Universal lab-on-a-chip platform for complex, perfused 3D cell cultures

被引:4
作者
Sonntag, F. [1 ]
Schmieder, F. [1 ,2 ]
Stroebel, J. [1 ]
Gruenzner, S. [1 ,2 ]
Busek, M. [1 ]
Guenther, K. [2 ]
Steege, T. [1 ]
Polk, C. [1 ]
Klotzbach, U. [1 ]
机构
[1] Fraunhofer Inst Mat & Beam Technol IWS, Winterbergstr 28, D-01277 Dresden, Germany
[2] Tech Univ Dresden, Inst Mfg Technol, George Bahr Str 3c, D-01069 Dresden, Germany
来源
MICROFLUIDICS, BIOMEMS, AND MEDICAL MICROSYSTEMS XIV | 2016年 / 9705卷
关键词
lab-on-a-chip; microfluidic; 3D cell culture; laser; 3D printing; micro structuring; micropump; tissue engineering;
D O I
10.1117/12.2218606
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The miniaturization, rapid prototyping and automation of lab-on-a-chip technology play nowadays a very important role. Lab-on-a-chip technology is successfully implemented not only for environmental analysis and medical diagnostics, but also as replacement of animals used for the testing of substances in the pharmaceutical and cosmetics industries. For that purpose the Fraunhofer IWS and partners developed a lab-on-a-chip platform for perfused cell-based assays in the last years, which includes different micropumps, valves, channels, reservoirs and customized cell culture modules. This technology is already implemented for the characterization of different human cell cultures and organoids, like skin, liver, endothelium, hair follicle and nephron. The advanced universal lab-on-a-chip platform for complex, perfused 3D cell cultures is divided into a multilayer basic chip with integrated micropump and application-specific 3D printed cell culture modules. Moreover a technology for surface modification of the printed cell culture modules by laser micro structuring and a complex and flexibly programmable controlling device based on an embedded Linux system was developed. A universal lab-on-a-chip platform with an optional oxygenator and a cell culture module for cubic scaffolds as well as first cell culture experiments within the cell culture device will be presented. The module is designed for direct interaction with robotic dispenser systems. This offers the opportunity to combine direct organ printing of cells and scaffolds with the microfluidic cell culture module. The characterization of the developed system was done by means of Micro-Particle Image Velocimetry (mu PIV) and an optical oxygen measuring system.
引用
收藏
页数:12
相关论文
共 28 条
[1]  
Amao Y, 2000, J PORPHYR PHTHALOCYA, V4, P292, DOI 10.1002/(SICI)1099-1409(200004/05)4:3<292::AID-JPP216>3.0.CO
[2]  
2-W
[3]   Skin and hair on-a-chip: in vitro skin models versus ex vivo tissue maintenance with dynamic perfusion [J].
Atac, Beren ;
Wagner, Ilka ;
Horland, Reyk ;
Lauster, Roland ;
Marx, Uwe ;
Tonevitsky, Alexander G. ;
Azar, Reza P. ;
Lindner, Gerd .
LAB ON A CHIP, 2013, 13 (18) :3555-3561
[4]   A LIVING SYSTEM ON A CHIP [J].
Baker, Monya .
NATURE, 2011, 471 (7340) :661-665
[5]   Microfluidic organs-on-chips [J].
Bhatia, Sangeeta N. ;
Ingber, Donald E. .
NATURE BIOTECHNOLOGY, 2014, 32 (08) :760-772
[6]   Characterization and simulation of peristaltic micropumps [J].
Busek, M. ;
Noetzel, M. ;
Polk, C. ;
Sonntag, F. .
JOURNAL OF SENSORS AND SENSOR SYSTEMS, 2013, 2 (02) :165-169
[7]  
Busek M., 2012, BIOMEDICAL ENG
[8]   Hollow fiber membrane contactor as a gas-liquid model contactor [J].
Dindore, VY ;
Brilman, DWF ;
Versteeg, GE .
CHEMICAL ENGINEERING SCIENCE, 2005, 60 (02) :467-479
[9]  
Henry W., 1832, Abstracts of the Papers Printed in the Philosophical Transactions of the Royal Society of London, V1, P103
[10]   Pneumatic micropumps with serially connected actuation chambers [J].
Huang, Chun-Wei ;
Huang, Song-Bin ;
Lee, Gwo-Bin .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2006, 16 (11) :2265-2272