First successful implantation of a biodegradable metal stent into the left pulmonary artery of a preterm baby

被引:336
作者
Zartner, P
Cesnjevar, R
Singer, H
Weyand, M
机构
[1] Deutsch Kinderherzzentrum Sankt Augustin, D-53757 St Augustin, Germany
[2] Univ Erlangen Nurnberg, Dept Cardiothorac Surg, Erlangen, Germany
[3] Univ Erlangen Nurnberg, Dept Pediat & Adolescent Med, Div Pediat Cardiol, Erlangen, Germany
关键词
corrosion; congenital heart disease; magnesium; arterial stenosis; pulmonary stenosis;
D O I
10.1002/ccd.20520
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Stent implantation in the youngest patients with a congenital heart disease implicates limitations concerning further vessel growth, the need of staged redilation, and later surgical removal. The search to overcome these restrictions led to open stent designs, with a wide adaptability to the vessel growth and recently to the development of bioabsorbable stent materials. A preterm baby born at 26 weeks of gestation was referred to our clinic following inadvertent ligation of the left pulmonary artery. Despite efficient debanding, the left lung perfusion was absent. Implantation of a biodegradable 3 mm magnesium stent was performed in a hybrid procedure when the baby weighed 1.7 kg. Reperfusion of the left lung was established and persisted throughout the 4-month follow-up period during which the gradual degradation process of the stent completed. Additional interventions, should they become necessary, seem not to be limited. Despite the small size of the baby, the degradation process was clinically well tolerated. The mechanical and degradation characteristics of the magnesium stent proved to be adequate to secure reperfusion of the previously occluded left pulmonary artery. Bioabsorbable stents with different diameters may help develop new strategies in the therapy of vessel stenosis in pediatric patients. (c) 2005 Wiley-Liss, Inc.
引用
收藏
页码:590 / 594
页数:5
相关论文
共 16 条
[1]   Cardiac surgery with extracorporeal circulation in 23 infants weighing 2500 g or less: short and intermediate term outcome [J].
Beyens, T ;
Biarent, D ;
Bouton, JM ;
Demanet, H ;
Viart, P ;
Dessy, H ;
Deville, A ;
Lamote, J ;
Deuvaert, FE .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1998, 14 (02) :165-172
[2]  
Di Mario Carlo, 2004, J Interv Cardiol, V17, P391, DOI 10.1111/j.1540-8183.2004.04081.x
[3]   Novel growth Stent for the permanent treatment of vessel stenosis in growing children: An experimental study [J].
Ewert, P ;
Riesenkampff, E ;
Neuss, M ;
Kretschmar, O ;
Nagdyman, N ;
Lange, PE .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (04) :506-510
[4]   Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology? [J].
Heublein, B ;
Rohde, R ;
Kaese, V ;
Niemeyer, M ;
Hartung, W ;
Haverich, A .
HEART, 2003, 89 (06) :651-656
[5]  
Hijazi ZM, 1996, CATHETER CARDIO DIAG, V38, P16
[6]   Stent implantation for aortic coarctation and recoarctation [J].
Magee, AG ;
Brzezinska-Rajszys, G ;
Qureshi, SA ;
Rosenthal, E ;
Zubrzycka, M ;
Ksiazyk, J ;
Tynan, M .
HEART, 1999, 82 (05) :600-606
[7]   Stenting of the ductus arteriosus and banding of the pulmonary arteries: basis for various surgical strategies in newborns with multiple left heart obstructive lesions [J].
Michel-Behnke, I ;
Akintuerk, H ;
Marquardt, I ;
Mueller, M ;
Thul, J ;
Bauer, J ;
Hagel, KJ ;
Kreuder, J ;
Vogt, P ;
Schranz, D .
HEART, 2003, 89 (06) :645-650
[8]   Catheterization treatment of stenosis and hypoplasia of pulmonary arteries [J].
O'Laughlin, MP .
PEDIATRIC CARDIOLOGY, 1998, 19 (01) :48-56
[9]   IMPLANTATION AND INTERMEDIATE-TERM FOLLOW-UP OF STENTS IN CONGENITAL HEART-DISEASE [J].
OLAUGHLIN, MP ;
SLACK, MC ;
GRIFKA, RG ;
PERRY, SB ;
LOCK, JE ;
MULLINS, CE .
CIRCULATION, 1993, 88 (02) :605-614
[10]   Outcome of cardiac surgery in low birth weight and premature infants [J].
Oppido, G ;
Napoleone, CP ;
Formigari, R ;
Gabbieri, D ;
Pacini, D ;
Frascaroli, G ;
Gargiulo, G .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 26 (01) :44-52