 ]
Ritch, Robert [4 ,5 ]
Liebmann, Jeffrey M. [2 ,3 ,4 ]
机构
[1] Mt Sinai Sch Med, New York, NY USA
[2] NYU, Sch Med, New York, NY USA
[3] Manhattan Eye Ear & Throat Hosp, New York, NY 10021 USA
[4] New York Eye & Ear Infirm, Einhorn Clin Res Ctr, New York, NY 10003 USA
[5] New York Med Coll, New York, NY USA
关键词
D O I
10.3928/15428877-20080701-15
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
BACKGROUND AND OBJECTIVE: To evaluate the intra-observer and inter-observer reproducibility of anterior chamber depth (ACD) and central corneal thickness (CCT) measurements taken by slit-lamp-adapted optical coherence tomography (SL-OCT) and to evaluate concordance of CCT and ACD values obtained by SL-OCT with gold standard devices. m PATIENTS AND METHODS: Three images of each eye were taken by two different operators using SL-OCT ACD and CCT values obtained by each operator were determined and compared to one another and to ultrasonic pachymetry (CCT) and axial OCT biometry (ACD) values obtained by a third, masked operator. Intra-observer and inter-operator reproducibility was assessed using the intraclass correlation coefficient (ICC). RESULTS: Forty-one eyes of healthy subjects were enrolled. Mean CCT was 0.556 +/- 0.020, 0.557 +/- 0.019, and 0.532 +/- 0.032 and mean ACD was 3.12 +/- 0.36, 3.13 +/- 0.38, and 3.50 +/- 0.44 for operators 1, 2, and 3, respectively (P <.001, for both CCT and ACD). Intra-operator reproducibility for CCT (ICC of 0.948 to 0.995) and ACD (ICC of 0.972 to 0.987) was excellent. CONCLUSION: SL-OCT provides highly reproducible measures for both CCT and ACD. SL-OCT yields higher values for CCT compared to ultrasonic pachymetry and lower values for ACD compared to axial OCT biometry. Caution should be taken when extrapolating these data into clinical practice until the differences between SL-OCT axial OCT biometry, and ultrasonic pachymetry can be further elucidated.
引用
收藏
页码:299 / 303
页数:5
相关论文
共 14 条
[1]   Static and dynamic analysis of the anterior segment with optical coherence tomography [J].
Baikoff, G ;
Lutun, E ;
Ferraz, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (09) :1843-1850
[2]   Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry [J].
Bechmann, M ;
Thiel, MJ ;
Neubauer, AS ;
Ullrich, S ;
Ludwig, K ;
Kenyon, KR ;
Ulbig, MW .
CORNEA, 2001, 20 (01) :50-54
[3]   Comparison of three methods of measuring corneal thickness and anterior chamber depth [J].
Buehl, W ;
Stojanac, D ;
Sacu, S ;
Drexler, W ;
Findl, O .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) :7-12
[4]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[5]   Comparison of corneal pachymetry using ultrasound and Orbscan II [J].
Fakhry, MA ;
Artola, A ;
Belda, JL ;
Ayala, MJ ;
Alió, JL .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :248-252
[6]  
Gordon MO, 2002, ARCH OPHTHALMOL-CHIC, V120, P714
[7]   Slit-lamp-adapted optical coherence tomography of the anterior segment [J].
Hoerauf, H ;
Wirbelauer, C ;
Scholz, C ;
Engelhardt, R ;
Koch, P ;
Laqua, H ;
Birngruber, R .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (01) :8-18
[8]   Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomography [J].
Leung, Dexter Y. L. ;
Lam, Douglas K. T. ;
Yeung, Barry Y. M. ;
Lam, Dennis S. C. .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 34 (08) :751-754
[9]   Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode [J].
Meinhardt, B ;
Stachs, O ;
Stave, J ;
Beck, R ;
Guthoff, R .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 244 (05) :559-564
[10]   Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness [J].
Miglior, S ;
Albe, E ;
Guareschi, M ;
Mandelli, G ;
Gomarasca, S ;
Orzalesi, N .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (02) :174-177